Page 3 - OUNDATION

Page 3 SPECIFICATIONS - GENERAL Case: Polycarbonate (panel and surface/pipe mount). NEMA 4X/CSA 4 (IP65) Dimensions Panel (code -10): 6.10 x 6.10 x 3.72 in. (155 x 155 x 94.5 mm) Surface/Pipe (code -11): 6.23 x 6.23 x 3.23 in. (158 x 158 x 82 mm); see page 15 for dimen-sions of pipe mounting bracket...

Page 4 - GENERAL SPECIFICATIONS FOR Xmt-A; steam sterilizable sensors); Temperature accuracy using RTD:; HAZARDOUS LOCATION APPROVALS; EEx ia IIC T4; MODEL Xmt-A AMPEROMETRIC TRANSMITTER; FEATURES; ATEX

Page 4 GENERAL SPECIFICATIONS FOR Xmt-A Input ranges: 0-330 nA, 0.3-4µA, 3.7-30 µA, 27-100 µA Repeatability (input): ±0.1% of range Linearity (input): ±0.3% of range Temperature range: -10 to 100°C (-10 to 150°C for steam sterilizable sensors) Temperature accuracy using RTD: ±0.5°C between 0 and 50°...

Page 6 - Preamplifier Location:; DIGITAL COMMUNICATIONS:

MODEL Xmt-P pH/ORP TRANSMITTER FEATURES The Model Xmt-P Transmitter with the appropriate sensor can be configured for either pH or ORP (oxidationreduction potential) measurement of aqueous solutions. Housed in a NEMA 4X case, the Model Xmt can be locat-ed close to the sensor even in the harshest env...