Emerson HASX2E-IM-HS - Manuals
Emerson HASX2E-IM-HS – Manual in PDF format online.
Manuals:
Manual Emerson HASX2E-IM-HS
Summary
ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, a...
Emerson Process Management GmbH & Co. OHG TOC-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 TO C Table of contents SHORT FORM GUIDE FOR THIS MANUAL To find information about see chapter Safety instructions ..................................................... S The different instruments d...
Emerson Process Management GmbH & Co. OHG TOC-2 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 T able of C onTenTs Introduction S-1 Definitions S-1 Terms Used in This Instruction Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-2 Symbols Used on and...
Emerson Process Management GmbH & Co. OHG TOC-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 TO C Table of contents Table of Contents 2.2.1 X-STREAM X2GK: ½19 Inch T able -Top Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5 2.2.2 X-STREAM X2GP: 19 Inch T able -Top ...
Emerson Process Management GmbH & Co. OHG TOC-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Table of Contents 5.2 Front Panel Elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-25.2.1 Display. . . . . . . . . . . . . . . . . . ...
Emerson Process Management GmbH & Co. OHG TOC-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 TO C Table of contents Table of Contents 7.4.6 Resetting a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-39 7.4.8 Cancelling an O ngoi...
Emerson Process Management GmbH & Co. OHG TOC-6 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Table of Contents Chapter 11 Dismounting and Disposal 11-1 11.1 Dismounting and Diposal of the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1 Appendix A-1 A.1 ...
Emerson Process Management GmbH & Co. OHG TOC-8 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Index of Figures Fig. 3.8 Cover for E O2 Sensor Block At Rear Panel .........................................................3-10 Fig. 3.9 Trace Oxygen Sensor Design Principle .......................
Emerson Process Management GmbH & Co. OHG TOC-9 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 TO C Table of contents Index of Figures Fig. 4-32: Loads in Series .....................................................................................................4-38Fig. 4-33: Loads in Paral...
Emerson Process Management GmbH & Co. OHG TOC-10 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 i ndex of T ables Tab. 3-1: Paramagnetic Sensor - Cross Interferences (examples) . . . . . . . . . . . . . . . . . . . . . 3-6 Tab. 3-2: Solvent Resistant Paramagnetic Sensor - Approved Solvents ...
Emerson Process Management GmbH & Co. OHG S-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 S Safety Instructions The instruction manual contains information about the component assembly, function, installation, operation and maintenance of the X-STREAM ® X2 series gas analyzers. The manual...
Emerson Process Management GmbH & Co. OHG S-2 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 TERMS USED IN THIS INSTRUCTION MANUAL Lower Explosion Limit (LEL) Volume ratio of flammable gas in air below which an explosive gas atmosphere will not be formed: the mixture of gas and air lacks su...
Emerson Process Management GmbH & Co. OHG S-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 S Safety Instructions SYMBOLS USED ON AND INSIDE THE UNIT This symbol at the instrument ... ... means dangerous voltages may be accessible. Remo - ving covers is permitted only, if the instrument is ...
Emerson Process Management GmbH & Co. OHG S-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 This symbol used in the manual ... ... means dangerous voltages may be exposed hot surfaces may be exposed possible danger of explosion toxic substances may be present substances harmful to health m...
Emerson Process Management GmbH & Co. OHG S-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 S Safety Instructions INTENDED USE STATEMENT X-STREAM series gas analyzers are intended to be used as analyzers for industrial pur-poses. They must not be used in medical, diagnostic or life support ...
Emerson Process Management GmbH & Co. OHG S-6 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 AUTHORIZED PERSONNEL In-depth specialist knowledge is an absolutely necessary condition for working with and on the analyzer. Authorized personnel for installing, operating, servicing and maintainin...
Emerson Process Management GmbH & Co. OHG S-7 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 S Safety Instructions INSTALLING AND CONNECTING THE UNIT The following notices should be carefollowed to ensure compliance with the low voltage directive . 1. Suitable grounding connections should be...
Emerson Process Management GmbH & Co. OHG S-9 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 S Safety Instructions TOXIC GASES This unit’s exhaust may contain toxic gases such as (but not limited to) e.g. sulfur dioxide. These gases can cause serious injuries. Avoid inhaling exhaust gases. C...
Emerson Process Management GmbH & Co. OHG S-12 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 • The unit must be installed in a clean and dry area protected from strong vibrations and frost. • The unit must not be exposed to direct sunlight and sources of heat. Admissable ambient temperatur...
Emerson Process Management GmbH & Co. OHG 1-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 Chapter 1 Technical description The following are the main features of the new Emerson Process Management X-STREAM gas analyzers in brief : • compact design with easily accessible internal components...
Emerson Process Management GmbH & Co. OHG 1-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1.1 Overview All X-STREAM gas analyzers feature an easy-to-use alphanumeric user interface, which displays measurement values, status and error messages, and menus for the input of parameters. For ea...
Emerson Process Management GmbH & Co. OHG 1-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 1.2 Configuration of Gas Lines 1.2 Configuration of Gas Lines Various materials are available to allow the analyzer to be customized to your needs. The materials used are selected based on the chara...
Emerson Process Management GmbH & Co. OHG 1-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1.2 Configuration of Gas Lines 1.2.6 Optional Components for Gas Lines The analyzers can, as an option, be fitted with further components. Not all components are available for all analyzer types:• in...
Emerson Process Management GmbH & Co. OHG 1-6 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 1.2 Configuration of Gas Lines 1.2.6.6 Internal Temperature Sensors The influence of varying temperatures can be compensated for by the use of internal temperature sensors ( measurement specificatio...
Emerson Process Management GmbH & Co. OHG 1-7 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1.2.6.7 Optional Heated Area The physical components can be optionally separated from the electrical components by means of a special box (not an option for ½ 19 in units). This can be done for one o...
Emerson Process Management GmbH & Co. OHG 1-8 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 1.2 Configuration of Gas Lines Depending on the application and the selected analyzer options, alternative gas line configu - rations are available, exemplified in the follo - wing diagram of a dual...
Emerson Process Management GmbH & Co. OHG 1-9 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1.3 Interfaces 1.3.1 Analog Outputs Each X-STREAM analyzer is fitted with one output per channel as standard, which can transmit data on concentration levels to an external data acquisition system.T ...
Emerson Process Management GmbH & Co. OHG 1-10 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 1.3.3.1 Modbus Interface, Serial A serial interface with the Modbus protocol allows communication with external data ac- quisition systems. The interface enables the exchange and modification of me...
Emerson Process Management GmbH & Co. OHG 1-11 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1.3 Interfaces 1.3.3.3 Digital Outputs Digital outputs can be used for various pur - poses: • Issuing concentration alarms: Process control systems can detect when limits are exceeded and trigger ap...
Emerson Process Management GmbH & Co. OHG 1-12 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 1.4 Overview of Analyzer Types 1.4 Comparison of the Various X-STREAM Analyzer Models 1 ⁄ 2 19 in housing, table-top or rackmountable, optional with carrying handle protection type: IP 20 1 ⁄ 1 19 ...
Emerson Process Management GmbH & Co. OHG 1-13 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1 ⁄ 2 19 in housing, table-top or rackmountable, optional with carrying handle protection type: IP 20 1 ⁄ 1 19 in housing, table-top or rackmountable, protection type: IP 20 Stainless steel wallmoun...
Emerson Process Management GmbH & Co. OHG 1-16 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 1.5 X-STREAM X2GK Fig. 1-5: X-STREAM X2GK - Views 1 4x20 character alphanumeric display 2 LED (red) 3 LED (red) 4 LED (green) 5 “Measure” key 6 “Enter” key 7 4 keys for adjustment and menu selectio...
Emerson Process Management GmbH & Co. OHG 1-19 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 Fig. 1-6: X-STREAM X2GP - Views 1.6 X-STREAM X2GP 1 4x20 character alphanumeric display 2 LED (red) 3 LED (red) 4 LED (green) 5 “Measure” key 6 “Enter” key 7 4 keys for adjustment and menu selection...
Emerson Process Management GmbH & Co. OHG 1-21 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 The service interface is elec-trically connected to the unit’s electronic components. If it is incorrectly handled, damage to the unit may result. The service Interface may only be used by EMERSON s...
Emerson Process Management GmbH & Co. OHG 1-22 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Fig. 1-7: X-STREAM X2XF Field Housings - Front Views 1.7 X-STREAM X2XF Field Housings HEAVY INSTRUMENT X-STREAM field housings, intended for outside and wall mounted use, weigh approx. (XLF) 26 kg ...
Emerson Process Management GmbH & Co. OHG 1-23 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1.7 X-STREAM X2XF Field Housings Fig. 1-8: X-STREAM X2XF Field Housings - Front Panel 1 4x20 character alphanumeric display 2 LED (red) 3 LED (red) 4 LED (green) 5 “Measure” key 6 “Enter” key 7 4 ke...
Emerson Process Management GmbH & Co. OHG 1-24 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 1.7 X-STREAM X2XF Field Housings Fig. 1-9: X-STREAM XLF - Bottom and Side View 1 Cable gland for power cable 2 Cable glands for signal cables 3 4 brackets for wall-mounting 4 Gas in- & outlets ...
Emerson Process Management GmbH & Co. OHG 1-25 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1.7 X-STREAM X2XF Field Housings Fig. 1-10: X-STREAM XLF - Power Supply and Signal Terminals 1 Screw-type terminals for signal cables 2 Power line filter 3 Cable glands 4 Power supply terminals with...
Emerson Process Management GmbH & Co. OHG 1-26 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 1.7.1 Field Housing for Installation in hazardous Areas (Ex-Zones) Special X-STREAM field housing analyzer mo - dels can be used in Ex-zones 2 or Division 2: X-STREAM XLFN/XXFN: A n a l y z e r w i...
Emerson Process Management GmbH & Co. OHG 1-27 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1.8 X-STREAM X2FD 1.8 X-STREAM X2FD: Cast Aluminum Flameproof Housing The most obvious X-STREAM 2FD analyzer feature is its flameproof housing (Fig. 1-10). This enables its use in Ex-zone 1 hazardou...
Emerson Process Management GmbH & Co. OHG 1-31 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 1.8 X-STREAM X2FD 1 Terminals for signal cables (shown fully populated) 2 Power line filter 3 Cable inlets for power and signal cables Fig. 1-14: X-STREAM X2FD - Terminals (shown with front panel re...
Emerson Process Management GmbH & Co. OHG 2-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 2 T echnical data Chapter 2 Technical data This chapter contains all the technical details of the analyzers, divided into common and model-specific data. Common technical data page 2-2 X-STREAM X2GK ...
Emerson Process Management GmbH & Co. OHG 2-2 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 2.1 Common Technical Data 2.1 Common Technical Data Installation site Humidity (non-condensing) < 90 % RH at +20 °C (68 °F) < 70 % RH at +40 °C (104 °F) Degree of pollution 2 Installation cate...
Emerson Process Management GmbH & Co. OHG 2-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 2 T echnical data Interfaces, signal inputs and outputs Interface signals are accessed in different ways depending on the analyzer model. (This does not apply to Ethernet, which always uses RJ45): X-...
Emerson Process Management GmbH & Co. OHG 2-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 2.1 Common Technical Data Optional interfaces for all models 1 Modbus interface RS 485 (2 or 4 wire) optional: RS 232 Ethernet (RJ45 socket) Digital inputs and outputs 7 or 14 digital inputs max. 30...
Emerson Process Management GmbH & Co. OHG 2-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 2 T echnical data 2.2.1 X-STREAM X2GK: ½19 Inch Table-Top Unit 2.2.1 Model-S pecific Technical Data: X-STREAM X2GK Fig. 2-1: X-STREAM X2GK - Dimensions 2.2 Model-Specific Technical Data All dimension...
Emerson Process Management GmbH & Co. OHG 2-6 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 2.2.1 Model-S pecific Technical Data: X-STREAM X2GK Front panel detail 6 screws (M 2.5) for rack mounting or affixing frame & handle Frame and handle detail Note!All shown variable options are i...
Emerson Process Management GmbH & Co. OHG 2-7 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 2 T echnical data 2.2.1 Model-S pecific Technical Data: X-STREAM X2GK Housing admissible operational ambient temperature: 0 °C to +50 °C (32 to 122 °F) admissible ambient temperature for storage: -20...
Emerson Process Management GmbH & Co. OHG 2-8 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 2.2.1 Model-S pecific Technical Data: X-STREAM X2GK 2.2.1.1 Data for Optional External Power Supply Units Model UPS 01 T This PSU can be ordered as an option for supplying power to one table-top uni...
Emerson Process Management GmbH & Co. OHG 2-9 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 2 T echnical data 2.2.1 Model-S pecific Technical Data: X-STREAM X2GK 275.3 105.9 54.2 57.7 Fig. 2-3: UPS 01 T Power Supply Unit IEC power input socket PE Dimensions (approx. values in [mm]) Pin 1: M...
Emerson Process Management GmbH & Co. OHG 2-10 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 This PSU can be optionally ordered for powering 2 table-top units with a single common power supply. Model: 5 A or 10 A table-top PSU Recommended clearance for cooling 15 mm (0.6 in) in front and b...
Emerson Process Management GmbH & Co. OHG 2-11 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 2 T echnical data Power input fuses Data for fuse link: AC 230 V / T 6.3 A / 5x20 mm Nominal output voltage 24 V (+ 5 %, -1 %) Power consumption max. 5 A or 10 A Output two 3-pin XLR sockets Efficie...
Emerson Process Management GmbH & Co. OHG 2-12 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 2.2.2 Model-S pecific Technical Data: X-STREAM X2GP 2.2.2 X-STREAM X2GP: 19 Inch Table-Top and Rack-Mount Models Fig. 2-5: X-STREAM X2GP - Dimensions approx. values in mm [ins] Strain relief bracke...
Emerson Process Management GmbH & Co. OHG 2-14 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Fig. 2-6: X-STREAM X2GP - Power Supply and Signal Connections Signal inputs and outputs As standard, signal cables are connected using Submin-D plugs or sockets on the unit’s rear panel (except Eth...
Emerson Process Management GmbH & Co. OHG 2-15 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 2 T echnical data 2.2.3 Model-S pecific Technical Data: X-STREAM X2XF Field Housings 2.2.3 X-STREAM X2XF: Single (XLF) or Dual (XXF) Compartment Field Housing Fig. 2-8: X-STREAM XLF - Dimensions All...
Emerson Process Management GmbH & Co. OHG 2-17 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 2 T echnical data Power supply Nominal input voltage 100–240 V 50/60 Hz, wide range power supply unit Input voltage range 85–264 V , 47–63 Hz Nominal input current XLF standard 1.3–0.7 A max. with t...
Emerson Process Management GmbH & Co. OHG 2-18 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Fig. 2-10: X-STREAM X2XF Field Housings - Power Supply Terminals / Fuse Holders Signal inputs and outputs All signal cables are connected to internal screw-type terminals located in the front area ...
Emerson Process Management GmbH & Co. OHG 2-19 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 2 T echnical data 2.2.4 X-STREAM X2FD: Flameproof Housing Housing Permissible operational ambient temperature -20 °C to +50 °C (-4 °F to 122 °F) Permissible ambient temperature for storage -20 °C to...
Emerson Process Management GmbH & Co. OHG 2-20 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Fig. 2-13: X-STREAM X2FD - Power Supply Terminals / Fuse Holders 3 2 4 1 Power supply Nominal input voltage 100–240 V 50/60 Hz, global power supply unit Input voltage range 85–264 V , 47–63 Hz Nomi...
Emerson Process Management GmbH & Co. OHG 2-22 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 2.3 Information on Name Plate 2.3 Information on Name Plate Fig. 2-15: Analyzer Name Plate (examples) The name plate provides details on the confi - guration of the unit, installed measuring tech-n...
Emerson Process Management GmbH & Co. OHG 3-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles Chapter 3 Measuring Principles 3.1 Infrared Measurement (IR) Ultraviolet Measurement (UV) 3.1.1 IntrinzX Technology X-STREAM series analyzers support several measuring principl...
Emerson Process Management GmbH & Co. OHG 3-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles 3.1 Infrared (IR) and Ultraviolet (UV) Measurement 3.1.2 NDIR Detector Fig. 3-2: Gas detector design principle Compensation chamber Fill nozzle Absorption chamber Connecting ch...
Emerson Process Management GmbH & Co. OHG 3-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 3.1 Infrared (IR) and Ultraviolet (UV) Measurement Fig. 3-3: Photometer assembly principle 3.1.3 Technical Implementation 1 UV source2 Adaptor cell3 Analysis cell (internal view)4 Filter cell5 UV de...
Emerson Process Management GmbH & Co. OHG 3-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles 3.2 Oxygen Measurement 3.2 Oxygen Measurement Two different principles are used for mea-suring oxygen concentrations. The currently used principle is given by the channel code ...
Emerson Process Management GmbH & Co. OHG 3-6 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 3.2 Oxygen Measurement Cross Interferences By Accompanying Gases The Table below by selected gases shows, how accompanying gases interfere the pa-ramagnetic oxygen measurement. A more comprehensive ...
Emerson Process Management GmbH & Co. OHG 3-7 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles 3.2 Oxygen Measurement Tab. 3-2: Solvent resistant paramagnetic sensor - approved solvents Tab. 3-3: Solvent resistant paramagnetic sensor - medium affected materials Applicati...
Emerson Process Management GmbH & Co. OHG 3-8 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 3.2 Oxygen Measurement 3.2.2 Electrochemical Measurement This sensor utilizes the principle of galvanic cells, fig. 3-5 shows the design. For storage and handling instructions, and safety data Chapt...
Emerson Process Management GmbH & Co. OHG 3-9 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles 3.2 Oxygen Measurement Fig. 3-7: Electrochemical reaction of Oxygen Sensor Resistance (6) Thermistor (5) (11) (red) (black) (+) Lead anode (1) (-) Gold cathode (2) Electrochemi...
Emerson Process Management GmbH & Co. OHG 3-10 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 3.2 Oxygen Measurement 3.2.2.1 Special Hints Due to the measuring principle the electro-chemical oxygen cell requires a minimum internal consumption of oxygen (residual humidity avoids drying of th...
Emerson Process Management GmbH & Co. OHG 3-11 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles 3.2.3.1 Special Hints For trace oxygen measurements (tO 2 )another electrochemical sensor technology is used, see Fig. 3-9. The sensor is a self contained disposable unit whic...
Emerson Process Management GmbH & Co. OHG 3-12 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 performance or damage to the sensor. Do not remove the sealing caps until all associated sample handling components are installed and the instrument is fully ready for installa-tion.After replaceme...
Emerson Process Management GmbH & Co. OHG 3-13 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles 3.3 Thermal Conductivity Measurement 3.3 Thermal Conductivity Measurement Thermal conductivity is the property of a ma- terial that indicates its ability to conduct heat. Ther...
Emerson Process Management GmbH & Co. OHG 3-14 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 3.3.2 Technical Implementation A block made either of aluminum, stainless steel or hastelloy contains two gas paths. Both, the volume of the block and the mass of the sensors have been minimized in...
Emerson Process Management GmbH & Co. OHG 3-15 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles 3.4 Trace Moisture Measurement Fig. 3-14: Trace Moisture Sensor Assembly 1 Sensor 2 Sensor block 3 Gas connections 3.4 Trace Moisture Measurement 1 2 3 T h e Tr a c e M o i s ...
Emerson Process Management GmbH & Co. OHG 3-16 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 3.4 Trace Moisture Measurement 3.4.1 Special Operating Conditions The sensor is completely calibrated with all calibration data stored in its flash memory and does not require recalibration:• If th...
Emerson Process Management GmbH & Co. OHG 3-17 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles 3.4 Trace Moisture Measurement 3.4.2 Accompanying Gases Component Maximum Permitted Concentration / ppm Maximum Permitted Dew- point / °C Acetylene (Ethyne) C 2 H 2 1 -20 Ammo...
Emerson Process Management GmbH & Co. OHG 3-19 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 3 Measuring Principles 3.5 Measurement Specifications 3.5 Measurement S pecifications Tab. 3-8: Gas Components and Measuring Ranges, Examples Sample gas components and measuring ranges (standard con...
Emerson Process Management GmbH & Co. OHG 3-20 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 3.5 Measurement Specifications Tab. 3-9: IR, UV, VIS, TCD - Measurement Performance Specifications The specifications given in these tables al - ways apply to the physical measuring ranges, as list...
Emerson Process Management GmbH & Co. OHG 4-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation Chapter 4 Installation This chapter describes the correct installation procedure for the various X-STREAM analyzer versions. On receipt, check the packaging and its contents thoroughly...
Emerson Process Management GmbH & Co. OHG 4-2 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 4.2 Installation - Introduction In order to comply with regulations on elec- tromagnetic compatibility, it is recommended to use only shielded cables which can be supplied by Emerson Process Managem...
Emerson Process Management GmbH & Co. OHG 4-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.3 Installation - Gas Conditioning 4.3 Gas Conditioning If moisture cannot be avoided, it is necessary to ensure that the dew point of the gases is at least 10 °C (18 °F) below the am...
Emerson Process Management GmbH & Co. OHG 4-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Case purge option The purge medium (e.g. to minimize CO 2 interference or for enhanced safety when measuring corrosive or poisonous gases) • must be dry, clean and free of corro-sives or components ...
Emerson Process Management GmbH & Co. OHG 4-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.4 Installation - Gas Connections TOXIC GAS HAZARDS Take care that all external gas pipes are connected in the described way and that they are gastight to avoid leakages! Faulty conne...
Emerson Process Management GmbH & Co. OHG 4-7 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.5 Electrical Connections 4.5 Installation - Electrical Connections ELECTRICAL SHOCK HAZARD Only qualified personnel, observing all applicable technical and legal requirements, may in...
Emerson Process Management GmbH & Co. OHG 4-8 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 4.6 Analyzer S pecific Instructions for Installation 4.6 Analyzer S pecific Instructions for Installation Important note for X-STREAM X2FD! Due to the special conditions which must be observed when ...
Emerson Process Management GmbH & Co. OHG 4-9 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation Plugs and sockets required for the electrical connections are on the rear panel of the unit (Fig. 4-5).X-STREAM X2GK analyzers are powered by an internal wide range power supply for wo...
Emerson Process Management GmbH & Co. OHG 4-10 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 The number and configuration of the gas inlets and outlets vary from model to model and are indicated on the notice on the rear of the instrument. To simplify installation, we recommend labelling t...
Emerson Process Management GmbH & Co. OHG 4-11 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.6.1 Installation - X-STREAM X2GK Signal inputs and outputs The number of signal outputs actually available varies according to the unit’s configuration. Analog signals Relay outputs...
Emerson Process Management GmbH & Co. OHG 4-12 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Fig. 4-7: Plug X2 - Modbus interface Modbus interface For specifications and notes on control, see chapter 7 4.6.1 Installation - X-STREAM X2GK Pin 1 Pin 8 Ethernet connector Notes!Consider the ins...
Emerson Process Management GmbH & Co. OHG 4-13 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.6.1 Installation - X-STREAM X2GK Fig. 4-8: Socket X4 - Pin configuration Digital inputs Type: 9 outputs, dry relay change-over ontacts, can be used as normally open (NO) or normally...
Emerson Process Management GmbH & Co. OHG 4-14 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 4.6.1 Installation - X-STREAM X2GK Power supply 24 VDC is supplied to the unit by means of a three-pin XLR connector on the rear panel of the instrument.Depending on the order, the following is sup...
Emerson Process Management GmbH & Co. OHG 4-16 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 The number and configuration of the gas inlets and outlets vary from model to model and are indicated on the notice on the rear of the instrument. To simplify installation, we recommend labelling t...
Emerson Process Management GmbH & Co. OHG 4-20 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Pin Signal P2.1 Channel 1, (+) 4 (0)–20 mA P2.2 Channel 1, GND P2.3 Channel 2, (+) 4 (0)–20 mA P2.4 Channel 2, GND P2.5 Channel 3, (+) 4 (0)–20 mA P2.6 Channel 3, GND P2.7 Channel 4, (+) 4 (0)–20 m...
Emerson Process Management GmbH & Co. OHG 4-22 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Pin Signal P3.1 Input 1 P3.2 Input 2 P3.3 Input 3 P3.4 Input 4 P3.5 Input 5 P3.6 Input 6 P3.7 Input 7 P3.8 GND for inputs 1-7 P3.9 unused P3.10 Output 5, NC P3.11 Output 5, NO P3.12 Output 5, COM P...
Emerson Process Management GmbH & Co. OHG 4-23 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.6.3 Installation - X-STREAM X2XF Field Housings 4.6.3 X-STREAM X2XF Field Housings (Single X LF; Dual XXF) Fitted with four eyebolts and featuring IP66 / Type 4X protection, the X-S...
Emerson Process Management GmbH & Co. OHG 4-25 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation Fig. 4-20: X-STREAM X2XF Field Housings - Arrangement of Terminals, Cable Glands and Gas Fittings 4.6.3 Installation - X-STREAM X2XF Field Housings Power and signal cables are connect...
Emerson Process Management GmbH & Co. OHG 4-26 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 All signal cables are connected to screw-type terminals located inside the housing. Access to the internal components is gained by relea - Preparation of signal cables Recommended wire gauge: 0.14…...
Emerson Process Management GmbH & Co. OHG 4-28 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 4.6.3 Installation - X-STREAM X2XF Field Housings Fig. 4-21: Terminal block X1 - Analog signals and relay outputs 1-4 Note!Consider the installation notes in section 4.5 and the notes on installing...
Emerson Process Management GmbH & Co. OHG 4-29 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.6.3 Installation - X-STREAM X2XF Field Housings Modbus interface Specification and interface control: Chapter 9 Fig. 4-22: Terminal block X1 - Modbus interface Notes!Consider the in...
Emerson Process Management GmbH & Co. OHG 4-30 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 4.6.3 Installation - X-STREAM X2XF Field Housings If fitted, the optional RJ45 connection is loca - ted on an electronics board in the card cage section of the unit (Fig. 4-9, pg. 4-23). To install...
Emerson Process Management GmbH & Co. OHG 4-31 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.6.3 Installation - X-STREAM X2XF Field Housings Fig. 4-24: Terminal blocks for digital inputs and outputs Digital inputs Quantity: 9 (1 terminal block) or 18 (2 terminal blocks), dr...
Emerson Process Management GmbH & Co. OHG 4-32 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Connecting the power cord The power cord is connected to screw-type terminals located inside the housing. Supported wire cross sections: 0.2…4 mm 2 (24…12 AWG), no need to use wire end sleeves Cabl...
Emerson Process Management GmbH & Co. OHG 4-33 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.6.3 Installation - X-STREAM X2XF Field Housings ELECTRICAL SHOCK HAZARD BY MISSING EARTHING CONDUCTOR Before completing the electrical connection of the instrument, verify cables ar...
Emerson Process Management GmbH & Co. OHG 4-34 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 4.7 Notes On Wiring Signal Inputs and Outputs 4.7 Installation - Notes on Wiring Emerson Process Managament has made every effort during the development process to ensure that the X-STREAM analyzer...
Emerson Process Management GmbH & Co. OHG 4-35 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation On-site conditions often differ from test environments and may require special pre-cautions. Such a case arises when strong electromagnetic fields which could induce an interference c...
Emerson Process Management GmbH & Co. OHG 4-36 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 X-STREAM X2GP with screw-type terminal adapters In order to avoid measured values being influ - enced by external interference signals when terminal adapters are in use, the signal cable shieldings...
Emerson Process Management GmbH & Co. OHG 4-37 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 4 Installation 4.7.2 Wiring Inductive Loads Switching inductive loads creates electroma-gnetic interference: When an inductive load (e.g. relay, valve) is switched off, the magnetic field resists th...
Emerson Process Management GmbH & Co. OHG 4-38 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Frequently, several loads in one system are controlled by several analyzer outputs, whe - reby the power for the loads derives from a common source. Special care is needed when wiring the loads to ...
Emerson Process Management GmbH & Co. OHG 5-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup Chapter 5 Startup Once the unit has been unpacked and in - stalled, it is recommended to first check the settings and if necessary adjust them to the user’s needs. e.g: • What hardware is i...
Emerson Process Management GmbH & Co. OHG 5-2 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 All X-STREAM X2 gas analyzers have an alphanumeric display with four lines of 20 characters to display measuring and status information and the easy-to-use menu-based user interface for entering par...
Emerson Process Management GmbH & Co. OHG 5-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup 5.2.1 Display The display has 4x20 characters, either li- quid crystal or vacuum fluorescent (LCD or VFD).What information is displayed depends on the currently displayed menu. 5.2.2 Status...
Emerson Process Management GmbH & Co. OHG 5-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.2.3 Keys Six keys enable the use of the menu system. Depending on the operational mode (measu- ring, browsing menus, editing) they have the following functions: ENTER key: Mode Function Measuring ...
Emerson Process Management GmbH & Co. OHG 5-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup LEFT key: Mode Function Measuring L e a v e s t h e m e a s u r e m e n t display Browsing Goes up 1 level or page in menu system Editing Moves cursor 1 space Leaves channel selection Cance...
Emerson Process Management GmbH & Co. OHG 5-6 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.3 Symbols Used 5.3 Symbols Used Symbol Description Within Process Descriptions Setup.. Menu title Setup.. Analog outputs.. Parent ( Setup ) and current Menu ( Analog outputs ) Analog outputs.. Out...
Emerson Process Management GmbH & Co. OHG 5-7 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup 5.4 Software The analyzer software displays measure-ment results and status messages, allows parameters to be set and edited and allows maintenance functions (e.g. calibration) to be carrie...
Emerson Process Management GmbH & Co. OHG 5-8 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Editing Editing mode enables the setting of a parame - ter. It is initiated by pressing the 8 key. The cursor is now placed over the last cha-racter of the current value. Pressing the keys change...
Emerson Process Management GmbH & Co. OHG 5-9 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup 5.4.2 Access Levels Access levels can be used to prevent changes to parameters by unauthorised personnel. The X-STREAM menu system supports four prioritized access levels which can be acti ...
Emerson Process Management GmbH & Co. OHG 5-10 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.4.3 Special Messages Depending on the last action performed by the user, one of the following messages may be displayed to assist or inform the user (the two confirmation messages are displayed o...
Emerson Process Management GmbH & Co. OHG 5-11 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup 5.5 Powering Up When the unit is powered up, a series of in-ternal tests is automatically performed. During this time the front panel keys are disabled, The measurement display is shown • ...
Emerson Process Management GmbH & Co. OHG 5-12 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 If the analyzer is operational and it becomes clear that the incorrect language has been set, which is unintelligible to the operator, the 5.6 Selecting the Language The DOWN key changes the langua...
Emerson Process Management GmbH & Co. OHG 5-13 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup 5.7 Checking the Settings Language.. Setup.. Display.. Starting with the measurement display (5.5.2, page 5-11), pressing any key except the MEASURE key will access the MAIN MENU; from her...
Emerson Process Management GmbH & Co. OHG 5-14 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.7 Checking the Settings The cursor is now in the “In/Outputs” line over an arrowhead. Press the DOWN key to display the next menu page and open the INSTALLED OPTIONS submenu. InstalledOptions.....
Emerson Process Management GmbH & Co. OHG 5-15 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup 5.7 Checking the Settings Protocol: MODB RTU MODB Mode: 32Bit ID number: 2 Interface: RS485/2w Baud rate: 19200 Parity: No In the two pages of the COMMUNICATION menu, you can verify th...
Emerson Process Management GmbH & Co. OHG 5-16 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.7 Checking the Settings Check the settings for the measurement dis-play, temperature and pressure units, and for menu access: use the DOWN and ETER keys to access the submenus . If a setting is n...
Emerson Process Management GmbH & Co. OHG 5-17 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup Multi-channel unit:Select the component to be set in the gas component selection menu. Once the display settings have been checked, press the LEFT key to return to the SETUP menu, then ope...
Emerson Process Management GmbH & Co. OHG 5-18 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.7 Checking the Settings This prevents calibration from being perfor - med when the incorrect gas is supplied (e.g. span gas calibration using zero gas), which would result in an incorrectly confi...
Emerson Process Management GmbH & Co. OHG 5-19 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup The first line allows the user to enter the cur - rent air pressure manually when no pressure sensor is installed, or to read the current pressure when a sensor is installed ( INSTALLED OP...
Emerson Process Management GmbH & Co. OHG 5-20 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.7 Checking the Settings Press the LEFT key to return to the SETUP menu, and then open the IN/OUTPUTS menu. and from there the ANALOG OUTPUTS menu. 5.7.4 Setting the Analog Outputs Analog outputs ...
Emerson Process Management GmbH & Co. OHG 5-21 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup 5.7 Checking the Settings Operational modes conforming to NAMUR 43 recommendations (NE 43) The modes described so far do not generate any signal which would allow the detection of a failur...
Emerson Process Management GmbH & Co. OHG 5-22 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.7 Checking the Settings Analog outputs SignalRange: 0-20mA Hold on cal: Yes Page 1 The behaviour of the outputs during calibrati - on can also be set on page 1 of the ANALOG OUTPUTS menu (“Hold...
Emerson Process Management GmbH & Co. OHG 5-23 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup Signal: Comp-1LowScale: 0 ppmHighScale: 1000 ppm Zoom.. The “Signal” parameter defines the value to be sent to the selected output. The following options (partly dependent on the number ...
Emerson Process Management GmbH & Co. OHG 5-24 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 The last line on this menu allows the zoom function of the analog output to be set when that output has been assigned the Zoom signal.This function allows a part of the signal range, specified by “...
Emerson Process Management GmbH & Co. OHG 5-27 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup 5.7 Checking the Settings 5.7.4.1 References to the Accuracy of the Analog Outputs Scaling or zooming relates to the analog outputs only and does not affect front panel display nor serial ...
Emerson Process Management GmbH & Co. OHG 5-28 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.7 Checking the Settings Level1: 100 ppm Function: Low Level2: 500 ppm Function: High Press the LEFT key until the SETUP menu is displayed, then select “Alarms” and open the submenu. If you are us...
Emerson Process Management GmbH & Co. OHG 5-29 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup 5.7 Checking the Settings The “SpanRange” parameter is displayed in the INFO - RANGE menu ( 6.2.5.1, page 6-62) and is always given as the percentage of the upper range limit of the select...
Emerson Process Management GmbH & Co. OHG 5-30 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 5.7 Checking the Settings Note!Off FS is preset by default unless otherwise specified on time of order. • Low pre-alarm and main alarm: A pre-alarm and a main alarm are set for falling concentraion...
Emerson Process Management GmbH & Co. OHG 5-31 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup Standard mode: An alarm results in the assigned relay being activated.Settings: • Level 1 > Level 2 • Level 1-Function: High • Level 2-Function: Low If an alarm is active, a correspondi...
Emerson Process Management GmbH & Co. OHG 5-32 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 • Defining high pre-alarm and main alarm Standard mode: An alarm results in the assigned relay being activated.Settings: • Level 1 > Level 2 • Level 1-Function: High • Level 2-Function: High If ...
Emerson Process Management GmbH & Co. OHG 5-33 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 5 Startup • Defining low pre-alarm and main alarm Standard mode: An alarm results in the assigned relay being activated.Settings: • Level 1 > Level 2 • Level 1-Function: Low • Level 2-Function: L...
Emerson Process Management GmbH & Co. OHG 5-34 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 The most important parameters have now been checked and the unit’s settings adjusted to your needs. A backup copy of these configuration data can now be made and saved. 5.7 Checking the Settings Pr...
Emerson Process Management GmbH & Co. OHG 6-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus Chapter 6 User Interface and Software Menus This chapter describes the structure and contents of the X-STREAM X2 gas analyzer software menus. While all the software menues are descri...
Emerson Process Management GmbH & Co. OHG 6-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System Fig. 6-1: X-STREAM Software Menu Structure 6.2 Menu System Zero calibration.. Span calibration.. Adv. calibration.. Apply gas.. Acknowledgements.. Zoom.. Language.. D...
Emerson Process Management GmbH & Co. OHG 6-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System CO2.1 135.1 ppmO2.2 201952 ppmTemp-1 58.8 °C (Messages) Measurement Display Control..Setup..Status..Info.. Setup menu 6.2.3, page 6-14 Status menu 6.2.4, page 6-50 Info menu 6.2.5, p...
Emerson Process Management GmbH & Co. OHG 6-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Control Menu 6.2.2 Control Menu Control.. If the system is set up accordingly, the access code for level 1 must be entered to gain ac-cess to this menu. Lock menu...
Emerson Process Management GmbH & Co. OHG 6-6 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Cancel calibration!Start calibration!Zero gas 0.0 ppm CO2.1 134.1 ppm 6.2.2.1 Zero Calibration Menu 6.2 Menu System - Control Menu Page 2 RESET.. Status.. CO2.1 134.1 ppm Current measured gas co...
Emerson Process Management GmbH & Co. OHG 6-8 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2.2.2.1 RESET Calibration Menu 6.2 Menu System - Control Menu RESET Are you sure?No!Yes! This menu appears if the user selected the menu line "Reset.." in either the ZERO CA - LIBRATION or...
Emerson Process Management GmbH & Co. OHG 6-9 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2.2.2.2 Calibration Status Menu 6.2 Menu System - Control Menu Gasflow Spangas CO.1 13.304 ppmProcedure NoneTime 0 s The first line indicates the currently used gas. This line show...
Emerson Process Management GmbH & Co. OHG 6-10 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Cancel calibration!ZeroAll!SpanAll!ZSCalAll! Control.. Adv. calibration.. 6.2.2.3 Advanced Calibration Menu 6.2 Menu System - Advanced Calibration Menu Pressing the 8 key in this line will abort th...
Emerson Process Management GmbH & Co. OHG 6-11 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2.2.4 Apply Gas Menu Use the and keys to switch between Sample , Zero , Span and None . When the selected value is confirmed with the 8 key, the corresponding valve will be op...
Emerson Process Management GmbH & Co. OHG 6-12 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2.2.5 Acknowledgements Menu 6.2 Menu System - Acknowledgements Menu Control.. Acknowledgements.. All status messages are acknowledged and reset here: simlpy press the 8 key, to ack- nowledge mess...
Emerson Process Management GmbH & Co. OHG 6-13 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2.2.6 zoom Menu 6.2 Menu System - Zoom Menu Control.. Zoom.. This menu allows the operator to determine which analog outputs are "zoomed”, if any: Off : Output is not zoomed. ...
Emerson Process Management GmbH & Co. OHG 6-14 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Setup Menu 6.2.3 Setup Menu Setup.. Display..Calibration..Measurement.. In/Outputs.. Display setup 6.2.3.1, page 6-15 Calibration setup 6.2.3.2, page 6-22 In/Output setup 6.2.3....
Emerson Process Management GmbH & Co. OHG 6-15 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Display Setup Menu 6.2.3.1 Display Setup If the system is set up accordingly, the access code for level 2 must be entered to gain ac-cess to this menu. If the syst...
Emerson Process Management GmbH & Co. OHG 6-16 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Display Setup Menu Component display setup 6.2.3.1.3, Page 6-19 Component.. Temperature..Pressure.. Gasflow... Page 3 Temperature display setup 6.2.3.1.4, Page 6-20 Gasflow di...
Emerson Process Management GmbH & Co. OHG 6-17 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus Language Language: EN 6.2.3.1.1 Display Language Setup 6.2 Menu System - Display Setup Menu Selects the preferred language for the ana- lyzer software. Available options may vary ac...
Emerson Process Management GmbH & Co. OHG 6-18 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Code 1: Off Code 2: Off Code 3: Off Activate On Code 1: 00000001 Code 2: 00000002 Code 3: 00000003 6.2.3.1.2 Menu Access Setup Lines 1 to 3 determine whether the correspon-ding access level is ...
Emerson Process Management GmbH & Co. OHG 6-19 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus Unit Text: ppmFactor: 1.0000Offset: 0.0000 ComponentTag: CO2.1DecimalPlaces: 1 CO2.1 134.1 ppm 6.2 Menu System - Display Setup Menu 6.2.3.1.3 Component Menu Multi-channel unit:S...
Emerson Process Management GmbH & Co. OHG 6-20 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Display Setup Menu 6.2.3.1.4 Temperature Menu Setup..Display..Temperature.. TemperatureUnit: °CDecimalPlaces: 1Temp-1 63.7 °C Current temperature, here: Sensor 1 value Set tempera...
Emerson Process Management GmbH & Co. OHG 6-21 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Display Setup Menu 6.2.3.1.6 Gasflow Menu Setup..Display.. Gasflow.. Gasflow Unit: l/hDecimalPlaces: 1Flow-1 63.7 l/h Current gas flow, here: Sensor 1 value Set ga...
Emerson Process Management GmbH & Co. OHG 6-23 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Calibration Setup Valve assignment.. Interval time.. Page 2 Assign valves 6.2.3.2.2, page 6-26 Set interval time for calibration 6.2.3.2.3, page 6-26 Note!This s...
Emerson Process Management GmbH & Co. OHG 6-24 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Calibration gases ZeroGas: 0.0 ppm SpanGas: 500.0 ppm 6.2 Menu System - Calibration Setup 6.2.3.2.1 Calibration Gases Menu Setup..Calibration..Calibration gases.. The concentration of the zero gas ...
Emerson Process Management GmbH & Co. OHG 6-25 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Valve Assignment Menu 6.2.3.2.2 Valve Assignment Menu Setup..In/Output..Valve assignment.. Multi-channel unit:Select the component to be set in the SELECT COMPONEN...
Emerson Process Management GmbH & Co. OHG 6-26 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Interval timeZeroAll: 2 h ZSCalAll: 10 h AutoCal in.. 6.2 Menu System - Calibration Setup 6.2.3.2.3 Interval Time Menu If the system is set up accordingly, the access code for level 3 must be enter...
Emerson Process Management GmbH & Co. OHG 6-27 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Calibration Setup 6.2.3.2.3.1 AutoCal in Menu Setup..Calibration..Interval time..AutoCal in.. Note!This menu appears only, if at least one interval time has been s...
Emerson Process Management GmbH & Co. OHG 6-28 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Pressure: 1014.0 hPaDamping.. 6.2 Menu System - Measurement Setup 6.2.3.3 Measurement Menu If no pressure sensor is installed (INSTALLED OPTIONS - PRESSURE.. set to Manual ), the current ambient pr...
Emerson Process Management GmbH & Co. OHG 6-29 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus Damping: 5 s 6.2.3.3.1 Signal Damping Settings The time for the electronic signal damping is entered here.Lower values give results which are updated more quickly, higher values sup...
Emerson Process Management GmbH & Co. OHG 6-30 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Analog outputs..DigitalOutputs..DigitalInputs..IntSHS.. 6.2.3.4 In/Outputs Setup 6.2 Menu System - I n/Outputs Setup Setup..In/Outputs.. Analog outputs menu 6.2.3.4.1, page 6-31 If the system is se...
Emerson Process Management GmbH & Co. OHG 6-31 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus Analog outputs SignalRange: 0-20mA Hold on cal: No 6.2 Menu System - Analog Outputs Menu 6.2.3.4.1 Analog Outputs Menu This line sets the output signal range for all outputs. Avai...
Emerson Process Management GmbH & Co. OHG 6-32 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Analog Outputs Menu Page 2 Page 3 Available analog outputs (min. 1, max. 4) can be set in greater detail via these submenus. 6.2.3.4.1.2, page 6-34 The lines on this page enable t...
Emerson Process Management GmbH & Co. OHG 6-34 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Analog Outputs Menu 6.2.3.4.1.2 Output1 (2 ... 4) Menu Selects the parameter to be displayed in the corresponding line in the measurement dis- play. Depending on the number of mea...
Emerson Process Management GmbH & Co. OHG 6-35 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus Enter this second menu page to trim the se-lected analog output signal: Thus, if the signal differs from the selected value (0/4 or 20 mA) it can be adjusted to compensate the offse...
Emerson Process Management GmbH & Co. OHG 6-36 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Analog outputs menu 6.2.3.4.1.2.1 Zoom Output Menu Setup..In/Outputs..Analog outputs..Output1 (2...4 ).. Zoom.. Switching: ManualZoom: 50 %Position: LowScaleStatus: Off Determines...
Emerson Process Management GmbH & Co. OHG 6-37 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Digital Outputs Menu 6.2.3.4.2 Digital Outputs Menu Setup..In/Outputs..DigitalOutputs.. Outputs are assigned functions in these lines. Avalable options: next page....
Emerson Process Management GmbH & Co. OHG 6-38 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Digital Outputs Menu The following functions can be assigned to each of the digital outputs (max. 22). Label Description Off Digital output unused On Digital output constantly on ...
Emerson Process Management GmbH & Co. OHG 6-39 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Digital Inputs Menu 6.2.3.4.3 Digital Inputs Menu Setup..In/Outputs..DigitalInputs.. Inputs are assigned to signals in these li-nes., Available options: next page....
Emerson Process Management GmbH & Co. OHG 6-41 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - IntSHS Menu 6.2.3.4.4 IntSHS Menu Setup..In/Outputs..IntSHS.. Gas5: Sample Gas6: V5Gas7: V6 Gas8: V7 Page 2 Gas1: V1Gas2: V2Gas3: V3 Gas4: V4 Page 1 For an o...
Emerson Process Management GmbH & Co. OHG 6-42 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - IntSHS Menu Note!Each virtual valve V1 to V8 is normally only assigned to one gas inlet, “Gas1” to “Gas8”. Double assignments are possible, but only meaningful if the unit is conf...
Emerson Process Management GmbH & Co. OHG 6-43 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2.3.5 Installed Options Menu 6.2 Menu System - Installed Options Menu Setup..Installed Options.. If the system is set up accordingly, the access code for level 3 must be entered t...
Emerson Process Management GmbH & Co. OHG 6-44 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Installed Options Menu 6.2.3.5.1 Installed Options, More Menu TempSensor: Temp-1FlowSensor: Flow-1 Setup..Installed Options..More.. Multi-channel unit:Select the component in the ...
Emerson Process Management GmbH & Co. OHG 6-47 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Alarm Setup Menu Multi-channel unit:Press the key to enter the SELECT COM- PONENT menu to change the settings for a different channel. Hysteresis: 5.00 % FlowL...
Emerson Process Management GmbH & Co. OHG 6-48 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Save-LoadCfgData > SvcPort!SvcPort > CfgData.. Verify! 6.2.3.8 Save-Load Menu This line initiates the procedure for saving configuration data to an external device: the data are sent to the...
Emerson Process Management GmbH & Co. OHG 6-49 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus Page 2 FactData > CfgData..CfgData > UserData..UserData > CfgData.. 6.2 Menu System - Save-Load Menu Comments: FactData T h i s r e p r e s e n t s t h e f a c t o r y se...
Emerson Process Management GmbH & Co. OHG 6-50 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Status Menu 6.2.4 Status Menu Status.. Note!All entries in this menu and its submenus are not editable and provide information only. Note!The numbers at the start of each line on ...
Emerson Process Management GmbH & Co. OHG 6-51 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Status Menu 6.2.4.1 Failures Status..1 Failures.. The number in the “Failures” line in the pre- vious menu indicates how many errors are currently active (here: 1)...
Emerson Process Management GmbH & Co. OHG 6-52 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2.4.2 Maintenance Requests Status..0 Check requests.. 6.2 Menu System - Status Menu The number in the “Check requests” line in the previous menu indicates how many errors are currently active (he...
Emerson Process Management GmbH & Co. OHG 6-53 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2.4.3 Function Checks Status..1 Function checks.. 6.2 Menu System - Status Menu The number in the “Function checks” line in the previous menu indicates how many errors are current...
Emerson Process Management GmbH & Co. OHG 6-55 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus Gasflow V2 Procedure PurgingTime 10 s AutoCal in.. 6.2 Menu System - Status Menu 6.2.4.5 Calibration Status Status..Calibration.. If the unit is currently being calibrated, this m...
Emerson Process Management GmbH & Co. OHG 6-56 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Note!For more detailed information about calibrati-on status, Chapter 7 Maintenance. Calibration Offset 0.000000 Factory settingsOffset 0.000000 These two lines show a comparison of the offsets c...
Emerson Process Management GmbH & Co. OHG 6-57 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2.4.5.1 Auto Calibration Status..Calibration..AutoCal in.. 6.2 Menu System - Status Menu ZeroAll 1 hZeroAll 15 minZSCalAll 1 hZSCalAll 15 min The first two lines show the remainin...
Emerson Process Management GmbH & Co. OHG 6-58 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Temp-1 54 °C Temp-2 44 °C Temp-3 44 °C Temp-4 44 °C 6.2.4.6 Measurement Status Page 1 Displays the current pressure, either from an internal or external sensor or manually entered. 6.2.3.5, page ...
Emerson Process Management GmbH & Co. OHG 6-60 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Alarms Level1 OffLevel2 Off 6.2.4.7 Alarm Status Multi-channel unit:Press the key to enter the SELECT COM- PONENT menu to open the same menu for a different channel. Multi-channel unit:Select the...
Emerson Process Management GmbH & Co. OHG 6-61 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2.5 Info Menu X-STREAM-2M Revision 1.0Sensor 1077 XCA05812345678 Page 1 Page 1b (optional) 6.2 Menu System - Info menu Installed software version Analyzer model Analyzer serial ...
Emerson Process Management GmbH & Co. OHG 6-62 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 6.2 Menu System - Info menu 6.2.5.1 Information On Measurement Range This menu displays information on the range for the selected channel. Info..Range.. These two lines display the measuring ranges...
Emerson Process Management GmbH & Co. OHG 6-63 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 6 Software Menus 6.2 Menu System - Info menu 6.2.5.2 Installed Options Installed OptionsValves Int+ExtCOM-Interf Yes Pump Yes Page 1 DigitalIO No Pressure InternalMore.. Page 2 These pages displ...
Emerson Process Management GmbH & Co. OHG 7-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures Chapter 7 Maintenance and other Procedures Maintenance carried out on a regular basis ensures long-term efficiency of your EMER - SON Process Management gas analyzer! 7...
Emerson Process Management GmbH & Co. OHG 7-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.2 General Maintenance Information Tests Component Interval Tubing, flexible Leak Test Once a year Pressure sensor, pressure switch Valve block Pump diaphragm Leak Tes...
Emerson Process Management GmbH & Co. OHG 7-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Max. pressure 7.25 psig (500 mbar)! Multi channel instruments:Analyzers with parallel tubing require separate leak tests for each gas path ! 7.3 Performing a Leak Test To achieve best and proper mea...
Emerson Process Management GmbH & Co. OHG 7-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4 Calibration Procedures 7.4 Calibration Procedures Note!To achieve best and proper measuring results it is recommended to perform zero and span calibrations on a reg...
Emerson Process Management GmbH & Co. OHG 7-6 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.4 Calibration Procedures Type of Procedure Menu Page (CONTROL - ) Valves Simulta- neously Calibrated Channels More Information Manual calibration ZERO CALIBRATION... optional single channel page 7...
Emerson Process Management GmbH & Co. OHG 7-7 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures Calibration gases ZeroGas: 0.0 ppm SpanGas: 500.0 ppm 7.4.1 Preparing Calibrations 7.4.1 Preparing Calibrations Before starting calibrations it is required to tell the ...
Emerson Process Management GmbH & Co. OHG 7-10 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Fig. 7-2: Calibration Improvement by Variable Valve Assignments 7.4.1.1 Valve Assignment for Valve Supported Calibrations 7.4.1 Preparing Calibrations As described earlier, several calibration proc...
Emerson Process Management GmbH & Co. OHG 7-11 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures Before starting to assign valves to gases and channels, you need to check what type of valves is used (external or internal): Setup..Installed Options.. Open the SETUP...
Emerson Process Management GmbH & Co. OHG 7-12 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.4.1 Preparing Calibrations 7.4.1.1.1 Internal Valve Assignment If your analyzer provides internal valves, at first open the SETUP - IN/OUTPUTS - INTSHS menu to assign the valves to the gas inlets...
Emerson Process Management GmbH & Co. OHG 7-13 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.1 Preparing Calibrations The exemplary settings on the previous page could relate to the following analyzer confi - guration: Fig. 7-3: Internal Valves Assignments...
Emerson Process Management GmbH & Co. OHG 7-14 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.4.1 Preparing Calibrations 7.4.1.1.2 External Valve to Digital Output Assignment If your analyzer has to control external valves, at first check if all valves required for calibra - tion are conn...
Emerson Process Management GmbH & Co. OHG 7-15 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.1 Preparing Calibrations 7.4.1.1.3 Calibration Valve Assignment For each channel a valve has to be assigned zero gas valve or span gas valve, whereat the valves ca...
Emerson Process Management GmbH & Co. OHG 7-17 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.2 Manual Calibration 7.4.2 Manual Calibration Zero calibration..Span calibration..Adv.Calibration.. Apply gas.. 7.4.2.1 Manual Zero Calibration CANCEL calibratio...
Emerson Process Management GmbH & Co. OHG 7-19 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.2 Manual Calibration Gasflow SpanGas CO2.1 20.000 ppmProcedure SpanningTime 10 s After having started the calibration watch the screen for status information: The ...
Emerson Process Management GmbH & Co. OHG 7-21 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.3 Advanced Calibration 7.4.3.1 Zero All Calibration Cancel calibration!ZeroAll!SpanAll!ZeroSpanAll! Before selecting any further line make sure the required calibr...
Emerson Process Management GmbH & Co. OHG 7-22 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.4.3 Advanced Calibration zero all channels with the current zero gas valve assigned select channel 1 increase channel no. by 1 select all channels with the same zero gas valve assigned is current...
Emerson Process Management GmbH & Co. OHG 7-23 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.3 Advanced Calibration Gasflow V4 Procedure PurgingTime 10 s The procedure starts showing the screen to the left, indicating that valve V4 is open. Currently the s...
Emerson Process Management GmbH & Co. OHG 7-24 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.4.3.2 Span All Calibrations 7.4.3 Advanced Calibration Cancel calibration!ZeroAll!SpanAll!ZeroSpanAll! To start a SPAN calibration for ALL channels select the third line. Note!Single channel anal...
Emerson Process Management GmbH & Co. OHG 7-25 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.3 Advanced Calibration span all channels with the current span gas valve assigned select channel 1 increase channel no. by 1 select all channels with the same span...
Emerson Process Management GmbH & Co. OHG 7-27 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.3.3 Zero & Span All Calibration 7.4.3 Advanced Calibration Before selecting any further line make sure the required calibra-tion gas is applied! Supply all cal...
Emerson Process Management GmbH & Co. OHG 7-28 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.4.3 Advanced Calibration Fig. 7-6: Zero Span All Calibration Procedure Flow Diagram zero all channels with the current zero gas valve assigned select channel 1 increase channel no. by 1 select al...
Emerson Process Management GmbH & Co. OHG 7-29 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.3 Advanced Calibration Gasflow V4 Procedure ZeroingTime 29 s When the prepurge time has elapsed, the instrument starts to calculate the zero point (here indicated ...
Emerson Process Management GmbH & Co. OHG 7-31 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.4 Remote Calibrations Remote calibrations may be initialized by di - gital inputs or Modbus commands, whereas both offer different functionalities:Remote calibrati...
Emerson Process Management GmbH & Co. OHG 7-33 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.4 Remote Calibrations Fig. 7-7: Digital Inputs - Initializing Calibrations Example 1: Signal A starts a ZeroAll calibrationSignal B is applied during the ongoing Z...
Emerson Process Management GmbH & Co. OHG 7-35 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.4.3 Modbus Activated Calibrations With Valves 7.4.4 Remote Calibrations Several Modbus commands allow to start calibrations ( Chapter 9, List of Modbus Commands). ...
Emerson Process Management GmbH & Co. OHG 7-37 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.5 Unattended Automatic Calibration Interval timeZeroAll: 5 h ZSCalAll: 5 h AutoCal in.. Setup..Calibration..Interval time.. Within the SETUP CALIBRATION menu the I...
Emerson Process Management GmbH & Co. OHG 7-38 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.4.5 Unattended Automatic Calibration ZeroAll: 1 h ZeroAll: 15 min ZSCalAll: 2 h ZSCalAll: 45 min ZeroAll: 0 h ZeroAll: 5 min ZSCalAll: 15 h ZSCalAll: 5 min At first the AUTOCAL IN.. menu serves a...
Emerson Process Management GmbH & Co. OHG 7-39 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4.6 Resetting a Calibration 7.4.7 Verifying a Calibration 7.4.6 Resetting a Calibration In case a wrong configuration was detected after calibration was performed (e...
Emerson Process Management GmbH & Co. OHG 7-40 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.4 Calibration Procedures 7.4.8 Cancelling an Ongoing Calibration Changing the APPLY GAS parameter opens the related valve. Available options: SpanGas, ZeroGas, Sample, None. The "Time" li...
Emerson Process Management GmbH & Co. OHG 7-41 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.4 Calibration Procedures CO2.1 0.000 ppm Procedure Zero 1Time 0 s Gasflow Sample CO2.1 0.000 ppm Procedure PurgingTime 10 s A confirmation popup appears, replaced by...
Emerson Process Management GmbH & Co. OHG 7-43 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures Fig. 7-9: X-STREAM X2GP If your instrument is equipped with a handle • loosen the 6 screws at the front panel, • to only get access to the cover screws, push frame and...
Emerson Process Management GmbH & Co. OHG 7-44 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Fig. 7-11: X-STREAM X2 Field Housings and X2FD - How to Open Sash fasteners Screws at the flange 7.5.2.4 How to Open X-STREAM X2FD To open a X-STREAM X2FD loosen the 20 screws located at the instru...
Emerson Process Management GmbH & Co. OHG 7-45 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.5 Replacing Worn Out Sensors Separate sections describe the replacement of the various sensors: Electrochemical oxygen sensor 7.5.3, page 7-46 Trace oxygen sensor (t...
Emerson Process Management GmbH & Co. OHG 7-47 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures Do not expose the sensor to a temperature other than the temperature range of -20 to +60‘C (-4 to +140 F). Exposing to a temperature outside the temperature range may ...
Emerson Process Management GmbH & Co. OHG 7-48 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 In consequence of it‘s design the sensor‘s lifetime is limited and depends on theoretical designed life and O xygen concentration. The sensor output can be taken as a rough criterion for end of lif...
Emerson Process Management GmbH & Co. OHG 7-49 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.5.3.1 Locating the Sensor The instruments provide two different varia-tions of internal designs (except 1 ⁄ 2 19“ instru - ments): • In instruments with internal hea...
Emerson Process Management GmbH & Co. OHG 7-50 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.5.3.2 Disassembling the Sensor Unit The sensor unit consists of a holder, an elec- tronics board and the sensor itself, all together installed on a base plate (Fig. 7-13). After loosening the nut...
Emerson Process Management GmbH & Co. OHG 7-51 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.5.3.3 Adjusting the Output Signal Fig. 7-14: Sensor At Rear Panel Fig. 7-15: OXS Board, Top View Procedure: • power on the open instrument.• Supply ambient air (appr...
Emerson Process Management GmbH & Co. OHG 7-52 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 • Disconnect the analyzer from power • Re-install the sensor unit into the analyzer • Close the housing. Take care to use all screws, especially if the instrument is to be used in hazardous areas! ...
Emerson Process Management GmbH & Co. OHG 7-53 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.5 Replacing Worn Out Sensors 7.5.4 Replacing the Trace Oxygen Sensor Replacing the trace oxygen sensor requires considering special instructions, shipped to-gether w...
Emerson Process Management GmbH & Co. OHG 7-54 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.5.5 Replacing the Trace Moisture Sensor • Locate the sensor unit within your ana-lyzer. • Open the fittings, connecting the unit to the piping. • Only loosen the small nuts (do not remove them), ...
Emerson Process Management GmbH & Co. OHG 7-55 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures Seal the open analyzer´s gas fit - tings utilizing PVC caps to avoid pollution of inner gas path. • Moisten the lint-free cloth with a mixture of 3 parts of water and ...
Emerson Process Management GmbH & Co. OHG 7-56 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.7 Save / Restore Configuration Data Sets 7.7 Save / Restore Configuration Data Sets After some time of operating the instrument, one can assume all the parameters (calibration gases setup, measur...
Emerson Process Management GmbH & Co. OHG 7-57 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures Fig. 7-17: Relationship of Supported Data Sets and Where to Find Further Information UserData configuration saved to FRAM Backup on user request(CfgData --> SvcPort...
Emerson Process Management GmbH & Co. OHG 7-59 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.7.2 Restore UserData to CfgData 7.7. Save / R estore C onfiguration Data Sets 2 nd menu page FactData > CfgData..CfgData > UserData..UserData > CfgData.. ...
Emerson Process Management GmbH & Co. OHG 7-61 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures 7.7.4 Save / Restore CfgData to External Device 7.7.4.1 Preparation 7.7. Save / R estore C onfiguration Data Sets Before starting any of these procedures an external d...
Emerson Process Management GmbH & Co. OHG 7-62 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.7.4.2 Save CfgData to SvcPort Open the terminal software on your external device. 7.7. Save / R estore C onfiguration Data Sets If system is setup accordingly, access level 3 code must be entered...
Emerson Process Management GmbH & Co. OHG 7-63 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 7 Maintenance & Procedures Save-LoadCfgData > SvcPort!SvcPort > CfgData.. Verify! 7.7. Save / R estore C onfiguration Data Sets The SAVE-LOAD.. menu also offers an option to verify that ...
Emerson Process Management GmbH & Co. OHG 7-64 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 7.7.4.3 Restore SvcPort to CfgData 7.7. Save / R estore C onfiguration Data Sets Open the terminal software on your external device. SvcPort>CfgData Are you sure?No!Yes! A screen appears to conf...
Emerson Process Management GmbH & Co. OHG 8-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting Chapter 8 Troubleshooting 8.1 Abstract This chapter covers troubleshooting the ana - lyzer: Section 8.2 describes messages possibly ap- pearing in the measuring screen‘s status line...
Emerson Process Management GmbH & Co. OHG 8-2 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.2 Solving Problems Indicated by Status Messages 8.2 Solving Problems Indicated By Status Messages In the following all possible status messages are listed in an alphabetical order together with hi...
Emerson Process Management GmbH & Co. OHG 8-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.2.1 Analyzer related messages 8.2.1 Analyzer Related Messages Message Status level Explanation Description Actions FlowAlm Check request Explanation: The internal flow control det...
Emerson Process Management GmbH & Co. OHG 8-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting Message Status level Explanation Description Actions Calibration Function check Explanation: Calibration active Currently a calibration proce - dure is ongoing Wait for calibration ...
Emerson Process Management GmbH & Co. OHG 8-6 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.2.2 Channel Related Messages (Preceded By Channel Tag, E.g. CO2.1) Message Status levelExplanation Description Actions ADC-Error Failure Explanation: A/D conversion overflow The A/D converter of t...
Emerson Process Management GmbH & Co. OHG 8-11 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.3 Solving Problems Not Indicated By Status Messages The following table lists possible faults not detectable by the instrument‘s software, gives hints on the potential causes and...
Emerson Process Management GmbH & Co. OHG 8-12 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.3 Solving Problems Not Indicated by Status Messages Situation Description Actions Display dark Power supply missing Check power connection Check power supply Check instrument‘s power fusesCheck p...
Emerson Process Management GmbH & Co. OHG 8-13 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.3 Solving Problems Not Indicated by Status Messages Situation Description Actions Digital outputs not working properly (cont.) Outputs on extension board(s) (XDIO) affected XDIO:...
Emerson Process Management GmbH & Co. OHG 8-15 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.3 Solving Problems Not Indicated by Status Messages Situation Description Actions Fluctuating or invalid readout (continued) Gas path(s) polluted Check analysis cells and win-dow...
Emerson Process Management GmbH & Co. OHG 8-16 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Situation Description Actions No gas flow Sample gas pump (option) switched off Switch on sample gas pump ( 6.2.2, page 6-5) Membrane of sample gas pump defective Replace sample pump mem- brane Sam...
Emerson Process Management GmbH & Co. OHG 8-17 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting On Components This section give information on how to check and replace internal components. Some work described on the next pages need to be carried out by qua...
Emerson Process Management GmbH & Co. OHG 8-19 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components Fig. 8-1: X-STREAM X2GP 8.4.1 Opening X-STREAM Analyzers 8.4.1.1 How To Open X-STREAM X2GP 8.4.1.2 How To Open X-STREAM X2GK Remove the top cover ...
Emerson Process Management GmbH & Co. OHG 8-20 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.4 Troubleshooting on Components Fig. 8-3: X-STREAM X2 Field housings and X2FD - How to Open Sash fasteners Screws at the flange 8.4.1.4 How To Open X-STREAM X2FD To open a X-STREAM X2FD loosen th...
Emerson Process Management GmbH & Co. OHG 8-23 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components Step 1: If applicable: Remove the screws S1 on both sides of the pump. Take off the cover. Required parts for the spare parts kit for the pump (PN...
Emerson Process Management GmbH & Co. OHG 8-25 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components Step 5: Remove the white block. Step 4: Mark the pump assy. before di - sassembly.
Emerson Process Management GmbH & Co. OHG 8-26 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.4 Troubleshooting on Components Step 6: Remove the teflon gasket. Step 7: Remove the remaining two pump parts.Clean the white plate for the gas in- and outlet.
Emerson Process Management GmbH & Co. OHG 8-27 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components Step 8: Disassemble the lower block and the clamp. Loosen the screw S4 and the nut N1 . Clamp S4 N1
Emerson Process Management GmbH & Co. OHG 8-28 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.4 Troubleshooting on Components Step 9: Remove the two washers on the diaphragm. Step 10: Replace the old with the new dia- phragm and assemble the was - hers and the clamp in reverse order (step...
Emerson Process Management GmbH & Co. OHG 8-29 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components Step 11: Remove the locking springs on both sides of the white block and take out the old diaphragms on both sides.
Emerson Process Management GmbH & Co. OHG 8-30 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.4 Troubleshooting on Components Step 13: Assemble the pump assy. Take care of your marker ( step 4) 1. Put the two upper plates under the clamp ( steps 6 & 7 for reference). 2. Put the white ...
Emerson Process Management GmbH & Co. OHG 8-31 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components Step 14: Assemble the pump assy in re - verse order. Put it in the pump housing and fix it with the screws S2 . Fix the clamp with screw S3 and th...
Emerson Process Management GmbH & Co. OHG 8-32 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.4 Troubleshooting on Components Step 15: If applicable: Install the cover and fix it with screws S1 at both sides. Finally re-install the pump into your analy-zer, to complete the replace-ment of...
Emerson Process Management GmbH & Co. OHG 8-33 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components 8.4.4 Paramagnetic Oxygen Cell: Adjustment Of Physical Zero Step 1: The adjoining figure shows a heated paramagnetic oxygen cell. Step 2: Open the...
Emerson Process Management GmbH & Co. OHG 8-34 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.4 Troubleshooting on Components Step 5: Tighten the screw S2 with care, close the cover and check the zero point again. Step 4: Carefully loosen the screw S2 . Now you can adjust the physical zer...
Emerson Process Management GmbH & Co. OHG 8-36 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.4 Troubleshooting on Components 8.4.5 Thermal Conductivity Cell: Adjustment Of Output Signal To adjust the zero signal of this measuring cell you need to have access to both sides of the related ...
Emerson Process Management GmbH & Co. OHG 8-37 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components Step 1: Check the solder bridges, lo - cated at the solder side of the board, for proper configuration: LB10 openLB4 2-5 closed LB21 1-4 closed LB...
Emerson Process Management GmbH & Co. OHG 8-38 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.4 Troubleshooting on Components R60 Test connector P4 P4.15 P4.16 + _ Step 3: Locate test connector P4 to measure the bridge voltage : P4.16 Bridge voltage (+)P4.15 Bridge voltage (-); GND CAUTIO...
Emerson Process Management GmbH & Co. OHG 8-39 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components P11/ P17 Step 4: To adjust the physical zero point : Apply zero gas to the analyzer. Connect the DVM to the follo-wing pins:P4.5 Raw signal (+)P4....
Emerson Process Management GmbH & Co. OHG 8-40 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 8.4 Troubleshooting on Components Step 5: To adjust the physical span : Apply span gas to the analyzer. Do not disconnect the DVM:P4.5 Raw signal (+)P4.15 Bridge voltage (-); GND CAUTION! Do not sh...
Emerson Process Management GmbH & Co. OHG 8-41 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 8 Troubleshooting 8.4 Troubleshooting on Components Step 7: To finetune the physical zero point : Close solder bridge LB10. Apply zero gas to the analyzer. Do not disconnect the DVM:P4.5 Raw signal ...
Emerson Process Management GmbH & Co. OHG 9-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 9 Modbus Commands Chapter 9 Modbus Functions This chapter lists all Modbus functions and registers supported by X-STREAM gas analyzers. Refer to the www.Modbus-IDA.org website for detailled documenta...
Emerson Process Management GmbH & Co. OHG 9-2 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 9.2 Supported Functions Modbus Function Function Code Note 1) decimal (hex) ReadCoils 01 (0x01) for registers of 2000 ReadDiscreteInputs 02 (0x02) for registers of 1000 ReadHoldingRegisters 03 (0x03...
Emerson Process Management GmbH & Co. OHG 9-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 9 Modbus Commands 9.3 List of Parameters and Registers - Sorted by Tag Name Tag Name Address Data Type Client Access Description Modicon Daniel Control.AcknowledgeStates 2038 2038 Boolean R/W 1=Ackno...
Emerson Process Management GmbH & Co. OHG 9-23 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 9 Modbus Commands 9.4 List of Parameters and Registers - Sorted by Registers Address Tag Name Data Type Client Access Description Daniel Modicon 2033 2033 Control.ApplyGas.PumpState Boolean R/W 0=Of...
Emerson Process Management GmbH & Co. OHG 10-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 10 Service Information Chapter 10 Service Information If factory repair of defective equipment is required, proceed as follows: 1. Secure a return authorization from a Rosemount Analytical Sales Off...
Emerson Process Management GmbH & Co. OHG 10-2 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 10.2 Customer Service F o r o r d e r a d m i n i s t r a t i o n , r e p l a c e m e n t p a r t s , a p p l i c a t o n a s s i s t a n c e , o n - s i t e o r f a c t o r y r e p a i r, s e r v ...
Emerson Process Management GmbH & Co. OHG 11-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 11 Dismounting & Disposal Chapter 11 Dismounting and Disposal 11.1 Dismounting and Diposal of the Analyzer HAZARDS FROM DISMOUNTING Dismounting instruments installed in hazardous area requires s...
Emerson Process Management GmbH & Co. OHG 11-2 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 11 Dismounting & Disposal When the instrument has reached the end of its useful life, When the instrument has reached the end of its useful life, do not throw it in a trash can! This instrument...
Emerson Process Management GmbH & Co. OHG A-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 A Appendix Appendix This chapter contains an excerpt from the Modbus publication "Modbus_over_serial_line“ A.1, page A-2 EC declaration of conformity A.2, page A-12 CSA Certificate of Compliance ...
Emerson Process Management GmbH & Co. OHG A-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 A Appendix A.1 Modbus Specification
Emerson Process Management GmbH & Co. OHG A-13 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 A Appendix A.2 EC Declaration of Conformity
Emerson Process Management GmbH & Co. OHG A-14 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 A.3 CSA C ertificate of Compliance Certificate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2425669 Date Issued: June 9, 2011 DQD 507 Rev. 2009-09-01 Page: 1 Issu...
Emerson Process Management GmbH & Co. OHG A-35 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 A Appendix Instrument details Analyzer model Serial no. Process details Temperature Pressure Declaration and Sender DataWe hereby declare that the retumed parts have been carefully cleaned. To the b...
Emerson Process Management GmbH & Co. OHG A-37 A Appendix X-STREAM X2 Instruction Manual HASX2E-IM-EX02/2012 A.7 Assignment of Terminals and Sockets A.7.1 Tabletop & Rack Mount Analyzers A.7 Assignment Of Terminals And Sockets DC 24 V Input (X2GK) Socket X1 - Analog outputs, relay outputs 1....
X-STREAM X2 Emerson Process Management GmbH & Co. OHG A-38 Instruction Manual HASX2E-IM-EX 02/2012 A.7 Assignment of Terminals and Socket A.7.2 Field Housings Pin Signal P3.1 Input 1 P3.2 Input 2 P3.3 Input 3 P3.4 Input 4 P3.5 Input 5 P3.6 Input 6 P3.7 Input 7 P3.8 GND for Inputs 1-7 P3.9 not us...
Emerson Process Management GmbH & Co. OHG I-1 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 IDX Index Index A Acknowledgements Menu See also Software: Menus Alarms See also Software: Menus: Alarm Setup Menu ; See also Software: Menus: Alarm Status Menu Concentration Alarms 5-28 SpanRange 5-...
Emerson Process Management GmbH & Co. OHG I-3 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 IDX Index Index O Off-Spec See also Software: Menus: Status Menu ; See also Troubleshooting Opening the Analyzers 7-43 X-STREAM X2FD 7-44 , 8-20 X-STREAM X2GK 7-43 X-STREAM X2GP 7-43 X-STREAM X2XF 7-...
Emerson Process Management GmbH & Co. OHG I-4 X-STREAM X2 Instruction Manual HASX2E-IM-HS 02/2012 Index Calibration Setup Menu 6-22 Calibration Status Menu 6-55 Calibration Status Menu 6-9 Communications Settings 6-45 Control Menu 6-5 Digital Inputs Menu 6-39 Digital Outputs Menu 6-37 Display Se...
Emerson Process Management GmbH & Co. OHG I-5 X-STREAM X2 Instruction Manual HASX2E-IM-HS02/2012 IDX Index Index X-STREAM XCA 1-18 X-STREAM XCC 1-15 Z Zero Calibration Menu See also Software: MenusZoom 6-5 , 6-13 , 6-34 See also Setup: Analog Outputs Zoom Menu See also Zoom
Emerson Manuals
-
Emerson LD195EM8
User Manual
-
Emerson EWR20V4
User Manual
-
Emerson EWR20V4
Manual
-
Emerson SLC195EM8
User Manual
-
Emerson SLC195EM8
Manual
-
Emerson CR202EM9
User Manual
-
Emerson HD14070
User Manual
-
Emerson RLC320EM9
User Manual
-
Emerson LC320EM9
User Manual
-
Emerson LC320EM9
Manual
-
Emerson PD6922
User Manual
-
Emerson PD6922
Manual
-
Emerson EPT427D
User Manual
-
Emerson EPT427D
Manual
-
Emerson 400 VMC
User Manual
-
Emerson 400 VMC
Manual
-
Emerson CF960CK
User Manual
-
Emerson MICRO MOTION 2700
Installation Manual
-
Emerson CF787AP00
User Manual
-
Emerson CF787AP00
Manual