1 Page 1
2 Page 2
3 Page 3
4 Page 4

Summary

Page 2 - Install the internal DMM circuit board into the chassis.; Position the internal; circuit board inside the instrument chassis; Lock the internal DMM circuit board in place.; Slide the internal; DMM; circuit board to the left (toward the rear of the

4 Install the internal DMM circuit board into the chassis. Position the internal DMM circuit board inside the instrument chassis by tilting it at an angle as shown below. Be sure to position the circuitboard over the tabs in the chassis as shown. Then, lower the assemblyinto the chassis until it is ...

Page 3 - Twist the internal DMM input cable.; Insert the DMM input cable into the backplane connector.; Enable the internal DMM.

6 Twist the internal DMM input cable. Twist the internal DMM input cable (labeled P101 ) at least one full turn before plugging it into the backplane connector. If you do not twist thecable as shown, certain specifications will be compromised. 7 Insert the DMM input cable into the backplane connecto...

Page 4 - Modify the Labeling on the Rear Panel.; through the black circle using a permanent marker as shown below.; 0 Replace the outer cover on the; Calibration Information; Internal DMM; Retrofit Instructions

9 Modify the Labeling on the Rear Panel. If you originally ordered the 34970A without the internal DMM, a solidblack circle should be visible above the rear-panel “Opt. 001” label. After installing the internal DMM retrofit kit, be sure to draw an “ X ” through the black circle using a permanent mar...