Page 2 - The electrical wiring for AC power; INSPECTION ON DELIVERY
Vista ™ basic Instructions for Use 1/25 950787 Rev H 02/03 The CE mark confirms compliance with the “Council Directive on Medical Products 93/42/EEC” dated the 14 th of June, 1993. B. Braun Melsungen AG RESPONSIBILITY OF MANUFACTURER Manufacturer, assembly, and installation personnel or instructors ...
Page 3 - CONTENTS; Program Mode
Vista ™ basic Contents Vista ™ basic Instructions for Use 2/25 950787 Rev H 02/03 CONTENTS Warnings & Cautions.................................................................. 3 Introduction ................................................................................ 5 Preparing the Infusio...
Page 4 - WARNINGS
WARNINGS & CAUTIONS Vista ™ basic Instructions for Use 3/25 950787 Rev H 02/03 The following should be observed to avoid patient/user injury. This Instructions for Use manual contains detailed instructions and warnings on the use of the Vista ™ basic Infusion Pump. Please read it completely prio...
Page 5 - SAFETY STANDARD
WARNINGS & CAUTIONS Vista ™ basic Instructions for Use 4/25 950787 Rev H 02/03 ♦ After starting an infusion, make sure drops are falling in the drip chamber. If no drops are falling, make sure the roller clamp is open. If the roller clamp is open and still no drops are falling, replace and dispo...
Page 6 - INTRODUCTION; for carrying
INTRODUCTION Vista ™ basic Instructions for Use 5/25 950787 Rev H 02/03 Handle for carrying Display All essential information is displayed in plain language. Green background lighting illuminates only when connected to AC power, or, in battery operation, only when a key is pressed. Keyboard Allows d...
Page 7 - Door Open Button
INTRODUCTION Vista ™ basic Instructions for Use 6/25 950787 Rev H 02/03 Door Open Button Aluminum Case Easy upkeep, drip proof, and resistant to disinfectants Function Keys Allow choices of YES, NO, OK, INFO, OPTS, etc. ON/OFF (Power) Key Alarm Silence Key Silences alarm tone for 2 minutes START/STO...
Page 8 - PREPARING THE INFUSION; Prepare the Pump
PREPARING THE INFUSION Vista ™ basic Instructions for Use 7/25 950787 Rev H 02/03 Prepare the Pump 1. Check for cleanliness, completeness, and damage. If unit appears damaged, do not use it. 2. Attach the pump to an IV pole by turning the knob on the pole clamp clockwise, or place the pump on a flat...
Page 10 - VTBD and TIME; Enter infusion time in hours and minutes.; pump calculates the rate and asks:; RATE and TIME; to stop the infusion.
Continuous Mode Vista ™ basic Instructions for Use 9/25 950787 Rev H 02/03 chamber. VTBD and TIME 1. Press VTBD. 2. Enter bag volume in ml. 3. Press OK. (To leave screen without saving changes, press EXIT.) 4. Press TIME. 5. Enter infusion time in hours and minutes. N OTE : For a 2 hour infusion, en...
Page 11 - INFORMATION KEY
Continuous Mode Vista ™ basic Instructions for Use 10/25 950787 Rev H 02/03 KEEP VEIN OPEN RATE After an infusion is complete, the pump will alarm and begin pumping at 1.0 ml/hr or the current infusion rate, whichever is less. 1. Stop the infusion. 2. Turn off the pump or reset the program. INFORMAT...
Page 12 - OPTIONS Piggyback; PIGGYBACK
OPTIONS Piggyback Vista ™ basic Instructions for Use 11/25 950787 Rev H 02/03 The Vista ™ basic contains many options to allow custom programming. To access the various options, press OPTS while the pump is stopped. Pressing OPTS each time will move you to the next option. Eventually, the OPTS key w...
Page 13 - Mode; ote; To Review or Change Ramp/Taper Mode Data
OPTIONS Piggyback, Ramp/Taper Mode Vista ™ basic Instructions for Use 12/25 950787 Rev H 02/03 To Exit Piggyback 1. Press to stop the infusion. 2. Press END. 3. Press OPTS. 4. Press OFF to turn Piggyback off 5. Press END to return to Continuous Mode. RAMP/TAPER MODE The Ramp/Taper Mode is designed t...
Page 14 - To Exit Ramp/Taper Mode; to start the infusion. Check for drops in the drip chamber.
OPTIONS Ramp/Taper Mode Vista ™ basic Instructions for Use 13/25 950787 Rev H 02/03 To Clear Ramp/Taper Mode 1. Press OPTS until display shows RAMP TAPER MODE. 2. Press CLR. 3. Press YES to clear all the data in Ramp/Taper Mode. 4. Press ON and continue to program a new Ramp/Taper program. To Exit R...
Page 15 - PROGRAM MODE; To Review or Change Program Mode data; To Exit Program Mode
OPTIONS Program Mode Vista ™ basic Instructions for Use 14/25 950787 Rev H 02/03 PROGRAM MODE Program Mode is for infusions requiring up to 9 separate rates and volumes to be delivered one after another. This is designed to deliver infusions for stepped oncology drugs, circadian medication delivery,...
Page 16 - DELAY A MODE; To Exit Delay A Mode
OPTIONS Delay A Mode Vista ™ basic Instructions for Use 15/25 950787 Rev H 02/03 DELAY A MODE The start of Continuous, Piggyback, Ramp/Taper, and Program Modes may all be delayed. The Vista ™ basic also provides an option to extend the Infusion Complete alarm at the end of the infusion. 1. Program t...
Page 17 - DOSAGE CALCULATION
OPTIONS Dosage Calculation Vista ™ basic Instructions for Use 16/25 950787 Rev H 02/03 DOSAGE CALCULATION This feature allows a rate to be calculated given the concentration and dose. 1. Press OPTS until display shows DOSAGE CALC. 2. Press ON. 3. If concentration unit desired is other than the one d...
Page 18 - OPTIONS; BOLUS MODE
OPTIONS Bolus Mode, Interval Bolus (Intermittent Mode) Vista ™ basic Instructions for Use 17/25 950787 Rev H 02/03 BOLUS MODE This feature allows a bolus to be given during a Continuous Mode infusion. 1. Press OPTS until display shows BOLUS MODE. 2. Press ON. 3. Press RATE. 4. Enter the rate at whic...
Page 19 - Total Infused, Standby Function; TOTAL INFUSED; been pumped since the last time the Total Infused was cleared.; To Clear the Total Infused; The pump will count down to zero and then alarm; To Exit Standby Function; Press EXIT to return to Continuous Mode.
OPTIONS Total Infused, Standby Function Vista ™ basic Instructions for Use 18/25 950787 Rev H 02/03 TOTAL INFUSED The Total Infused information screen displays the total amount of fluid in ml that has been pumped since the last time the Total Infused was cleared. To Clear the Total Infused 1. Press ...
Page 20 - OCCLUSION PRESSURE; To Turn On the Drop Sensor; To Turn Off the Drop Sensor; Press OPTS until display shows DROP SENSOR.; BATTERY CAPACITY; Press END to return to Continuous Mode
OPTIONS Occlusion Pressure, Drop Sensor, Battery Vista ™ basic Instructions for Use 19/25 950787 Rev H 02/03 OCCLUSION PRESSURE The pressure at which a downstream occlusion alarm will be triggered may be varied. The higher the pressure limit, the less sensitive the pump is to changes in fluid resist...
Page 21 - DATA LOCK
OPTIONS Data Lock, Contrast Selection Vista ™ basic Instructions for Use 20/25 950787 Rev H 02/03 DATA LOCK This features prevents unauthorized use by locking out all keys except , ,and . 1. Press OPTS until display shows DATA LOCK. 2. Press ON. 3. Press END to return to Continuous Mode. To Turn Off...
Page 22 - ADJUST CLOCK
OPTIONS Adjust Clock, Loudness Vista ™ basic Instructions for Use 21/25 950787 Rev H 02/03 ADJUST CLOCK This feature allows the time and date to be changed. 1. Press OPTS until display shows ADJUST CLOCK. 2. Press DAT to enter the date. (Press C to clear the current data and re- enter new data.) 3. ...
Page 23 - OPTIONS Alarms; In alarm situations, press; Alarm Possible
OPTIONS Alarms Vista ™ basic Instructions for Use 22/25 950787 Rev H 02/03 In alarm situations, press to silence the alarm. Correct the problem. Resume the infusion by pressing . Alarm Possible Cause Action Air-In-Line! Air bubble in air-in-line sensor Remove air from system Check Drop Sensor! Free ...
Page 24 - Battery Operation
AC, 12 V, BATTERY, & CC INTERFACE Vista ™ basic Instructions for Use 23/25 950787 Rev H 02/03 AC and 12 V ♦ Check AC voltage as per rating label. ♦ Insert/screw on power lead (12 V lead in ambulance car) on back. Battery Operation ♦ Switching to internal battery is automatic in the event of powe...
Page 25 - Maintenance; Storage
MAINTENANCE, STORAGE, & SERVICE Vista ™ basic Instructions for Use 24/25 950787 Rev H 02/03 Maintenance Clean the pump with a soft, lint-free cloth or swab dampened with soap and water, a general non staining chemical disinfectant, isopropyl alcohol (15%), derivatives of phenylphenol (5%) or amm...
Page 26 - SPECIFICATIONS; Linear peristaltic volumetric infusion pump
SPECIFICATIONS Vista ™ basic Instructions for Use 25/25 950787 Rev H 02/03 Mechanism Linear peristaltic volumetric infusion pump Dimensions WxHxD 5.5 in (14cm) x 9.4 in (24cm) x 7.8 in (20cm) Weight 6.8 lbs. (3.1 kg) Power Requirements 100-120 VAC, 50-60 Hz, 13 Watts Leakage Current Max 30µA Battery...