Hanna Instruments HI 2300 - Manual
Hanna Instruments HI 2300 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.
Table of Contents:
- Page 2 – is guaranteed for two years against defects in workmanship and
- Page 3 – P R E L I M I N A R Y E X A M I N A T I O N; Conductivity / TDS probe; G E N E R A L D E S C R I P T I O N; Hanna Instruments reserves the right to modify the design,
- Page 4 – F U N C T I O N A L D E S C R I P T I O N; ) key, to manually decrease temperature value or other parameters.; Front Panel; Rear Panel
- Page 5 – S P E C I F I C A T I O N S
- Page 7 – O P E R A T I O N A L G U I D E; C O N N E C T I O N S
- Page 8 – A U T O R A N G I N G; T E M P E R A T U R E C O M P E N S A T I O N; ARROW
- Page 9 – E C / T D S C A L I B R A T I O N
- Page 10 – N a C l C A L I B R A T I O N
- Page 13 – Co
- Page 14 – S E T U P
- Page 15 – G O O D L A B O R A T O R Y P R A C T I C E
- Page 17 – D A T A T R A N S F E R T O P C
- Page 18 – P R O B E M A I N T E N A N C E; CONDUCTIVITY BUFFER SOLUTIONS; A C C E S S O R I E S
- Page 20 – SALES AND TECHNICAL SERVICE CONTACTS
1
Instruction Manual
HI 2300
Auto-ranging
Microprocessor
EC/TDS/NaCl/ºC
Bench Meter
C
HI2300
Microprocessor Conductivity Meter
ALT Functions
Functions
TC
ATC
ALT
LOCK
SETUP
GLP
RANGE
CFM
CAL
CALT
Display TC Factor
TC
View Calibration
Data
GLP
SETUP
Enter Setup Mode
Start T
emperature
Calibration
CAL
T
Temperature
Compensation
ATC
Confirm V
alue
CFM
Change V
alue
/
Lock Current Range
LOCK
Start Calibration
CAL
Select Measurement
Range
RANGE
w w w . h a n n a i n s t . c o m
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
2 T A B L E O F C O N T E N T S W A R R A N T Y HI 2300 is guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according toinstructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or re...
3 P R E L I M I N A R Y E X A M I N A T I O N Remove the instrument from the packing material and examine it carefully tomake sure that no damage has occurred during shipping. If there is anydamage, notify your Dealer or the nearest Hanna Customer Service Center.Each instrument is supplied with: • H...
4 F U N C T I O N A L D E S C R I P T I O N 1) Liquid Crystal Display (LCD).2) CAL key, to enter/exit calibration mode. CALT key (alternate function), to enter temperature calibration mode. 3) CFM key, to confirm different values. 4) key, to manually decrease temperature value or other parameters. S...