Emerson Process ManagementRosemount Analytical Inc.Process Analytic Division 1201 N. Main St.Orrville, OH 44667-0901T (330) 682-9010F (330) 684-4434e-mail: [email protected]://www.processanalytic.com ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, ...
HIGHLIGHTS OF CHANGES Effective May, 1998 Rev. 2.0 Page Summary — Modified existing manual to accommodate integrally mounted SPS 4000Single Probe Autocalibration Sequencer information. — Added locks for the electronic housing covers throughout the manual. — Changed test gas to calibration gas throug...
HIGHLIGHTS OF CHANGES (CONTINUED) Effective May, 1998 Rev. 2.0 (Continued) Page Summary Page 6-4 Updated Figure 6-3 to include last calibration failed alarm. Page 6-6 Added fault 14 to Table 6-1. Page 6-14 Added NOTE to determine when to replace the diffusion element. Page 7-1 Added general troubles...
HIGHLIGHTS OF CHANGES (CONTINUED) Effective Oct., 1998 Rev. 2.1 (Continued) Page Summary Page 9-2 Changed Figure and Index No. Page 9-3 Changed Figure and Index No. and Description. Added Abrasive ShieldAssemblies to parts list. Page 9-5 through 9-7 Changed Figure and Index No. and Part Numbers. Eff...
HIGHLIGHTS OF CHANGES (CONTINUED) Effective Nov., 1999 Rev. 2.3 Page Summary Pages P-11 thru P-15 Added new Quick Start Guide. Page 1-10 Added information on electronics operating temperatures and parts formounting. Page 1-14 Removed Table 1-4, renumbered subsequent tables in Section 1. Page 3-3 Cha...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management i Oxymitter 4000 TABLE OF CONTENTS PREFACE........................................................................................................................ P-1 Definitions ......
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management iii Oxymitter 4000 LIST OF ILLUSTRATIONS Figure 1-1. Typical System Package ....................................................................................... 1-2Figure 1-2. Oxy...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 iv Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 Figure 7-12. Fault 12, EEPROM Corrupt ................................................................................. 7-15Figure 7-13. Fault 13, Invalid Slope .......
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management P-1 Oxymitter 4000 PREFACE The purpose of this manual is to provide information concerning the components, func-tions, installation and maintenance of the Oxymitter 4000. Some sectio...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 P-2 Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 IMPORTANT SAFETY INSTRUCTIONS FOR THE WIRING AND INSTALLATION OF THIS APPARATUS The following safety instructions apply specifically to all EU member states. They s...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management P-3 Oxymitter 4000 CERAMIC FIBER PRODUCTS MATERIAL SAFETY DATA SHEET JULY 1, 1996 SECTION I. IDENTIFICATION PRODUCT NAME Ceramic Fiber Heaters, Molded Insulation Modules and Ceramic ...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 P-4 Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 SECTION II. PHYSICAL DATA APPEARANCE AND ODOR Cream to white colored fiber shapes. With or without optional white to gray granular surface coating and/or optionalbl...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management P-5 Oxymitter 4000 SECTION V. HEALTH HAZARD DATA THRESHOLD LIMIT VALUE (See Section III) EFFECTS OF OVER EXPOSURE EYE Avoid contact with eyes. Slightly to moderately irritating. Abra...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 P-6 Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 4) Pleural plaques (thickening along the chest wall) have been observed in a small number of employees who had along duration of employment. There are several occup...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management P-7 Oxymitter 4000 INCOMPATIBILITY/MATERIALS TO AVOID Incompatible with hydrofluoric acid and concentrated alkali. HAZARDOUS DECOMPOSITION PRODUCTS N.A. SECTION VII. SPILL OR LEAK PR...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 P-8 Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 SECTION IX. SPECIAL PRECAUTIONS PRECAUTIONS TO BE TAKEN IN HANDLING AND STORING General cleanliness should be followed.The Toxicology data indicate that ceramic fib...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 P-10 Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 WHAT YOU NEED TO KNOW BEFORE INSTALLING AND WIRING A ROSEMOUNT OXYMITTER 4000 OXYGEN TRANSMITTER 1. What type of installation does your system require? Use the fol...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management P-11 Oxymitter 4000 CAN YOU USE THE FOLLOWING QUICK START GUIDE? Use this Quick Start Guide if ... 1. Your system requires a STANDARD or INTEGRAL SPS 4000 OPTION installation. Instal...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 P-12 Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 QUICK START GUIDE FOR OXYMITTER 4000 SYSTEMS Before using the Quick Start Guide, please read “WHAT YOU NEED TO KNOW BEFOREINSTALLING AND WIRING A ROSEMOUNT OXYMITT...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management P-13 Oxymitter 4000 - - - - + + + + FACTORY WIRED 4-20 mACONNECTION TERMINALS 5 AND 6NOT USED LINEGROUND NEUTRAL CAL INITIATE5 VDC SELF POWEREDTO REMOTE CONTACTINPUT CONNECTION CALFA...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 P-14 Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 QUICK REFERENCE GUIDE OXYMITTER 4000 OXYGEN TRANSMITTER Performing a Manual Calibration 1. Place the control loop in manual. 2. Press the CAL key. The CAL LED will...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management P-15 Oxymitter 4000 HART COMMUNICATOR FAST KEY SEQUENCES Perform Calibration O 2 Upper Range Value 2 3 1 1 3 2 1 Trim Analog Output Analog Output Lower Range Value 2 4 3 2 2 Toggle A...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1 Oxymitter 4000 SECTION 1 DESCRIPTION AND SPECIFICATIONS 1-1 COMPONENT CHECKLIST OF TYPICALSYSTEM (PACKAGE CONTENTS) A typical Rosemount Oxymitter 4...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 2 7 1 26170008 o HART Communicator FISHER-ROSEMOUNT TM MAN 4275A00 English October 1994 6 5 3 4 1. Instruction Bulletin 2. IMPS 4000 ...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3 Oxymitter 4000 c. System Configuration Oxymitter 4000 units are available in fivelength options, giving the user the flexibilityto use an in situ p...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5 Oxymitter 4000 e. Handling the Oxymitter 4000 It is important that printed circuitboards and integrated circuits arehandled only when adequate anti...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 1-8 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 MANIFOLD INTERFACE BOARD CALIBRATION GAS 2 (LOW CALIBRATION GAS) SOLENOID CALIBRATION GAS 1 SOLENOID (HIGH CALIBRATION GAS) PRESSURE ...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-9 Oxymitter 4000 c. Operation The SPS 4000 works in conjunction with theOxymitter 4000’s CALIBRATION RECOM-MENDED feature to perform an autocalibra-t...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-13 Oxymitter 4000 Table 1-1. Product Matrix (Continued) Cont'd Code Language 1 English 2 German 3 French 4 Spanish 5 Italian Code Filtered Customer T...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 1-14 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 Table 1-3. Intelligent Multiprobe Test Gas Sequencer Versions Part Number Description Number of Oxymitter 4000 Units 3D39695G01 IMPS...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1 Oxymitter 4000 SECTION 2 INSTALLATION Before installing this equipment, readthe “Safety instructions for the wiringand installation of this apparatus” atthe front of...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 CO VER REMO V AL & A CCESS (305) (305) 12 12 ADD T O DIM “A” FOR PR OBE WITH CERAMIC DIFFUSER AND FLAME ARREST OR 5.14(131) WITH ST AND ARD SNUBBER...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3 Oxymitter 4000 1/4 IN. TUBE FITTING FOR INSTRUMENT AIR IN (OPTIONAL) TERMINAL COVER NOTE: DIMENSIONS ARE IN INCHES WITH MILLIMETERS IN PARENTHESES. 12.00 (304.80) NO...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 P S U E I T P I C R H W E N T H G C K E N I - E E R W A V I S O L P - X O M T A G NI N - R I T L A I VE - E E H DEFLECT OR ASSY 9 FT 12 FT (3112) (4026...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5 Oxymitter 4000 22.5 o B C 8 THREADED HOLES EQUALL Y S PACED ON D DIA B.C. ABRASIVE SHIELD FLANGE O.D. A A TABLE 5. ADAPTER PLA TE* DIMENSIONS FOR OXYMITTER 4000 TABL...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT ADAPTER PLATE NOT FURNISHED BY ROSEMOUNT. INSTALLATION FOR MASONRY WALL STACK CONSTRUCTION INSTALLATION ...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7 Oxymitter 4000 BRACE BARS (NOT BY ROSEMOUNT) 2.00 (51) NOTE: DIMENSIONS IN INCHES WITH MILLIMETERS IN PARETHESES. VERTICAL BRACE CLAMP ASSY. ABRASIVE SHIELD HORIZONT...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 4. If using the optional ceramic diffusion element, the vee deflector must be cor-rectly oriented. Before inserting theOxymitter 4000, check the direct...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9 Oxymitter 4000 6. If the system has an abrasive shield, check the dust seal gaskets. The jointsin the two gaskets must be staggered180°. Also, make sure the gaskets ...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 2-10 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 b. Connect Line Voltage Connect the line, or L1, wire to the L1 ter-minal and the neutral, or L2 wire, to the Nterminal (Figure 2-9). The Oxymitter 40...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11 Oxymitter 4000 2-3 ELECTRICAL INSTALLATION (FOR OXY-MITTER 4000 WITH SPS 4000) All wiring must conform to local and nationalcodes. Disconnect and lock out power bef...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 2-12 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 f. Connect Remote Contact Input Wiring To set up the SPS 4000 to initiate a calibra-tion from a remote location, route the 5 VDCcalibration initiate c...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 2-14 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 TO ELECTRONICS REF AIR SET 263C152G01 1 FLOWMETER 0.2-2.0 SCFH 771B635H02 2 2" PRESSURE GAGE 0-15 PSIG 275431-006 3 COMBINATION FILTER-REG. 0-30 P...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-15 Oxymitter 4000 Do not use 100% nitrogen as a low gas(zero gas). It is suggested that gas forthe low (zero) be between 0.4% and2.0% O2. Do not use gases with hy-droc...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Startup 3-1 Oxymitter 4000 SECTION 3 STARTUP Install all protective equipment coversand safety ground leads beforeequipment startup. Failure to installcovers and ground leads could r...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 3-2 Startup Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 c. Verify Oxymitter 4000 Configuration Located on the microprocessor board, thetop board, are two switches that configureoutputs for the Oxymitter 4000 (Fig...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 3-4 Startup Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 3-2 LOGIC I/O This two-terminal logic contact can be config-ured either as a solid-state relay-activated alarmor as a bi-directional calibration handshake s...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Startup 3-5 Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC INC DEC...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 3-6 Startup Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 3. CALIBRATION FAILED. One contact per probe from an SPS 4000 or IMPS4000 to the control room for notificationthat the calibration procedure failed.Grouped ...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Operation 4-1 Oxymitter 4000 SECTION 4 OPERATION 4-1 GENERAL a. Overview Ensure the Oxymitter 4000 is at normal op-eration. The diagnostic LEDs will display theoperating cycle. All o...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-1 Oxymitter 4000 SECTION 5 HART/AMS 5-1 OVERVIEW The HART Communicator is a handheld com-munications interface device. It provides acommon communications link to all micro...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 5-4 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 Table 5-1. Logic I/O Configuration Mode Configuration 0 The unit is not configured for any alarm condition. 1 The unit is configured for a Unit Alarm. 2 Th...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 5-8 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 5-7 HART COMMUNICATOR MANUALO 2 CAL METHOD To perform a calibration using the HART Com-municator, use the following procedure. If nec-essary, use the menu ...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-1 Oxymitter 4000 SECTION 6 MAINTENANCE AND SERVICE 6-1 OVERVIEW This section identifies the calibration methodsavailable and provides the procedures to main...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 6-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 DIA GN OS TIC ALAR MS TE ST PO IN TS HE ATE R T /C HE ATER 02 C ELL CA LIB RAT IO N CA LIB RATIO N R EC OM ME ND ED 02 C ELL m V + 02 C ELL ...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 6-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 SW2 DIP SWITCH HART LOCAL 0-10% 0-25% 4mA 20mA NOT USED NOT USED LED HEATER O2 CELL CALIBRATION HEATER T/C STATUS OPEN SHORTED REVERSED HIGH...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-5 Oxymitter 4000 readings using the first cali-bration gas and the CAL LEDwill flash continuously. Theflashing indicates the Oxy-mitter 4000 is ready to tak...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 6-6 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 Table 6-1. Diagnostic/Unit Alarms LED Flashes Status Fault 1 OPEN 1 2 SHORTED 2 3 REVERSED 3 HEATER T/C 4 A/D COMM ERROR 4 1 OPEN 5 2 HIGH H...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-7 Oxymitter 4000 It is recommended that the Oxymitter4000 be removed from the stack for allservice activities. The unit should beallowed to cool and be take...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-9 Oxymitter 4000 500 VA R SERIAL NO.TAG NO. OXYMITTER 4000 VOLTS: WATTS: OUTPUT: LINE FUSE: Rosemount Analytical Inc.Orrville, OH 44667-0901 85-264 VAC 48-6...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 6-10 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 10. Holding the J1 connector leads, slide the electronic assembly the rest of theway into the housing. Align the elec-tronic assembly so th...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 6-12 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 6-7 HEATER STRUT REPLACEMENT This paragraph covers heater strut replacement.Do not attempt to replace the heater strut untilall other possi...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-13 Oxymitter 4000 f. When replacing the strut, align the slot onthe heater plate with the calibration gas linein the probe tube. Slide the strut into thepro...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-15 Oxymitter 4000 OPTIONAL CERAMIC DIFFUSION ELEMENT VEE DEFLECTOR HUB SETSCREW RETAINER CEMENT PORT CEMENT FILLET 22220029 Figure 6-10. Ceramic Diffusion E...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 6-16 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 Do not get cement on ceramic diffu-sion element except where it touchesthe hub. Any cement on ceramic diffu-sion element blocks airflow thr...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-19 Oxymitter 4000 - - + + + CAL INITIATE BLACK CAL FAIL LINE IN TO PRESSURE SWITCH IN CAL NEUTRALGROUND N RELAY OUTPUTS 26170004 REMOTE CONTACT INPUT LINE V...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 6-20 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 7. Tag and unplug solenoid (13 or 20) leads from power supply board (18).Refer to Figure 6-12. Calibration gas 1solenoid wires connect to c...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-21 Oxymitter 4000 f. Pressure Regulator (Optional) Maintenance 1. Pressure Adjustments. Reference air pressure regulator (8, Figure 6-13) isfactory set to 2...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 6-22 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 6. Remove screws (6) and bracket (5) se- curing flowmeter (16 or 17) to flowme-ter bracket (25). 7. Remove flowmeter (16 or 17), with in- s...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-23 Oxymitter 4000 26170012 4 NOTE: A STANDARD SPS 4000 IS EQUIPPED WITH TEFLON TUBING AND BRASS FITTINGS. OPTIONAL STAINLESS STEEL TUBING AND FITTINGS ARE A...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-1 Oxymitter 4000 SECTION 7 TROUBLESHOOTING Install all protective equipment coversand safety ground leads after trouble-shooting. Failure to install covers andgroun...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 The logic contact is self-powered, +5 VDC,340 ohm series resistance. An interposingrelay will be required if this contact is to beutilized to annunc...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-4 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC ...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-5 Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-6 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC ...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-8 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC ...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-9 Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-10 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-11 Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - IN...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-12 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-13 Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - IN...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-14 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-15 Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - IN...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-16 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-17 Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - IN...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-18 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 DIAGNOSTIC ALARMS TEST POINTS HEATER T/C HEATER O2 CELL CALIBRATION CALIBRATION RECOMMENDED O2 CELL mV + O2 CELL mv - HEATER T/C + HEATER T/C - INC...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-19 Oxymitter 4000 Install all protective equipment coversand safety ground leads after trouble-shooting. Failure to replace coversand ground leads could result in s...
Instruction Manual IB-106-340 Rev. 2.4April, 2001 7-20 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Oxymitter 4000 Table 7-2. SPS 4000 Fault Finding Symptom Check Fault Remedy No calibrationgas flow Wiring Improper wire connections,loose connections, ordamaged w...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-21 Oxymitter 4000 CHECK ALL WIRING BETWEEN OXYMITTER 4000 AND SPS 4000. CHECK FUSE F1 ON POWER SUPPLY BOARD. CHECK LOGIC I/O SETTING VIA HART/AMS. DISCONNECT CAL GA...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 8-1 Oxymitter 4000 SECTION 8 RETURN OF MATERIAL 8-1 If factory repair of defective equipment is re-quired, proceed as follows: a. Secure a return authorization num...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 9-1 Oxymitter 4000 SECTION 9 REPLACEMENT PARTS Table 9-1. Replacement Parts for Probe Figure and Index No. Part Number Description No Dust Seal Dust Seal 6-1, 1 thr...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 9-5 Oxymitter 4000 Table 9-2. Replacement Parts for Electronics Figure and Index No. Part Number Description 4850B10G01 Electronics English-Standard 6-1, 21, 27 484...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Optional Accessories 10-1 Oxymitter 4000 SECTION 10 OPTIONAL ACCESSORIES HART HANDHELD 275 COMMUNICATOR The HART Handheld 275 Communicator is an interface devicethat provides a commo...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Optional Accessories 10-3 Oxymitter 4000 O 2 CALIBRATION GAS SEQUENCER Rosemount Analytical’s O 2 Calibration Gas and Service Kits have been carefully designed to provide a more conv...
Instruction Manual IB-106-340 Rev. 2.4 April, 2001 Rosemount Analytical Inc. A Division of Emerson Process Management Optional Accessories 11-1 Oxymitter 4000 SECTION 11 INDEX This index is an alphabetized listing of parts, terms, and procedures having to do with the Haz-ardous Area Oxygen/Combustib...
3536 3583/4-01 WARRANTY Rosemount warrants that the equipment manufactured and sold by it will, upon shipment, be freeof defects in workmanship or material. Should any failure to conform to this warranty become ap-parent during a period of one year after the date of shipment, Rosemount shall, upon p...
Emerson Manuals
-
Emerson LD195EM8
User Manual
-
Emerson EWR20V4
User Manual
-
Emerson EWR20V4
Manual
-
Emerson SLC195EM8
User Manual
-
Emerson SLC195EM8
Manual
-
Emerson CR202EM9
User Manual
-
Emerson HD14070
User Manual
-
Emerson RLC320EM9
User Manual
-
Emerson LC320EM9
User Manual
-
Emerson LC320EM9
Manual
-
Emerson PD6922
User Manual
-
Emerson PD6922
Manual
-
Emerson EPT427D
User Manual
-
Emerson EPT427D
Manual
-
Emerson 400 VMC
User Manual
-
Emerson 400 VMC
Manual
-
Emerson CF960CK
User Manual
-
Emerson MICRO MOTION 2700
Installation Manual
-
Emerson CF787AP00
User Manual
-
Emerson CF787AP00
Manual