Invacare CG8000 - Manual

Invacare CG8000

Invacare CG8000 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
Page: / 8

Table of Contents:

  • Page 2 – THERAPY PAD; WARNINGS; FIGURE 1; INSTALLING THE INVACARE 8000 (FIGURES 1 AND 2); FIGURE 2
  • Page 3 – INSTALLING; FIGURE 3; REMOVING; FIGURE 2- INFLATING THE MATTRESS
  • Page 4 – SECURING MATTRESS BASE TO BED FRAME (FIGURE 4); Strap; PLACING PATIENT ON MATTRESS AND ADJUSTING
  • Page 5 – STANDARD DEFLATION; NOTE: Using the “PULL CPR” straps will disconnect
  • Page 6 – in this instruction sheet.; INFLATING THE MATTRESS; in this instruction; CARE AND MAINTENANCE
  • Page 7 – LIMITED WARRANTY
Loading the manual

NOTE: The above model number contains one (1) Alternating Pressure Mattress and one (1)

Power Unit.
NOTE: Check ALL parts for shipping damage. In case of shipping damage, DO NOT use. Contact

Carrier/Dealer for further instruction.

SAFETY SUMMARY

To ensure the safe installation of the Invacare 8000 Alternating Pressure

Mattress and Power Unit, these instructions MUST be followed:

WARNING

DO NOT install or use this equipment without first reading and

understanding this instruction sheet. If you are unable to understand the

Warnings, Cautions or Instructions, contact a healthcare professional,

dealer or technical personnel before attempting to install this equipment -

otherwise, injury or damage may occur.
DO NOT use the mattress near an open flame or while smoking.
Inspect power cord, plug and air hoses of the Invacare 8000 Alternating

Pressure Mattress/Power Unit for breaks, tears, or kinks. If damage is

found, DO NOT use.
DO NOT, under any circumstances, cut or remove the round grounding

prong from any plug used with or for Invacare products. Devices are

equipped with three-prong (grounding) plugs for protection against

possible shock hazards. Where a two-prong wall receptacle is encountered,

it is the personal responsibility and obligation of the customer to contact a

qualified electrician and have the two-prong receptacle replaced with a

properly grounded three-prong wall receptacle in accordance with the

National Electrical Code. If you must use an extension cord, use ONLY a

three-wire extension cord having the same or higher electrical rating as the

device being connected.
DO NOT place hot objects (for example: heating pads, electric blankets,

etc.) on top of the mattress. This may weaken the mattress material.
Use extreme care when handling sharp objects such as needles near the

mattress.
Do NOT inflate mattress of an OCCUPIED bed.
The Alternating Pressure Mattress has a weight limitation of 250 lbs. (114 kg.).
The Invacare 8000 APM MUST be installed on medical bed frames or medical-

grade home care beds with side rails. After placing the patient on the mattress,

verify the side rails provide adequate protection for the patient. The side rails

MUST be in the raised position, whenever a patient is on the bed.

Invacare 8000 Alternating Pressure Mattress

Assembly, Installation and Operating Instructions

Model No. CG8000

1

INVACARE 8000 PRESSURE MATTRESS

Part No. 1078996 Rev. B

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - THERAPY PAD; WARNINGS; FIGURE 1; INSTALLING THE INVACARE 8000 (FIGURES 1 AND 2); FIGURE 2

2 INVACARE 8000 PRESSURE MATTRESS Part No. 1078996 Rev. B INFLATING THE MATTRESS (FIGURE 2) NOTE: Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet. NOTE: Before inflating, ensure that ALL cushion fittings, both air hose quick disconnects and both CPR plugs are secured, otherwise, in...

Page 3 - INSTALLING; FIGURE 3; REMOVING; FIGURE 2- INFLATING THE MATTRESS

3 Part No. 1078996 Rev. B INVACARE 8000 PRESSURE MATTRESS INSTALLING 1. Position therapy pad over mattress.2. Insert one (1) of the ten (10) straps into open latch as shown in FIGURE 3 . 3. Pull strap until therapy pad fits tight around mattress. 4. Push down on latch to secure.5. Repeat STEPS 2 - 4...

Page 4 - SECURING MATTRESS BASE TO BED FRAME (FIGURE 4); Strap; PLACING PATIENT ON MATTRESS AND ADJUSTING

4 INVACARE 8000 PRESSURE MATTRESS Part No. 1078996 Rev. B FIGURE 4 - SECURING MATTRESS BASE TO BED FRAME 1. Wrap one (1) of the ten (10) straps on the mattress around the bed frame. 2. Insert strap into the two (2) double D- rings. 3. Feed strap back through only one (1) of the double D-rings. 4. Pu...