Extech PH210 - Manual
Extech PH210 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.
Table of Contents:
- Page 2 – Introduction; Features; Meter
- Page 3 – Calibration
- Page 8 – Copyright
User
Guide
Model
PH210
pH
/
ORP
/
Temperature
Meter
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
PH210-en-GB_v1.3 11/14 2 12 3 45 67 8 9 10 Introduction Thank you for selecting the Extech PH210 pH/ORP/Temperature meter. This microprocessor ‐ based device with tactile push ‐ buttons is battery operated and ideal for field use. The PH210 simulta...
PH210-en-GB_v1.3 11/14 3 Getting Started 1. Calibrate the instrument often, as described in a later section. Note that only a pH calibration is required; mV (ORP) is calibrated as a by ‐ product of the pH calibration. 2. Set the unit ...
PH210-en-GB_v1.3 11/14 8 Troubleshooting an Electrode Symptom Cause Recommendation Long response time or reading drift Clogged Junction Soak in 4.07 M Potassium Chloride (KCl) @ 60 o C for 30 minutes. Strong Alkaline measurement Soak in 0...