Model 499A CL-01
Free Chlorine Sensor
Instruction Sheet
PN 51A-499ACL01/rev.K
September 2010
SPECIFICATIONS - SENSOR
Pressure:
0 to 65 psig (0 to 549 kPa abs)
Temperature:
32 to 122°F (0 to 50°C)
Process Connection:
1 inch MNPT
Wetted Parts:
Noryl®
1
, Viton®
2
, platinum, polyethersulfone,
polyester, wood, silicone
Type
PN
Wetted Materials
Process Connection
Max.Temperature
Max. Pressure
1-1/2 in. tee
23567-00
CPVC, Buna N
1-1/2 in. socket
122°F (50°C)
65 psig (549 kPa abs)
915240-03
3/4 in. NFPT
2 in. tee
915240-04
PVC, Buna N
1 in. NFPT
120°F (49°C)
60 psig (515 kPa abs)
915240-05
1-1/2 in. NFPT
Low Flow*
24091-00
Polycarbonate/
Compression fitting for
158°F (70°C)
90 psig (722 kPa abs)
polyester, 316SS,
1/4 in. OD tubing
silicone
24091-01 (with
Polycarbonate/
Compression fitting for
158°F (70°C)
90 psig (722 kPa abs)
bubble sweep- polyester, 316SS,
1/4 in. OD tubing
ing nozzle)
silicone
1 Noryl is a registered trademark of General Electric.
2 Viton is a registered trademark of DuPont Performance Elastomers
SPECIFICATIONS - FLOW CELLS
For additional information, please visit our website
at
www.emersonprocess.com/raihome/liquid/
.
CAUTION
Do not exceed pressure and temperature specifications.
Pressure: 65 psig max (549 kPa abs max)
Temperature: 32 to 122°F (0 to 50°C)
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
The wetted sensor materials may not be compatible
with process com
position and operating conditions.
Application compat ibility is entirely the responsibility of
the user.
* Pressure and temperature specifications for the low flow cell exceed the temperature and pressure specifications for the sensor.
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
2 MODEL 499A CL-01 INSTALLATION FIGURE 4. Low Flow Cell (PN 24091-00 or 24091-01) INSTALLATION Install the sensor in a flowing sample. Keep the sample flowas constant as possible at a value within the limits below. Flow through 1 to 5 gpm (3.8 to 19 L/min) Open channel 1 ft/sec (0.3 m/sec) Low flow ...
MODEL 499A CL-01 WIRING WIRING 3 FIGURE 5. Wiring Sensor with Standard Cable toModel 1056 and Model 56 Analyzer FIGURE 6. Wiring Sensor with Optimum EMI/RFI orVariopol Cable to Model 1056 and Model 56 Analyzer FIGURE 7. Wiring Sensor with Standard Cable toModel 54eA Analyzer FIGURE 8. Wiring Sensor ...
FIGURE 13. Wiring Sensor with Standard Cable toModel 1066 Transmitter. FIGURE 14. Wiring Sensor with Optimum EMI/RFI orVariopol Cable to Model 1066 Transmitter. CALIBRATION Zero point: Even in the absence of free chlorine, the 499ACL-01 sensor generates a small signal called thezero current. Failing...