Table of Contents:

  • Page 4 – ROUTINE ALIGNMENT PROCEDURES FOR PHILIPS CM-20 TEM; Tilt indicator is at 0; Press MODES and TEM to enter the TEM bright field page.; III; (to apply filament current) after the HT
  • Page 5 – DO NOT CHANGE; (5) Check EMISSION current. It should be somewhere between 5 to 8
  • Page 6 – (B) Center Filament Image and Alignment of Condenser System:; the filament image.
  • Page 8 – (iii) Intermediate and Project Lens Alignment
  • Page 10 – Final Comments; Happy Hunting and Enjoying the Small World!
Loading the manual

OPERATION PROCEDURES FOR PHILIPS/FEI CM-20

TRANSMISSION ELECTRON MICROSCOPE

General Comments


Normally the microscope will be left in the power ON (standby ready to operate)

condition; i.e., all circuits will be continuously energized and the microscope will be under full
vacuum (MAIN, VACUUM, HIGH TENSION lamps on the right control panel are all lit if the
PANEL DIM is pressed). If it is not in this condition do not operate the microscope. Seek help
from authorized personnel.

If the microscope is in standby ready condition, proceed with the following abbreviated

procedures for high voltage application and alignment before beginning any observation. If the
microscope is not properly aligned, good results will not be obtained.

I.

Start-Up (This procedure is when the microscope starts from cold)


(1)

Run the cooling water (for vacuum and lens systems).


(2)

Turn on the main power line (in mechanical room, behind the microscope room).

(3)

Press the ON power switch on MICROSCOPE. The microscope will go through a series
of self check by the computer program. It will automatically turn on VACUUM
SYSTEM, and reach to the high vacuum state as indicated by the UHV and HI VAC
lights.

II.

Specimen Exchange


(i) Specimen Unloading:

(1)

Make sure specimen position is at center, i.e., SHIFT (right at 0, left at 10), TILT
indicator at 0 and 11 (HOLDER/azimuth) and tilt drive is engaged.

(2)

From holding position (B): Pull out the specimen holder gently until it no longer moves
and then turn it clockwise about 180 degree until it stops (A position, make sure it no
longer turns, the indicator box face down).

(3)

Gently extract the holder from the airlock.

(4) Place the specimen holder onto a proper sample holder stand/support for specimen

loading/unloading.

(ii) Specimen Loading:


(1)

Set specimen onto suitable specimen holder.

USE GLOVES! Caution:

No direct

contact of your hand with a specimen holder and any other part which will be placed
inside the microscope.


WAC, 09/27/02

1

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 4 - ROUTINE ALIGNMENT PROCEDURES FOR PHILIPS CM-20 TEM; Tilt indicator is at 0; Press MODES and TEM to enter the TEM bright field page.; III; (to apply filament current) after the HT

ROUTINE ALIGNMENT PROCEDURES FOR PHILIPS CM-20 TEM The following protocol is not intended to be specific. It is intended to be used as a mean to refresh one’s memory in carry out step by step procedures which are necessary to obtain reasonably good results with the transmission electron microscopes....

Page 5 - DO NOT CHANGE; (5) Check EMISSION current. It should be somewhere between 5 to 8

(6) Set voltage at 20. Push HIGH TENSION button (the green LED should light) Slowly increase voltage from 20 kV to 160 kV using soft keys (wait 1 minute between each step). (7) After voltage reaches 160 kV, press FREE CONTROL soft key. Use focus knob, slowly increase voltage from 160kV to 200kV. If ...

Page 6 - (B) Center Filament Image and Alignment of Condenser System:; the filament image.

by first making crossover image with INTENSITY knob, then open up condenser II (INTENSITY)/expand the beam and close condenser II on the opposite side of crossover. If the beam appears to swing this indicates the aperture is not properly centered. Center the aperture by moving the X and Y controls o...

Other Philips Models