Page 2 - TABLE OF CONTENTS; Before using the monitor
TABLE OF CONTENTS Before using the monitor Introduction .............................................................................................................3 Safety information ...................................................................................................4 Intended use....
Page 3 - INTRODUCTION; Fill in for future reference.; SAVE THESE INSTRUCTIONS
INTRODUCTION Thank you for purchasing the ReliOn TM BP200 Blood Pressure Monitor, Model # HEM-741CRELN4. Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the mov...
Page 4 - SAFETY INFORMATION; INTENDED USE
SAFETY INFORMATION INTENDED USE This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with the readings. To assure the correct u...
Page 5 - GENERAL USAGE
5 SAFETY INFORMATION GENERAL USAGE Always consult your physician. Self-diagnosis of measurement values and self-treatment are dangerous. Consult your physician before using the device for any of the following conditions: • If you have had a mastectomy. • People with severe blood flow problems or blo...
Page 6 - GENERAL PRECAUTIONS
6 SAFETY INFORMATION GENERAL PRECAUTIONS • Do not forcibly crease the arm cuff or the air tube excessively. • Do not press the air tube while taking a measurement. • To unplug the air plug, pull on the air plug at the connection with the monitor, not the tube itself. • Do not drop the monitor or sub...
Page 7 - BEFORE TAKING A MEASUREMENT
7 BEFORE TAKING A MEASUREMENT To ensure a reliable reading follow these recommendations: 1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes before taking a measurement. Rest for at least 5 minutes before taking the measurement. 2. Stress raises blood pressure. Avoid t...
Page 8 - KNOW YOUR UNIT
8 KNOW YOUR UNIT Main Unit: Components: Quick Start Guide Instruction Manual Air plug Air tube Wide Range Cuff (# CD-WR17)(Model: HEM-RML31) Arm Cuff (Arm circumference9” - 17” (22 - 42 cm)) Date/Time setting button Memory button Start/Stop button "AA" Alkaline Batteries ( / ) Air jack Displ...
Page 9 - UNIT DISPLAY
UNIT DISPLAY Average reading symbol User symbol (A/B) Battery symbol (low/depleted) Blood pressure level indicator(High/Normal)Date/Time display Memory symbol Diastolic blood pressure Systolic blood pressure Irregular heartbeat symbol Movement error symbol Heartbeat symbol (Flashes during measuremen...
Page 10 - DISPLAY SYMBOLS
DISPLAY SYMBOLS IRREGULAR HEARTBEAT SYMBOL ( ) When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Heartbeat Symbol will appear on the display with the measurement values.An irregular heartbeat rhythm is defined as a rhythm that varies by 25% less or ...
Page 11 - BLOOD PRESSURE LEVEL INDICATOR ( )
DISPLAY SYMBOLS BLOOD PRESSURE LEVEL INDICATOR ( ) If your Systolic Blood Pressure is 135 mmHg or above and/or the Diastolic Blood Pressure is 85 mmHg or above, the High Blood pressure level indicator ( ) will display when the measurement value is displayed. If the measurements are within the standa...
Page 12 - BATTERY INSTALLATION
BATTERY INSTALLATION 1. Press the hook on the top of the battery cover and lift the cover off in the direction of the arrow. 2. Install 4 "AA" size batteries so the + (positive) and - (negative) polarities match the polarities of the battery compartment as indicated. Battery Cover 3. Replace...
Page 13 - SETTING THE DATE AND TIME
SETTING THE DATE AND TIME Set the monitor to the correct date and time before taking a measurement for the first time or after replacing the batteries. 1. Press the button to show the year. Year Month Day Hour Minute Change Confirm Change Confirm Change Confirm Change Confirm Change Confirm : Forwar...
Page 14 - APPLYING THE CUFF ON THE LEFT ARM; APPLYING THE ARM CUFF; Left Arm
APPLYING THE CUFF ON THE LEFT ARM 1. Make sure the air plug is securely inserted in the main unit. 2. Remove tight-fitting clothing from your upper arm. 3. Apply the cuff to your left upper arm so the air tube is centered on the inside of your arm and aligned with your middle finger. The air tube ru...
Page 15 - APPLYING THE CUFF ON THE RIGHT ARM; Right Arm
APPLYING THE ARM CUFF APPLYING THE CUFF ON THE RIGHT ARM When taking a measurement using the right arm, use this instruction for Step 3. Apply the cuff to your right upper arm so the air tube runs along the inside of your arm. The bottom of the cuff should be approximately 1/2 inch (1 - 2 cm) above ...
Page 16 - TAKING A MEASUREMENT
TAKING A MEASUREMENT A B The monitor is designed to store the blood pressure and the pulse rate in the memory for two people (USER A or USER B) every time a measurement is completed. 1. Sit on a chair with your feet flat on the floor. Place your arm on a table so the cuff is level with your heart. R...
Page 17 - If your systolic pressure is more than 210 mmHg
3. Press the Start/Stop button. All symbols appear on the display. The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically determines your ideal inflation level. Remain still and do not move your arm until the entire measurement process is completed. NOTE: To stop t...
Page 19 - USING THE MEMORY FUNCTION; TO VIEW THE MEASUREMENT VALUES STORED IN MEMORY
: To view the older values : To view the more recent values 3. Press the or button to view the values stored in the memory. 4. Press the Start/Stop button to turn the monitor off. • If there are no measurement values stored in the memory, the screen to the right is displayed. < > > USING TH...
Page 20 - TO DELETE ALL VALUES STORED IN THE MEMORY
USING THE MEMORY FUNCTION TO DELETE ALL VALUES STORED IN THE MEMORY The values stored in the memory are deleted by USER ID. You cannot partially delete values stored in the memory. All values for the user you select will be deleted. 1. Select your USER ID (A or B). 2. Press the button to turn the mo...
Page 21 - CARE AND MAINTENANCE; Do not forcefully
CARE AND MAINTENANCE To keep your blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed below: Do not forcefully bend the arm cuff or air tube. Do not fold tightly. Clean the monitor with a soft dry cloth. Do not use any abrasive or volatile clea...
Page 22 - ERROR INDICATORS; SYMBOL
ERROR INDICATORS Air plug disconnected. Arm cuff not applied correctly.Air is leaking from the arm cuff. Cuff over or under inflated. Batteries are depleted. Insert the plug securely. Refer to "APPLYING THE ARM CUFF". Irregular heartbeats are detected. Remove the arm cuff. Wait 2-3 minutes a...
Page 24 - TROUBLESHOOTING TIPS; PROBLEM
24 TROUBLESHOOTING TIPS PROBLEM No power. No display appears on the unit. Measurement values appear too high or too low. CAUSES AND SOLUTIONS Replace all four batteries with new ones. Check the battery installation for proper placement of the battery polarities.Blood pressure varies constantly. Many...
Page 25 - FCC STATEMENT; FCC CAUTION
FCC STATEMENT FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of th...
Page 26 - GUIDANCE AND MANUFACTURER’S DECLARATION; ReliOn Blood Pressure Monitor
GUIDANCE AND MANUFACTURER’S DECLARATION ReliOn Blood Pressure Monitor Model: HEM-741CRELN4 Information for accompanying documents in the scope of IEC60601-1-2:2007 Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC’s and mo...
Page 27 - Emissions test; Electromagnetic environment; Compliance
GUIDANCE AND MANUFACTURER’S DECLARATION Guidance and manufacturer’s declaration – electromagnetic emissions Emissions test Harmonic emissions IEC 61000-3-2Voltage fluctuations/ flicker emissions IEC61000-3-3 ReliOn TM HEM-741CRELN4 is intended for use in the electromagnetic environment specified bel...
Page 28 - Immunity test
GUIDANCE AND MANUFACTURER’S DECLARATION Electrostatic discharge (ESD) IEC 61000-4-2 ±6 kV contact ±8 kV air ±6 kV contact ±8 kV air Floor should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. Electrical fast transient...
Page 32 - Output Power
80 MHz to 800 MHz 800 MHz to 2.5GHz 150 kHz to 80 MHz Not Applicable. d = 1.2 P d = 2.3 P Recommended separation distance between portable and mobile RF communications equipment and the ReliOn TM HEM-741CRELN4 ReliOn TM HEM-741CRELN4 is intended for use in an electromagnetic environment in which rad...
Page 33 - LIMITED WARRANTY
LIMITED WARRANTY Your ReliOn TM BP200 Blood Pressure Monitor, Model # HEM-741CRELN4, excluding the arm cuff and batteries, is warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase, when used in accordance with the instructions provided with...
Page 34 - FOR CUSTOMER SERVICE
LIMITED WARRANTY This warranty provides you with specific legal rights, and you may have other rights which vary from state to state. Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you. Some states do not allow the exclusion or li...
Page 35 - SPECIFICATIONS; NOTE: These specifications are subject to change without notice.
SPECIFICATIONS Model:Display:Measurement Range:Accuracy/Calibration: Inflation:Deflation:Rapid Air Release:Measurement Method:IP classification: Power Source:Battery Life:Operating Temperature /Humidity:Storage Temperature /Humidity/Air Pressure:Main Unit Weight: Main Unit Dimensions: Cuff Size: Cuf...