Emerson NGA2000 TO2- Manuals
Emerson NGA2000 TO2– User Manual in PDF format online.
Manuals:
User Manual Emerson NGA2000 TO2
Summary
The information contained in this document is subject to change without notice. Rosemount Analytical's system of NGA 2000 Modular Gas Analyzers and Controllers are patented, under U.S. Patent5.787.015. Manual Part Number 748410-AAugust 1999Printed in U.S.A. Rosemount Analytical Inc. 4125 East La Pal...
C ONTENTS 748410-AAugust 1999 i Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 P REFACE Purpose/Safety Summary .................................................................................P1 Glossary .........................................................................
C ONTENTS 748410-AAugust 1999 iii Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 S ECTION 4. M AINTENANCE AND T ROUBLESHOOTING 4.1 Overview ....................................................................................................27 4.2 Fuses ........................
C ONTENTS 748410-A August 1999 iv Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 F IGURES 1-1 CLD Analyzer Module Flow Diagram .......................................................... 31-2 CLD Analyzer Module - Top View ........................................................
P REFACE 748410-AAugust 1999 P1 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 P URPOSE /S AFETY S UMMARY The purpose of this manual is to provide the procedures for the installation, operation andmaintenance of the Platform and the System Accessories of the NGA 2000 System...
P REFACE 748410-A August 1999 P2 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 WARNING: TOXIC AND OXIDIZING GAS This module generates ozone which is toxic by inhalation and is a strong irritantto throat and lungs. Ozone is also a strong oxidizing agent. Its presence isdetec...
P REFACE 748410-AAugust 1999 P3 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 WARNING: INTERNAL ULTRAVIOLET LIGHT Ultraviolet light from the ozone generator can cause permanent eye damage.Do not look directly at the ultraviolet source. Use of ultraviolet filtering glassesi...
P REFACE 748410-A August 1999 P4 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 WARNING: POSSIBLE EXPLOSION HAZARD Ensure that all gas connections are made as labeled and are leak free.Improper gas connections could result in explosion or death. CAUTION: OVER-VOLTAGE SPIKING...
P REFACE 748410-AAugust 1999 P5 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 G LOSSARY Analyzer Module The module that contains all sensor/detector components for development of a PrimaryVariable signal; includes all signal conditioning and temperature control circuitry. ...
P REFACE 748410-A August 1999 P6 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 Power Supply Any of a variety of components that provides conditioned power to other NGA 2000components, from the Power Supply Board that plugs into the front of the Backplane in astand-alone ins...
P REFACE 748410-AAugust 1999 P7 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 S PECIFICATIONS - G ENERAL M EASUREMENT S PECIES : NO, NOx R ANGES : 0 to 10 ppm through 0 to 10,000 ppm NO, NOx R EPEATABILITY : ± 0.5% of fullscale (at constant temperature) M IN . D ETECTABLE ...
P REFACE 748410-A August 1999 P8 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 S PECIFICATIONS - S AMPLE T EMPERATURE : 0 ° C to 55 ° C (32 ° F to 131 ° F) T OTAL F LOW RATE : (Externally measured) 900 to 2200 cc/min. withbackpressure regulator pressure adjusted to 344 hPa(...
P REFACE 748410-AAugust 1999 P9 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 C USTOMER S ERVICE , T ECHNICAL A SSISTANCE AND F IELD S ERVICE For order administration, replacement Parts, application assistance, on-site or factory repair,service or maintenance contract info...
P REFACE 748410-A August 1999 P10 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 NAMUR 97-C219 N96 D OCUMENTATION The following Chemiluminescense Detector Analyzer Module instruction materials areavailable. Contact Customer Service or the local representative to order. 74841...
I NTRODUCTION 748410-AAugust 1999 1 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 1 1.1 O VERVIEW This manual describes the Chemiluminescence (CLD) Analyzer Module of RosemountAnalytical's NGA 2000 Series of gas analysis components (See Figures 1-1 and 1-2). The CLD Analyz...
I NTRODUCTION 748410-A August 1999 2 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 1 NO + O 3 → NO2* + O 2 NO 2 * → NO 2 + red light As NO and O3 mix in the reaction chamber, the intensity of the emitted red light ismeasured by a photodiode and is proportional to the concen...
I NTRODUCTION 748410-AAugust 1999 3 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 1 TEE 1/8T KYNAR VENT CAPILLARY 70cc/min 4 PSIG PUMP CAPILLARY 1/8 OD TEFLON VALVE C NC NO TEE 1/8TKYNAR TOP CONNECTOR,GLASS BLUE NOx TO NO CONVERTER CONNECTOR,GLASSBLUE CONNECTOR, MALE1/8T-3...
I NSTALLATION 748410-AAugust 1999 5 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 2 2.1 U NPACKING If the Chemiluminescence (CLD) Analyzer Module is received as a separate unit, carefullyexamine the shipping carton and contents for signs of damage. Immediately notify thesh...
I NSTALLATION 748410-A August 1999 6 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 2 F IGURE 2-1. A NALYZER M ODULE I NSTALLATION INTO I NSTRUMENT P LATFORM Unrestricted air flow to the rear of the Analyzer Module is critical to itsperformance and reliability. Operating amb...
I NSTALLATION 748410-A August 1999 8 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 2 2.4.3 S PECIFICATIONS Calibration Gases Either U.S.P. breathing grade air or nitrogen is recommended for use as zero gas. A mixtureof NO or NO 2 in a background of nitrogen is recommended a...
I NSTALLATION 748410-AAugust 1999 9 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 2 At no time should ozonator gas pressure exceed 2070 hPa-gauge (30 psig).Damage to internal components may occur if this pressure level is exceeded. Leak Test The CLD Analyzer Module is comp...
I NSTALLATION 748410-A August 1999 10 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 2 F IGURE 2-3. F RONT P ANEL C ONTROLS AND E LECTRICAL C ONNECTIONS Metering Valve Flow Balance Sample Regulator 250 V 24V 1 +2 -3 GND LON 1 LON 2 CLD T 6A Regulator AdjustSlot Network1 Conn...
S TARTUP AND O PERATION 748410-AAugust 1999 13 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 3 3.1 OVERVIEW Prior to initial startup, the user should leak test the module as outlined in Section 2. For the remainder of this section, Analyzer Module interconnection with a Pl...
S TARTUP AND O PERATION 748410-A August 1999 14 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 3 DISPLAY MESSAGE DESCRIPTION TYPE BAROMETER System Barometer WARNING BLOCK FET Heater Current WARNING BLOCK THRESH CLD Threshold Temperature WARNING BLOK TEMP Block Temperature WA...
S TARTUP AND O PERATION 748410-AAugust 1999 15 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 3 After the warm-up period (about one hour for the CLD Analyzer Module), all modules arecompletely functional. Establish that correct ozonator air pressure and sample flow rate are...
S TARTUP AND O PERATION 748410-AAugust 1999 17 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 3 F IGURE 3-6. T YPICAL H ELP S CREEN 3.5 C ALIBRATION Under the Expert Controls and Setup menu, use the ↓ arrow key to select Analyzer Module Setup, then select Calibration Gas Li...
S TARTUP AND O PERATION 748410-A August 1999 18 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 3 It is recommended, after initial startup, that the CLD Analyzer Module is calibrated at leastonce every eight hours. This practice should continue until experience indicates that...
S TARTUP AND O PERATION 748410-A August 1999 20 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 3 3.7 M EASUREMENT OF C ONVERTER E FFICIENCY It is the responsibility of the user to measure efficiency of the NO 2 -to-NO converter during initial startup and thereafter at interv...
S TARTUP AND O PERATION 748410-A August 1999 22 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 3 In the initial measurement, after lowering the temperature setpoint in Step 2, theefficiency will normally be less than 92%. 14. Reset converter temperature setpoint 20 ° C highe...
S TARTUP AND O PERATION 748410-AAugust 1999 23 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 3 F IGURE 3-7. M EASURING E FFICIENCY OF NO 2 TO NO C ONVERTER NO in AIR ADDED OZONATOR OFF a-b b a c-d c OZONATORON N 2 90 80 85 20 0 A. TYPICAL TEST SETUP B. TYPICAL TEST RESULTS...
S TARTUP AND O PERATION 748410-A August 1999 24 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 3 F IGURE 3-8. C ONVERSION E FFICIENCY AS A F UNCTION OF C ONVERTER T EMPERATURE 3.7.4 R EPLACEMENT OF C ONVERTER If the subnormal conversion efficiency is real, and not due to mea...
S TARTUP AND O PERATION 748410-AAugust 1999 25 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 3 3.8 R OUTINE O PERATION The CLD Analyzer Module is designed to continuously analyze the sample stream.Normally, it is never powered off except for servicing or for a prolonged sh...
M AINTENANCE AND T ROUBLESHOOTING 748410-AAugust 1999 27 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 4 WARNING: ELECTRICAL SHOCK HAZARD Disconnect power to the module(s) prior to replacing components. This equipment should not be adjusted or repaired by anyone except pro...
M AINTENANCE AND T ROUBLESHOOTING 748410-AAugust 1999 29 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 4 ASSEMBLED SIDE VIEW Converter Tube 655227 Wrap withaluminum foil Glass Cloth Sensor Heater Jacket 655228 Sensor 655282 4.3 F ANS Refer to Figure 4-1. To replace either ...
M AINTENANCE AND T ROUBLESHOOTING 748410-A August 1999 30 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 4 4.5 O ZONATOR Refer to Figure 4-1.To replace the ozonator, remove the two large straps and all tie-wraps,and disconnect the one electrical connection. Reassemble in rev...
M AINTENANCE AND T ROUBLESHOOTING 748410-AAugust 1999 31 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 4 Reaction Chamber Installation To reinstall, hold the housing in the inverted position while sliding the Reaction ChamberO-ring and window into position and the tubing i...
R EPLACEMENT P ARTS 748410-AAugust 1999 35 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 5 CAUTION: PARTS INTEGRITY Tampering with or unauthorized substitution of components may adverselyaffect safety of this product. Use only factory-approved components for repair. 5.1 RE...
R EPLACEMENT P ARTS 748410-A August 1999 36 Chemiluminescense Detector Analyzer ModuleRosemount Analytical NGA 2000 5 655250 CONVERTER REPLACEMENT PARTS NL655228 Heater Jacket NL655227 Tube, Aged NL655282 Temperature Sensor 659753, 659754 DETECTOR REPLACEMENT PARTS NL655235 Heater/Thermostat Assembl...
CLD I DENTIFICATION M ATRIX 748410-AAugust 1999 A1 Chemiluminescense Detector Analyzer Module Rosemount Analytical NGA 2000 A Each analyzer is configured per the customer sales order. Below is the CLD sales matrixwhich lists the various configurations available. To identify the configuration of an a...
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free fromdefects in workmanship and material under normal use and service for a period of twelve (12)months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquidjunctions, electrolyte...
Field service and repair facilities are located worldwide. U.S.A. To obtain field service on-site or assistance with a service problem, contact (24 hours, 7days a week): National Response Center 1-800-654-7768 I NTERNATIONAL Contact your local Rosemount Sales and Service office for service support. ...
I NSTRUCTIONS 015-748335-B Rosemount Analytical SAMPLE INLET FLOW RATE RESTRICTION FOR NGA 2000 CLD ANALYZER MODULE Rosemount Analytical Inc. 4125 East La Palma Avenue • Anaheim, California 92807-1802 • (714) 986-7600 • FAX (714) 577-8006 June 1997 015-748335-B Printed in U.S.A. CAUTION NOTICE EXTER...
I NSTRUCTIONS 015-748430-A Rosemount Analytical ADDENDUM CLD I NSTRUCTION M ANUAL 748410 Rosemount Analytical Inc. 4125 E AST L A P ALMA A VENUE • A NAHEIM , C ALIFORNIA 92807-1802 • (714) 986-7600 • FAX: (714) 577-8006 A UGUST 1999 • 015-748430-A • P RINTED IN USA This addendum serves as an amendme...
ADDENDUM 44 of 44 Rosemount Analytical August 1999 748430-A N OTES
Emerson Manuals
-
Emerson LD195EM8
User Manual
-
Emerson EWR20V4
User Manual
-
Emerson EWR20V4
Manual
-
Emerson SLC195EM8
User Manual
-
Emerson SLC195EM8
Manual
-
Emerson CR202EM9
User Manual
-
Emerson HD14070
User Manual
-
Emerson RLC320EM9
User Manual
-
Emerson LC320EM9
User Manual
-
Emerson LC320EM9
Manual
-
Emerson PD6922
User Manual
-
Emerson PD6922
Manual
-
Emerson EPT427D
User Manual
-
Emerson EPT427D
Manual
-
Emerson 400 VMC
User Manual
-
Emerson 400 VMC
Manual
-
Emerson CF960CK
User Manual
-
Emerson MICRO MOTION 2700
Installation Manual
-
Emerson CF787AP00
User Manual
-
Emerson CF787AP00
Manual