Page 2 - Contents; Commissioning
Contents ACT 250 addendum Alcatel Vacuum Technology France - ACT 250 addendum 1 January 2001 Commissioning Presentation ■ The ACT 250 controller . . . . . . . . . . . . . . ■ p. 3 ■ Main characteristics . . . . . . . . . . . . . . . . ■ p. 3 ■ Accessories . . . . . . . . . . . . . . . . . . . . . . ...
Page 3 - Use; Controlling the pump from the
Contents ACT 250 addendum Alcatel Vacuum Technology France - ACT 250 addendum 2 January 2001 Use Commissioning (continued) ■ Controlling the pump from the ACT 250 controller . . . . . . . . . . . . . . . . . ■ p. 24 ■ Controller functions . . . . . . . . . . . . . . . . . ■ p. 26 ■ ■ Precautions ■ ■...
Page 4 - Presentation; The ACT 250 controller; Main characteristics
Presentation Alcatel Vacuum Technology France - ACT 250 addendum 3 January 2001 The ACT 250 controller Dear Customer,You have just bought anACT 250 controller. Inorder to ensure the bestpossible performance of theequipment and yourcomplete satisfaction inusing it, we advise you toread this addendumc...
Page 5 - Accessories; Pump power supply; Kit
Presentation Alcatel Vacuum Technology France - ACT 250 addendum 4 January 2001 Accessories Pump power supply cable The controller is connected to the pump using acorresponding power supply cable, ordered separately. Cable length (m) Part No. 1 105086 1.5 A458885 3.5 101812 5 101810 10 101811 15 105...
Page 6 - Local control mode; Remote control; The remote control
Presentation Alcatel Vacuum Technology France - ACT 250 addendum 5 January 2001 ACT 250 “box” controller control modes Local control mode A C T 2 5 0 A C T 2 5 0 POWER FAULT SPEED START STOP 1 2 Pump start-up switch Pump stop switch Pump status indicators Controller power and fault indicators Remote...
Page 8 - Characteristic; dimensions
Presentation Alcatel Vacuum Technology France - ACT 250 addendum 7 January 2001 ACT 250 controller technical characteristics Characteristic Unit ACT 250 “box” ACT 250 “OEM” Weight kg 1.8 1.3 Dimensions H x W x D mm 128.4 × 106.3 × 220 95 × 100 × 212 3U × 1/4 Rack Nominal voltage V 85 - 265 Frequency...
Page 10 - Safety instructions
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 9 January 2001 Safety instructions Before switching on the controller, the user shouldstudy the manual and follow the safetyinstructions listed in the compliance certificate booklet supplied with the pump.See sheet B10 in the ATP pump...
Page 11 - If the controller is remote controlled, make the various
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 10 January 2001 ACT 250 controllerelectrical connections Connect the RS232/485 serial link cable to the connector (cable and PCsupplied by customer) ■ If the controller is remote controlled, make the various connections on the REMOTE ...
Page 12 - compliance with safety regulations.; Voltage control mode; Select Mode
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 11 January 2001 ACT 250 remote connector characteristics When units containing the control circuits areequipped with dr y contact outputs, it is thecustomer's responsibility to use the outputs in compliance with safety regulations. Vo...
Page 13 - Signalling on output; These are dry contacts
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 12 January 2001 ACT 250 remote connector characteristics Signalling on output contacts These are dry contacts (48 V AC - 1 A) which replicate pump status information. When the output contact is closed: 1 6 2 7 3 8 4 9 1 - 6 The pump h...
Page 14 - Use in local mode; Without galvanic isolation; Wiring seen from solder side
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 13 January 2001 ACT 250 remote connector wiring Use in local mode ■ Standard connector plug (factor y wired) Use in remote control mode ■ Without galvanic isolation (not recommended) Wiring seen from solder side. ■ With galvanic isola...
Page 15 - Configuring the RS 232; Switch
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 14 January 2001 RS 232 / 485 serial link Wiring the RS 232 or RS 485 serial link See sheet B 110 in the ATP Series Pump User Manual. Configuring the RS 232 or RS 485 serial link The link is configured using switches on the rear panel ...
Page 16 - Detailed description of RS commands; Syntax conventions; : command executed correctly
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 15 January 2001 Detailed description of RS commands Syntax conventions applicable to all commands Status values Error messages ADR Syntax Result adr = address, from 000 to 255 <CR> Carriage Return (ascii 13) <LF> Line Feed...
Page 17 - CKS; for CKSON
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 16 January 2001 Detailed description of RS commands CKS Syntax Result CYC Syntax Result Enables or disables checksums on reply strings #adr CKS ON<CR>[<LF>] Enables the ASCII checksum character at the end of a reply string...
Page 18 - DLI
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 17 January 2001 Detailed description of RS commands DLI Syntax Result See also: DLR DLR Syntax Result See also: DLI, LNG, SEP, SHT Enables DataLogger operation (RS232 only) #adr DLR <CR>[<LF>] #adr,sssss,nnnnn,iiii,ttttt,u...
Page 19 - HDR; if error; or
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 18 January 2001 Detailed description of RS commands HDR Syntax Result IDN Syntax Result Identifies the device which is communicating, and itssoftware version #adr IDN <CR>[<LF>] #adr, VS.... - Vx.zz’or#adr, VS.... - Vx.zz ...
Page 20 - LEV; NSP
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 19 January 2001 Detailed description of RS commands LEV Syntax Result LNG Syntax Result See also: SHT Returns the state of the parameters defined by SET #adr LEV <CR>[<LF>] #adr,nnnnn,sssss,aaaa,hhhhh or #adr,nnnnn rpm,sss...
Page 21 - OPT; RPM; SAV
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 20 January 2001 Detailed description of RS commands OPT Syntax Result See also: SEL Used to select possible user choices #adr OPT2 n<CR>[<LF>] choice of temperature unit: n = 0 : degrees Centigrade n = 1 : degrees Fahrenhe...
Page 22 - SBY; SEL; SEP
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 21 January 2001 Detailed description of RS commands SBY Syntax Result See also: NSP, RPM Switches the speed set point to the stand-by value #adr SBY <CR>[<LF>] #adr,ok Resets the stand-by speed to its last stored value, an...
Page 23 - SET
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 22 January 2001 Detailed description of RS commands SET Syntax Result See also: LEV SHT Syntax Result See also: LNG SPD Syntax Result See also: LNG , SHT Defines the internal operating parameters #adr SET 1 hhhhh<CR>[<LF>]...
Page 24 - STA; ” character at the start of the reply
Commissioning Alcatel Vacuum Technology France - ACT 250 addendum 23 January 2001 Detailed description of RS commands STA Syntax Result TMP Syntax Result Returns the status of the internal dynamic parameters #adr STA <CR>[<LF>] #adr,xxxxxx,yyyyyy,zzzzzz,sssss,iiii,www,ppp,vvv,ttttt<CR...
Page 25 - Indicator lights
Use Alcatel Vacuum Technology France - ACT 250 addendum 24 January 2001 Controlling the pump from the ACT 250controller Once all connections are complete, plug the controller intothe mains. A C T 2 5 0 A C T 2 5 0 POWER FAULT SPEED START STOP Indicator lights “box” version “OEM” version Yellow lit C...
Page 26 - START; Starting the pump with
Use Alcatel Vacuum Technology France - ACT 250 addendum 25 January 2001 Controlling the pump from the ACT 250controller START STOP Starting the pump with the START switch Stopping the pump with the STOP switch The pump is started up to reach the selected speed. Theyellow rising speed indicator light...
Page 27 - Controller functions; Precautions; Indication of rotation
Use Alcatel Vacuum Technology France - ACT 250 addendum 26 January 2001 Controller functions Precautions Local mode operation The Start and Stop functions use switches located on thefront panel of the “box” version and on the board for the“OEM” version. The “OEM” version of the ACT 250 has beendesig...
Page 28 - Fault monitoring; Alerts
Use Alcatel Vacuum Technology France - ACT 250 addendum 27 January 2001 Controller functions Fault monitoring ■ Alerts are indicated by: - flashing red indicator. The pump power supply ismaintained. ■ Faults are indicated by: - lit red indicator. The pump is stopped as soon as a faultis detected. Al...