Agilent 34970-80010 - Manual

Agilent 34970-80010

Agilent 34970-80010 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
Page: / 4

Table of Contents:

  • Page 2 – Install the internal DMM circuit board into the chassis.; Position the internal; circuit board inside the instrument chassis; Lock the internal DMM circuit board in place.; Slide the internal; DMM; circuit board to the left (toward the rear of the
  • Page 3 – Twist the internal DMM input cable.; Insert the DMM input cable into the backplane connector.; Enable the internal DMM.
  • Page 4 – Modify the Labeling on the Rear Panel.; through the black circle using a permanent marker as shown below.; 0 Replace the outer cover on the; Calibration Information; Internal DMM; Retrofit Instructions
Loading the manual

1 Disconnect the power cord.

2 Remove the outer cover from the

Agilent 34970A.

To remove the cover from the instrument, follow the procedures shown
in chapter 6 of the 34970A Service Guide.

3 Connect the internal DMM ribbon cable to the main circuit board.

Connect the loose end of the grey ribbon cable to connector

J202

on the

34970A main circuit board (34970-66511). Press down firmly on the
ribbon cable connector and secure the cable in place with the locking clips.

Snap the Locking Clips over
the top of the connector to
secure the Ribbon Cable

Ribbon Cable

1

** Corner Staple

** Trim to 7.5 x 9 inches

** Double Sided

Agilent 34970A / Option 001
Internal DMM Assembly

Installation Instructions for 34970-80010 Retrofit Kit

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - Install the internal DMM circuit board into the chassis.; Position the internal; circuit board inside the instrument chassis; Lock the internal DMM circuit board in place.; Slide the internal; DMM; circuit board to the left (toward the rear of the

4 Install the internal DMM circuit board into the chassis. Position the internal DMM circuit board inside the instrument chassis by tilting it at an angle as shown below. Be sure to position the circuitboard over the tabs in the chassis as shown. Then, lower the assemblyinto the chassis until it is ...

Page 3 - Twist the internal DMM input cable.; Insert the DMM input cable into the backplane connector.; Enable the internal DMM.

6 Twist the internal DMM input cable. Twist the internal DMM input cable (labeled P101 ) at least one full turn before plugging it into the backplane connector. If you do not twist thecable as shown, certain specifications will be compromised. 7 Insert the DMM input cable into the backplane connecto...

Page 4 - Modify the Labeling on the Rear Panel.; through the black circle using a permanent marker as shown below.; 0 Replace the outer cover on the; Calibration Information; Internal DMM; Retrofit Instructions

9 Modify the Labeling on the Rear Panel. If you originally ordered the 34970A without the internal DMM, a solidblack circle should be visible above the rear-panel “Opt. 001” label. After installing the internal DMM retrofit kit, be sure to draw an “ X ” through the black circle using a permanent mar...