Fluke 3458/HFL - Manual

Fluke 3458/HFL

Fluke 3458/HFL – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12
13 Page 13
Page: / 13

Table of Contents:

  • Page 2 – D C V o lta g e; R a n g e
  • Page 3 – DC Current; Range
  • Page 4 – Resistance
  • Page 5 – AC Current
  • Page 8 – this compares with 10u
  • Page 9 – of last ACAL
  • Page 12 – 508A Options and Accessories:
Loading the manual



Multimeter Selection Guide
Fluke 8508A & Agilent 3458/HFL

TECHNICAL INFORMATION



When comparing two products features and specification using sales information
it is often difficult to determine which product will meet the needs of your specific
application. This is mostly because each supplier will define specifications using
alternate methods. This Selection Guide will help you compare the specifications
and features of the Fluke 8508A Reference multimeter with the nearest
equivalent digital multimeter the Agilent 3458/HFL. Both the 8508A and the
special 3458/HFL multimeters are available through the Fluke sales channel.

The following pages summarize & segment the features of each meter into the
following three categories.

Accuracy

Functionality & Versatility

Ease of Use


The 'Comments' column found in the following table provides a brief explanation
against each feature and its relevance to the intended metrology applications.

Further reference information:


1) Agilent 3458A/HFL Technical Data - 1275558 D-ENG-N

2) Fluke 8508A Extended Specifications - 1887212 D-ENG-N rev A

3) 1281Operators Handbook – 850090 issue 6



"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - D C V o lta g e; R a n g e

D C V o lta g e ± (p p m o f s e ttin g + p p m o f ra n g e ), re la tiv e to c a lib ra tio n s ta n d a rd s u s e d . C a lib ra tio n In te rva l 3 6 5 d a y, ± 1 °C o f la s t T C A L R a n g e 1 2 8 1 H P 3 4 5 8 /H F L 8 5 0 8 A 1 2 0 .0 0 0 0 0 m V - 4 + 3 - 1 9 9 .9 9 9 9 9 9 m V 6 + 0 .5 ...

Page 3 - DC Current; Range

DC Current ±(ppm of setting + ppm of range), relative to calibration standards used. Calibration Interval 365 day, ±1°C of last TCAL Range 1281 HP3458/HFL 8508A 120.000nA - 30 + 400 - 1.200000µA - 20 + 40 - 12.000000µA - 20 + 10 - 120.00000µA - 20 + 8 - 199.99999µA 25 + 2 - 6.5 + 2.0 1.2000000mA - 2...

Page 4 - Resistance

Resistance ±(ppm of setting + ppm of range), relative to calibration standards used. Calibration Interval 365 day, ±1°C of last TCAL Range 1281 HP3458/HFL 8508A 1.99999999 Ohm - - 10 + 2.0 12.00000 Ohm - 10 + 5 - 19.9999999 Ohm 12 + 1.0 - 7.0 + 0.7 120.00000 Ohm - 8 + 5 - 199.999999 Ohm 8 + 0.3 7.0 ...