Page 2 - Quick Installation Guide; About this guide; Explosions could result in death or serious injury; Electrical shock can result in death or serious injury
March 2013 2 Quick Installation Guide About this guide This installation guide provides basic guidelines for Rosemount 5300 Series transmitters. It does not provide instructions for configuration, diagnostics, maintenance, service, troubleshooting, Explosion-proof, Flameproof, or intrinsically safe ...
Page 3 - Contents
Quick Installation Guide 3 March 2013 Any substitution of non-authorized parts or repair, other than exchanging the complete transmitter head or probe assembly, may jeopardize safety and is prohibited. Unauthorized changes to the product are strictly prohibited as they may unintentionally and unpred...
Page 4 - Step 1: Confirm system readiness; Confirm HART revision capability; Confirm correct Device Driver; ensure proper communication.; Rosemount 5300 device revisions and drivers; Device Driver and documentation for your device.; Table 1. Rosemount 5300 device revisions and files; Switch HART revision mode; To change to HART Revision 5, Enter: “HART5” in the Message field
March 2013 4 Quick Installation Guide Step 1: Confirm system readiness Confirm HART revision capability If using HART-based control or asset management systems, confirm the HART capability of those systems prior to transmitter installation. Not all systems are capable of communicating with the HAR...
Page 5 - Tank connection with flange
Quick Installation Guide 5 March 2013 Step 2: Mount the transmitter head/probe Tank connection with flange 1. Place a gasket on top of the tank flange. 2. Lower the probe with flange into the tank. 3. Tighten the bolts and nuts with sufficient torque regarding flange and gasket choice. 4. Loosen the...
Page 6 - Bracket mounting, on wall
March 2013 6 Quick Installation Guide Step 2 continued... Bracket mounting, on wall 1. Mount the bracket directly to the wall with screws suitable for the purpose. 2. Mount the transmitter with probe to the bracket and secure the installation with the three supplied screws. Bracket mounting, on pipe...
Page 7 - Remote housing
Quick Installation Guide 7 March 2013 Step 2 continued... Remote housing 1. Remove the transmitter head from the probe by unscrewing the M50 nut. For safety information regarding ESD discharge, see the applicable warning on page 3 . 2. Mount the probe in the tank. 3. Mount the bracket to the pole, m...
Page 8 - Step 3: Connect the wiring; Grounding
March 2013 8 Quick Installation Guide Step 3: Connect the wiring Wiring and power supply requirements can be dependent upon the approval certification. As with all F OUNDATION fieldbus requirements, a conditioned power supply and terminating resistors are required for proper operation. It is recomme...
Page 9 - To connect the transmitter; Verify that the power supply is disconnected.; Terminal block
Quick Installation Guide 9 March 2013 To connect the transmitter 1. Verify that the power supply is disconnected. 2. Remove the terminal block cover (see picture below). 3. Pull the cable through the cable gland/conduit. For Explosion-proof / Flameproof installations, only use cable glands or condui...
Page 10 - HART communication; within the loop in order to function properly, see diagrams below.; Non-intrinsically safe power supply
March 2013 10 Quick Installation Guide Step 3 continued... HART communication The Rosemount 5300 Series transmitter operates with a power supply ranging from 16 - 42.4 Vdc (16 - 30 Vdc in IS applications, 20-42.4 Vdc in Explosion-proof / Flameproof applications and in Non-Sparking / Energy Limited a...
Page 11 - Type n approvals : non-sparking / energy limited power supply
Quick Installation Guide 11 March 2013 Step 3 continued... Type n approvals : non-sparking / energy limited power supply Transient protection terminal block Load Resistance 250 Power Supply HART Modem Field Communicator PC Rosemount 5300 Series Transmitter HART: U n = 42.4 V Power Supply Load Resi...
Page 12 - Load limitations; The Field Communicator requires a minimum load resistance of 250
March 2013 12 Quick Installation Guide Step 3 continued... Load limitations The Field Communicator requires a minimum load resistance of 250 within the loop to function properly. The maximum load resistance can be obtained from the following diagrams. NOTE: For Flameproof / Explosion-proof install...
Page 13 - fieldbus; The Rosemount 5300 Series transmitter, F; fieldbus version, operates
Quick Installation Guide 13 March 2013 Step 3 continued... F OUNDATION fieldbus The Rosemount 5300 Series transmitter, F OUNDATION fieldbus version, operates with a power supply ranging from 9 - 32 Vdc (9 - 30 Vdc in IS applications and 16 - 32 Vdc in Explosion-proof / Flameproof and in Non-Sparking...
Page 14 - RS-485 with Modbus communication
March 2013 14 Quick Installation Guide Step 3 continued... Type n approvals : non-sparking / energy limited power supply RS-485 with Modbus communication The 5300 Series RS-485 with Modbus communication transmitter version operates using a power supply ranging from 8 - 30 Vdc (max. rating). See the ...
Page 15 - Field Communicator, the AMSTM Suite, DeltaV; fieldbus parameters are given.; Installing the Rosemount Radar Master software
Quick Installation Guide 15 March 2013 Step 4: Configure NOTE: If the transmitter is pre-configured at the factory, it is only necessary to proceed with the following steps if you need to verify or change the settings. Basic configuration can easily be done either with Rosemount Radar Master, a Fiel...
Page 16 - Master
March 2013 16 Quick Installation Guide Configuration using the Rosemount Radar Master software 1. Start Rosemount Radar Master ( Programs > Rosemount > Rosemount Radar Master ). 2. Connect to the desired transmitter. Once the transmitter is connected, the Guided Setup window appears automatica...
Page 17 - Tank Height; , the distance from the Upper Reference Point to the Lower; Mounting Type; if Mounting Type is Nozzle or; Nozzle Height; if Mounting Type is Nozzle.; fieldbus parameters:
Quick Installation Guide 17 March 2013 are normally configured at the factory, but can be changed if, for example, the probe is cut in the field, or if disturbing objects in the near zone need to be blocked out (Hold Off Distance/Upper Null Zone (UNZ)). Handheld HART Communicator: Fast Key Sequence ...
Page 19 - Guided Setup
Quick Installation Guide 19 March 2013 the available pre-defined tank options, or if high volume accuracy is desired.Choose None if volume calculation is not desired. Handheld HART Communicator: Fast Key Sequence [2, 1, 5] F OUNDATION fieldbus parameters: TRANSDUCER_1300 > VOL_VOLUME_CALC_METHOD ...
Page 20 - You may be required to calibrate Vapor Compensation, to set up; Setup - Modbus communication parameters; communication parameters:; Setup; menu select; General; Enter the desired Modbus communication settings.
March 2013 20 Quick Installation Guide 12. The Basic Configuration with the Radar Master Wizard is now finished. Proceed with the Device Specific Setup to see which additional configuration is needed. You may be required to calibrate Vapor Compensation, to set up Probe End Projection, or to perform ...
Page 21 - Installation; ) at different currents; Configuration
Quick Installation Guide 21 March 2013 Safety Instrumented Systems (4-20 mA only) The following section is applicable to the Rosemount 5300 Prior-Use option (Special certification: QS). Additional Safety Instrumented Systems information is available in the Rosemount 5300 Series Reference Manual (Doc...
Page 22 - Damping; damping values + response time; Alarm and saturation levels; Write protection; Table 3. Alarm levels and operation values; Rosemount Alarm Level; Namur Alarm Level
March 2013 22 Quick Installation Guide Damping User adjusted damping affects the transmitter’s ability to respond to process changes. Therefore, the damping values + response time should not exceed the Safety loop requirements. Alarm and saturation levels DCS or safety logic solver should be configu...
Page 23 - Site acceptance; Operation and maintenance; Proof test; Disable write protection if the function is enabled.; This step tests for possible quiescent current related failures.
Quick Installation Guide 23 March 2013 Site acceptance After the installation and configuration, proper operation of the transmitter should be verified. A site acceptance test is therefore recommended. The proof test outlined in this section can be used for this. Operation and maintenance The Rosemo...
Page 24 - Document the test result for future reference.; Inspection; reported. Feedback can be submitted electronically at; Contact Us; References; estimates. The full report is accessible at
March 2013 24 Quick Installation Guide 6. Enable write protection. 7. Restore the loop to full operation. 8. Remove the bypass from the safety logic solver or otherwise restore normal operation. 9. Document the test result for future reference. Inspection Visual inspection It is recommended to inspe...
Page 25 - EU conformity; The EC declaration of conformity can be found on; Hazardous Locations certifications; North-American certifications; Specific conditions of use:
Quick Installation Guide 25 March 2013 Product certifications EU conformity The EC declaration of conformity can be found on page 33 . The most recent revision of the EC declaration of conformity can be found at www.rosemount.com . Safety Instrumented Systems (SIS) The Rosemount 5300 Series has been...
Page 26 - Canadian Standards Association (CSA) approvals; Dual seal annunciation; Intrinsically Safe Exia:
March 2013 26 Quick Installation Guide I5, IE 1 Intrinsically safe and FISCO model: Intrinsically Safe for Class I, II, III, Div. 1, Groups A, B, C, D, E, F, and G, Class I, Zone 0, AEx ia IIC T4 when installed per Control Drawing: 9240 030-936. Non-Incendive Class I, Div. 2, Groups A, B, C, D, F, a...
Page 27 - European certifications; ATEX approvals; Intrinsically safe and FISCO model:
Quick Installation Guide 27 March 2013 European certifications ATEX approvals Special conditions for safe use (X) The intrinsically safe circuits do not withstand the 500 V AC test as specified in IEC 60079-11 clause 6.4.12. Impact and friction hazards need to be considered according to EN 60079-0 c...
Page 28 - Brazilian certifications; INMETRO approvals
March 2013 28 Quick Installation Guide Brazilian certifications INMETRO approvals Special conditions for safe use (X) Letter X in the number of the certificate indicates the following special condition(s) for safe use: For models 530xFxxxxxxxxxE1..., 530xFxxxxxxxxxKA..., 530xFxxxxxxxxxKB... or 530...
Page 29 - IB; Chinese certifications; National Supervision and Inspection Center for Explosion
Quick Installation Guide 29 March 2013 I2 1 Intrinsically safe: 4-20 mA/HART model: Ex ia IIC T4 GaEx ta IIIC T79 °C-50 °C < T a < +70 °C U i : 30 V, I i : 130 mA, P i : 1.0 W, L i : 0 μ H, C i : 7.26 nF F OUNDATION fieldbus model: Ex ia IIC T4 GaEx ta IIIC T69 °C-50 °C < T a < +60 °C U ...
Page 30 - fieldbus FISCO model:; Japanese certifications; Technology Institution of Industrial Safety (TIIS) approval
March 2013 30 Quick Installation Guide IC 1 F OUNDATION fieldbus FISCO model: Ex ia IIC T4 (-50 °C < T a < +60 °C) DIP A20 T A 69 °C U i = 17.5 V, I i = 380 mA, P i = 5.32 W, C i 4.95 nF, L i 0.1 μ H N3 Type n: HART model: Ex nL IIC T4 (-50 °C < T a < +70 °C) U i = 30 V, I i = 130...
Page 31 - IECEx certifications; IECEx approvals
Quick Installation Guide 31 March 2013 IECEx certifications IECEx approvals Special conditions for safe use (X) The intrinsically safe circuits do not withstand the 500 V AC test as specified in IEC 60079-11 clause 6.4.12. Impact and friction hazards need to be considered according to EN 60079-0 cla...
Page 32 - Other certifications; Overfill protection; Suitability for intended use
March 2013 32 Quick Installation Guide Other certifications Overfill protection Cert. no: Z-65.16-476 U1 TÜV-tested and approved by DIBt for overfill protection according to the German WHG regulations Approval valid for HART and F OUNDATION fieldbus options. Suitability for intended use Compliant wi...
Page 33 - EC Declaration of Conformity; Figure 1. EC Declaration of Conformity - page 1
Quick Installation Guide 33 March 2013 EC Declaration of Conformity Figure 1. EC Declaration of Conformity - page 1
Page 34 - Figure 2. EC Declaration of Conformity - page 2
March 2013 34 Quick Installation Guide Figure 2. EC Declaration of Conformity - page 2
Page 35 - Figure 3. EC Declaration of Conformity - page 3
Quick Installation Guide 35 March 2013 Figure 3. EC Declaration of Conformity - page 3