Page 2 - Introduction; The supplied thermocouple is rated to a temperature of 482
42560-EU-EN-V5.1-9/11 2 Introduction Congratulations on your purchase of the Model 42560 IR Thermometer. The Model 42560 IR thermometer measures and displays non-contact (infrared) surface temperature to a maximum of 1922 o F (1050 o C). The built-in laser pointer increases target accuracy while the...
Page 3 - Descriptions; please read the entire User Guide.
42560-EU-EN-V5.1-9/11 3 Descriptions Meter Description 1. Laser pointer 2. IR sensor 3. Measurement trigger 4. Battery compartment 5. LCD display 6. Function buttons* 7. Thermocouple 8. Optical interface (meter bottom) 9. Thermocouple input jacks (meter bottom) *Function Buttons MODE : Select from...
Page 5 - Operating Instructions; Power; the LCD) after which the meter will automatically shut off.; Backlight/Laser Pointer; Press again to turn the backlight off.
42560-EU-EN-V5.1-9/11 5 Operating Instructions Power 1. The meter is powered by one (1) 9V battery. 2. The transmitter base can also be powered by the 9V battery in the meter or by the supplied AC adaptor. IR (non-contact) Measurements 1. Hold the meter by its handle and point it toward the surface ...
Page 7 - IR Measurement Notes; diagram (printed on the side of the meter and in this guide).; Field of View
42560-EU-EN-V5.1-9/11 7 IR Measurement Notes 1. The object under test should be larger than the spot (target) size calculated by the field of view diagram (printed on the side of the meter and in this guide). 2. Before measuring, be sure to clean surfaces that are covered with frost, oil, grime, etc...
Page 9 - Datalogging
42560-EU-EN-V5.1-9/11 9 Datalogging Wireless Datalogging The Model 42560 is supplied with a Wireless Datalogging system for use with a PC. The kit contains software (CD-ROM), Transmitter base, and a USB Data Receiver. Readings can be automatically transferred from the IR thermometer to a PC using th...
Page 10 - —Consult the dealer or an experienced radio/TV technician for help.
42560-EU-EN-V5.1-9/11 10 Analog Output The Model 42560 is supplied with a banana plug pair to 3.5mm mini-plug cable. The mini-plug end of the cable connects to the meter’s Transmitter base. The banana plug test lead end can be connected to an external device such as a MultiMeter, chart recorder, mag...
Page 11 - Specifications; or target; Type K Thermocouple (Contact) Specifications
42560-EU-EN-V5.1-9/11 11 Specifications Infrared Thermometer (Non-contact) Specifications Range / Resolution -50.0 to 1050.0 o C (-58.0 to 1922.0 o F) Resolution: 0.1 o Accuracy ± 5 o C (± 9 o F) from -50 to -20 o C (-58 to -4 o F) ± (1.5% of reading + 2 o C [3.6 o F]) from -20 to 200 o C (-4 to 392...
Page 12 - Battery Replacement; Battery ordinance; ) to return all used batteries; disposal in the household garbage is prohibited!; end of its lifecycle; Copyright
42560-EU-EN-V5.1-9/11 12 Battery Replacement When the low battery symbol appears on the LCD, replace the meter’s battery (9V). The battery compartment is located behind the panel that surrounds the meter’s trigger (see diagram). Open the compartment by carefully pulling the panel down away from the ...