Page 3 - RISK OF ELECTRIC SHOCK DO NOT OPEN.
The lightning flash with arrowhead symbol, within an equilateral triangle, isintended to alert the user to the presence of uninsulated "dangerous voltage"within the product's enclosure that may be of sufficient magnitude to constitute arisk of electric shock. The exclamation point within an ...
Page 4 - CONTENTS
1 1 1 1 1 1 CONTENTS .................................................................. 1 2 2 2 2 2 PRECAUTIONS ....................................................... 2 - 4 3 3 3 3 3 UNPACKING ................................................................. 5 4 4 4 4 4 FEATURES AND FUNCTIONS Front...
Page 5 - PRECAUTIONS; INSTALLATION LOCATIONS; PROTECTIVE MEASURES
2 2 PRECAUTIONS MAINTAIN GOOD VENTILATIONVentilation slots and holes are provided on sides of thisunit.Place the unit on a hard and level surface and locate atleast 4" (10cm) from walls to ensure proper ventilation. UNSUITABLE LOCATIONSAvoid shaky places or hot-springs areas where hydrogensulfid...
Page 7 - TECHNICAL DESCRIPTION
4 SAFETY TECHNICAL CHECKS Periods: According to the recommendations of the manufacturer of medical device.Scope: a) Visual check Housing, cables, operator controls, readout device ( displays, LED etc.), labels,accessories, instruction manual. b) Function test Performance check acc. instruction manua...
Page 8 - UNPACKING; Open the top of the box.; Take the unit out of the box carefully.; Accessories
5 UNPACKING 3 1 1 1 1 1 Open the top of the box. 2 2 2 2 2 Remove the cushion above the unit. • Make sure to keep the unithorizontal. 3 3 3 3 3 Take the unit out of the box carefully. Take the unit out of the box by the following procedures. Make sure to check thecontents. 4 4 4 4 4 Unwrap the unit....
Page 9 - FEATURES AND FUNCTIONS; Front Panel; Function; A B C
6 4 FEATURES AND FUNCTIONS OPEN PRINT COPY FEED FUNC CONT BRT ADJUST BRT CONT EXP. SIZE MODE SIDE MULTI NOR 1:1 SINGLE EXP. E 5 Front Panel 12 • 31 Power switch Turns on/off the power. 1 3 Function Name Reference Page 2 Selects the size of images to print. MODE switch Selects the extended function o...
Page 10 - Rear Panel; POTENTIAL EQUALIZATION; G H; Name; Reference
7 AC LINE REMOTE ON OFF VIDEO IN OUT DIP SW DIP SW FUNCTION TABLE NO. 1 2 3 4 5 6 FUNCTION IMP TRAP MEMORY RESERVED SW-ON 75 Ω ON FIELD SW-OFF HIGH OFF FRAME Rear Panel POTENTIAL EQUALIZATION TERMINAL This is used to equalize the potential of theequipment connected to this unit.For details refer to ...
Page 11 - INSTALLATION OF PAPER; Pull out the paper end.
8 5 INSTALLATION OF PAPER Moisture, fingerprints or dust on thepaper surface may cause a noise atprinting or deterioration in printquality. Set the paper by the follow-ing procedure to prevent adhesionof fingerprint or dust on the papersurface. Note: The printing surface is the outside. Place the pa...
Page 12 - CAUTION; Do not use defective paper.; INCORRECT; Adjust the paper position correctly.
9 • Keep the high-density paper away from fingerprint, dust or moisture when storing it.• Do not touch the rubber roller. Do not stain or damage the roller surface.• Do not touch the thermal head (located behind the cutter). When printing, the thermal head is heated to high temperature. • Do not tou...
Page 13 - EXAMPLE OF CONNECTION / SETTING OF SWITCHES; Composite Video Signal; Setting of Switches
10 6 EXAMPLE OF CONNECTION / SETTING OF SWITCHES Connecting to various composite video signal equipment suchas medical equipment. 1 Turn off the power switches of the Video copy processor andthe equipment to be connected. Composite Video Signal 2 Connect the VIDEO input terminal of the Video copypro...
Page 14 - Medical Video Signal
11 Medical Video Signal 1 Turn off the power switches of the Video copy processor andthe equipment to be connected. 2 Connect the VIDEO input terminal of the Video copyprocessor to the video output terminal of the connectedequipment. Rear panel Medicalequip-ment AC LINE REMOTE ON OFF VIDEO IN OUT DI...
Page 15 - PRINTING; Printing Procedure; Cut the printed paper.; Paper Feeding
12 7 PRINTING Printing Procedure • Display a picture to be printed on themonitor screen, andpress the PRINTbutton. When printingfinishes, abuzzer tone willbe heard. COPY FEED cutter 2 You can set the number of copies bypressing the COPY button on the frontpanel.You can copy the same picture asmany t...
Page 16 - Use of Remote Control; Precautions on Printing; Paper saving mode
13 Use of Remote Control Connect the wired remote control tothe remote control terminal on therear panel.Press the button on the remotecontrol to print pictures.This button functions the same as thePRINT button.When no signal is input, the button isinvalid. To remote control terminal Precautions on ...
Page 17 - ADJUSTMENT OF PRINT PICTURE; Adjustment of Brightness/Contrast; Press the buttons to adjust brightness or contrast.
14 8 ADJUSTMENT OF PRINT PICTURE Adjustment of Brightness/Contrast You can adjust brightness and contrast of the printed picturewhile observing the monitor screen. Control panel COPY FEED FUNC CONT BRT ADJUST BRT CONT EXP. SIZE MODE SIDE MULTI NOR 1:1 SINGLE EXP. • For adjustment, use the BRT button...
Page 19 - SETTING FUNCTION MODE; Function mode
16 9 SETTING FUNCTION MODE Function mode In this mode, the initial setting value of each function can bechanged. Each time the FUNC button is pressed, the mode isswitched as follows. Press the FUNC button for less than 1 second more than 1 second curve setting Selection of the memorized image for co...
Page 20 - Indicator; Setting the function mode; curve setting; Purpose and description
17 Indicator Setting the function mode γ curve setting Purpose and description • To select the gamma ( γ ) curve (relation between the density and the brightness of the image) to obtain an optimum densitydepending on the connected device. Five options are available. • The default setting is 5. Indic...
Page 25 - SIZE switch setting; SIZE
22 SIZE switch setting The size of the printed image can be selected by the SIZEswitch. SIZE SIDE NOR 1:1 Example A A A SIDE NOR. 1:1
Page 26 - MODE; To print a multi image, press the PRINT button twice.
23 MODE switch setting MODE switch works as the extended function of SIZE switch. MODE MULTI SINGLE EXP. Set this switch to SINGLE for normal use. The image is printed in the size specified by theSIZE switch. When selecting MULTI, you can print 2-multi images. • To print a multi image, press the PRI...
Page 27 - A B
24 Print patterns depending on the combination of the settings ofthe SIZE and MODE switches B A A B A A A A Printing is not available. Printing is not available. SIZE MODE SIDE NOR 1:1 SINGLE MULTI EXP.
Page 29 - Overheat; ERROR DISPLAY; COPY button
26 10 In case of an error in the unit during operation, you are warnedby an alarm tone or the LED indicator. Cause/Error display 1 1 1 1 1 Overheat ERROR DISPLAY Symptom/Remedy [Symptom]• When the head gets overheated, the indicator blinks. In this case, the following buttons function as described b...
Page 30 - No paper; Button input
27 Cause/Error display Cause/Error display 2 2 2 2 2 No paper [Symptom]• When the paper runs out or the paper is not installed, printing becomes impossible and an alarm tone is heard.In this case, all the buttons become invalid. • If this error occurs while more than one copy is being printed or the...
Page 32 - DIP SWITCH FUNCTIONS; DIP SWITCH; FRAME
29 11 DIP SWITCH FUNCTIONS DIP SWITCH Functions 1 IMP(IMPEDANCE)75 Ω / HIGH 2 TRAPON / OFF 3 MEMORYFIELD / FRAME 4 RESERVED 5 RESERVED 6 RESERVED Set to "75 Ω " in normal use. Set to "HIGH" when making branch connection of a monitor orother units to the VIDEO IN connector. Set to ...
Page 33 - STATUS AND MODES
30 12 Set state/Mode LED display Left Contents of right side LED display Power off Stand-by Selection of the memorizedimage for copy printing Setting of the vertical print area γ -curve set mode ( γ -curve) PRINT button function setting Error detect state Power off Underscan / Overscan Printing of t...
Page 34 - Insert the cleaning paper.; USE OF CLEANING PAPER
31 13 When the thermal head is dirty with dust, etc., white spots or stripes may appear on theprint. In this case, clean the thermal head by the following procedure BY USING THESUPPLIED CLEANING PAPER. Roll the cleaning paper and installinto the set. Switch the left side lever to the OPENposition. P...
Page 35 - MAINTENANCE; Turn off the power for maintenance.
32 14 MAINTENANCE Turn off the power for maintenance. Maintenance of Main Unit Wipe off stains of the front panel with a soft cloth.When the panel is heavily stained, wipe of with the clothmoistened with neutral cleanser diluted by water and finish witha dry cloth. Maintenance of Rubber Roller When ...
Page 36 - SPECIFICATIONS; Operating conditions:; SERVICE INFORMATION
33 15 SPECIFICATIONS Type: Video Copy Processor Model: P93W / P93E Power supply andpower consumption: 100-240V AC, 50/60Hz, 1.5 - 0.8A Connection terminal: Video input terminal (BNC contact plug)Video output terminal (BNC contact plug) Resolution: Horizontal 1280 pixels x Vertical 500 lines (Standar...