Invacare 7400036 - Manual
Invacare 7400036 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.
Table of Contents:
- Page 2 – WARNING
- Page 3 – FIGURE 4 - CONVERTING FROM PATIENT OPERATED WHEEL LOCKS
- Page 4 – QTY; PARTS LIST
1
Installation, Assembly and Operating Instructions
ATTENDANT OPERATED WHEEL LOCKS
FOR 2000,4000 RECLINER WHEELCHAIRS -
KIT NUMBERS 7400036 AND 7400037
NOTE: Check ALL parts for shipping damage. In case
of shipping damage, DO NOT use. Contact Carrier/
Dealer for further instruction.
SAFETY SUMMARY
The following recommendations are made for safe in-
stallation and adjustment of the Attendant Operated
Wheel Locks:
GENERAL WARNINGS
DO NOT install or use this equipment without
first reading and understanding this instruction
sheet. If you are unable to understand the Warn-
ings, Cautions or Instructions, contact a health-
care professional, dealer or technical person-
nel before attempting to install this equipment
- otherwise, injury or damage may occur.
DO NOT attempt to stop a moving wheelchair
with the wheel locks. WHEEL LOCKS ARE NOT
BRAKES.
INSTALLATION WARNING
After making adjustments and BEFORE use,
make sure all attaching hardware is securely
tightened.
REPLACING THE ATTENDANT
OPERATED WHEEL LOCKS
(FIGURE 1)
NOTE: Refer to INSTALLATION WARNING in the
SAFETY SUMMARY in this instruction sheet.
1. Remove the mounting screw and locknut that se-
cures the locking rod to the back tube mounting
bracket.
2. Slide anti-rattle up and off of locking rod.
3. Perform one (1) of the following:
a. If converting from Patient Operated Wheel
Locks, proceed to STEP 6.
b. If replacing current Attendant Operated Wheel
Locks, continue with STEP 4.
4. Loosen the locknut that secures the attendant op-
erated wheel lock to the recliner tube.
5. Slide the existing attendant operated wheel lock
up and off of the locking rod.
6. Slide the NEW attendant operated wheel lock
down locking rod and onto the recliner tube.
7. Reposition the anti-rattle on top of recliner tube.
8. Resecure the locking rod to the back tube mount-
ing bracket with the existing mounting screw and
locknut. Securely tighten.
9. Adjust the Attendant Operated Wheel Lock. Re-
fer to ADJUSTING THE ATTENDANT OPER-
ATED WHEEL LOCKS in this instruction sheet.
10. Repeat STEPS 1-9 for opposite wheel lock, if nec-
essary.
FIGURE 1 -REPLACING THE ATTENDANT
OPERATED WHEEL LOCKS
Locknut
(STEPS 1,8)
Locking Rod
(STEPS 1,2,5,6,8)
BackTube
Mounting
Bracket
(STEPS 1,8)
Mounting
Screw
(STEPS 1,8)
Anti-Rattle
(STEPS 1,7)
Recliner Tube
(STEPS 4,6,7)
Locknut
(STEP 4)
Attendant Operated
Wheel Lock
(STEPS 4-6)
Recliner Tube
(STEPS 4,6,7)
Locking Rod
(STEPS 1,2,
5,6,8)
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
2 Wheel Lock (STEP 1) Wheel Lock Shoe (STEP 2) 5/32 to 5/16- inches (STEP 2) Locknut (STEP 1) Rear Wheel Recliner Tube (STEP 1) FIGURE 2 - ADJUSTING THE ATTENDANT OPERATED WHEEL LOCKS Tire Section Tire Contact Area DETAIL "A" - BACK VIEW OF TIRE Wheel Lock Shoe ADJUSTING THE ATTENDANTOPERATE...
3 CONVERTING FROM PATIENTOPERATED WHEEL LOCKS(FIGURE 4) NOTE: Refer to INSTALLATION WARNING in theSAFETY SUMMARY in this instruction sheet. 1. Remove the existing axle mounting screw and locknut that secures the rear portion of the patientoperated wheel lock assembly, pivot arm spacerand rear wheel ...
INVACARE CORPORATION l ONE INVACARE WAY l P.O. Box 4028 l Elyria, Ohio 44036-2125 l Phone 1-(800)-333-6900 Form No. 98-101 Part No. 1076130 Rev B (2) - 6/99 Printed in U.S.A. QTY ITEM DESCRIPTION 7400036 7400037 1 Attendant Operated Wheel Lock 1 1 2 3/4-inch Spacer 1 1 3 4-3/4-inch Axle Mounting Scr...