Page 2 - International Safety Symbols; Meter Description
2 SDL300-EU-EN V1.5 07/13 Introduction Congratulations on your purchase of the Extech SDL300 Thermo-Anemometer Datalogger. This meter displays and stores air velocity and temperature readings. Temperature readings from the thermometer built into the vane probe or from an externally connected type K ...
Page 3 - Operation
3 SDL300-EU-EN V1.5 07/13 Operation Power ON-OFF Power the meter by pressing and holding the power key for at least 1.5 seconds. Press and hold the power key for at least 1.5 seconds to power OFF the meter. Meter Functions This meter has two modes of operation. Press and hold the FUNCTION key fo...
Page 4 - Display Backlight; System Reset
4 SDL300-EU-EN V1.5 07/13 MAX-MIN Readings For a given measurement session, this meter can record the highest (MAX) and the lowest (MIN) readings for later recall. 1. Press the MAX-MIN key momentarily to access this mode of operation (REC icon appears) 2. The meter is now recording the MAX and MIN r...
Page 5 - Datalogger; Types of Data Recording
5 SDL300-EU-EN V1.5 07/13 Datalogger Types of Data Recording Manual Datalogging: Manually log up to 99 readings onto an SD card via push-key press. Automatic Datalogging: Automatically log data onto an SD memory card where the number of data points is limited only by the card size. Readings are ...
Page 6 - SD Data Card to PC Data Transfer
6 SDL300-EU-EN V1.5 07/13 SD Data Card to PC Data Transfer 1. Complete a datalogging session as covered above. For the first test, simply record a small amount of test data. 2. With the meter switched OFF, remove the SD Card. 3. Plug the SD Card directly into a PC SD card reader. If the PC does not ...
Page 7 - Setup Mode
7 SDL300-EU-EN V1.5 07/13 Setup Mode Basic settings at a glance To view the current configuration of the meter with regard to time, date, thermocouple type, and datalogging sampling rate, press the SET/CLOCK ICON key momentarily. The meter will now display the configuration in quick succession. Repe...
Page 9 - Battery Replacement and Disposal; Other Battery Safety Reminders
9 SDL300-EU-EN V1.5 07/13 Battery Replacement and Disposal When the low battery icon appears on the LCD, the batteries must be replaced. Several hours of accurate readings are still possible in this condition; however batteries should be replaced as soon as possible: 1. Remove the two (2) Phillips s...
Page 10 - Specifications; General Specifications
10 SDL300-EU-EN V1.5 07/13 Specifications General Specifications Display Backlit LCD; LCD size: 52 x 38mm (2 x 1.5”) Measurement Units Air velocity: m/S (meters per second) Km/h (kilometers per hour) Ft/min (FPM; feet per minute), Knots (nautical miles per hour), Mile/h (mph; miles per hour) Air tem...
Page 11 - Copyright
11 SDL300-EU-EN V1.5 07/13 Electrical Specifications (Ambient Temperature 23°C ± 5°C) Air velocity Measurement Range Resolution Accuracy m/S 0.4 - 35.0 m/s 0.1 m/S ± (2% rdg + 0.2 m/S) Km/h 1.4 - 126.0 km/h 0.1 Km/h ± (2% rdg + 0.8 Km/h) mph (Mile/h) 0.9 – 78.3 mile/h 0.1 mph ± (2% rdg + 0.4 mph) Kn...