Page 2 - ESSENTIAL INSTRUCTIONS; READ THIS PAGE BEFORE PROCEEDING!; WARNING; About This Document; Date
ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its products tomeet many national and international standards. Because these instru-ments are sophisticated technical products, you must properly install, use,and maintain them to ensure th...
Page 4 - MODEL 54eA ANALYZER/CONTROLLER; TABLE OF CONTENTS
MODEL 54eA TABLE OF CONTENTS MODEL 54eA ANALYZER/CONTROLLER TABLE OF CONTENTS Section Title Page 1.0 SPECIFICATIONS ................................................................................................... 1 1.1 Features and Applications .......................................................
Page 5 - Section
9.0 CALIBRATION - FREE CHLORINE (498 ACL-01) ................................................ 60 9.1 Introduction .............................................................................................................. 60 9.2 Zeroing the Sensor ....................................................
Page 6 - LIST OF FIGURES
iii MODEL 54eA TABLE OF CONTENTS LIST OF FIGURES Section Title Page 2-1 Wall Mounting .......................................................................................................... 5 2-2 Pipe Mounting...........................................................................................
Page 7 - SPECIFICATIONS; FEATURES AND APPLICATIONS
1 MODEL 54eA SECTION 1.0 SPECIFICATIONS SECTION 1.0 SPECIFICATIONS 1.1 FEATURES AND APPLICATIONS The Model 54eA Analyzer/Controller with the appro-priate sensor monitors and controls dissolved oxy-gen (ppm and ppb level), free chlorine, totalchlorine, and ozone in a variety of process liquids.The an...
Page 10 - ORDERING INFORMATION
MODEL 54eA SECTION 1.0 SPECIFICATIONS 4 1.3 ORDERING INFORMATION The Model 54eA Microprocessor Analyzer measures dissolved oxygen, free chlorine, total chlorine, ozone, and pH. pH is available for free chlorine only. The analyzer has an on-board pressure sensor for automatic aircalibration of oxygen...
Page 11 - INSTALLATION; UNPACKING AND INSPECTION
5 MODEL 54eA SECTION 2.0 INSTALLATION SECTION 2.0 INSTALLATION 2.1 UNPACKING AND INSPECTION Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. Ifthere is no apparent damage, unpack the container. Be sure all items shown on the packing li...
Page 14 - WIRING; NOTE; Be sure to turn off power to the analyzer before wiring the sensor.
8 MODEL 54eA SECTION 3.0 WIRING SECTION 3.0 WIRING 3.1 GENERAL WARNING Electrical installation must conform to the National Electrical Code, all state and local codes, and all plantcodes and standards for electrical equipment. Electrical installation and wiring must be done by qualifiedpersonnel. Th...
Page 22 - DISPLAY AND OPERATION; GENERAL DESCRIPTION; ppm
16 MODEL 54eA SECTION 4.0 DISPLAY AND OPERATION SECTION 4.0 DISPLAY AND OPERATION 4.1 GENERAL DESCRIPTION The 54eA analyzer/controller is a normally a singleinput, dual output instrument. It does, however, accepta second input for pH when the main measurement isfree chlorine. Figure 4-1 shows how th...
Page 23 - SOFTWARE CONFIGURATION
17 MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION SECTION 5.0 SOFTWARE CONFIGURATION The instrument is configured at the factory to measure oxygen. Figure 5-1 is an outline of the menu structure. Table 5-1 lists the default settings and the range of choices available for each setting. To reduce the c...
Page 30 - CHANGING ALARM SETPOINTS; Alarm setpoints
24 5.1 CHANGING ALARM SETPOINTS 2. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 3. Press Enter (F4). 4. Move the cursor to the desired alarm and press Enter (F4). 5. The screen appearing at this point depends on how the alarm was configured. 6. If ...
Page 31 - RANGING THE OUTPUTS; Output 1 setpoints
MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.2 RANGING THE OUTPUTS 2. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 3. Move the cursor to "Output setpoints" and press Enter (F4). 4. Move the cursor to the desired output and press Enter (F...
Page 32 - Setpoint
26 MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.3 CHANGING OUTPUT SETPOINTS (PID ONLY) 2. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 3. Move the cursor to "Output setpoints" and press Enter (F4). 4. Move the cursor to the desired outpu...
Page 33 - TESTING OUTPUTS AND ALARMS; Output setpoints; S i m u l a t i n g o u t p u t 1; Test alarm 1; S i m u l a t i n g a l a r m 1
MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.4 TESTING OUTPUTS AND ALARMS 2. Press any key to enter the main menu. Move the cursor to "Program"and press Enter (F4). 3. Move the cursor to "Simulated tests" and press Enter (F4). 4. Move the cursor to the desired output or alarm. Bot...
Page 34 - Configure; CHOOSING DISPLAY OPTIONS; Display
28 Output setpointsSimulated tests Configure Exit Enter MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.5 CHOOSING DISPLAY OPTIONS 3. Press any key to enter the main menu. Move the cursor to "Program"and press Enter (F4). 4. Move the cursor to "Configure" and press Enter (F4). 5. Wit...
Page 36 - CHANGING OUTPUT PARAMETERS; Output 1 control
30 MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.6 CHANGING OUTPUT PARAMETERS 3. Press any key to enter the main menu. Move the cursor to "Program"and press Enter (F4). 4. Move the cursor to "Configure" and press Enter (F4). 5. Move the cursor to "Outputs" and press Enter (...
Page 37 - Ctrl mode
MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.6 CHANGING OUTPUT PARAMETERS (continued) e. The display returns to the “Output: Process” screen. Press Exit(F1). The display returns to the “Output Measurement” screen.Move the cursor to "Control mode" and press Enter (F4). f. Press Edit (F4). ...
Page 38 - Hold feature setup
32 MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION Proportional: Proportional is the same as proportional band and is the range over which control is being used. It is the opposite ofprocess gain. Smaller values provide tighter control. Integral: Integral is the number of seconds over which devia- tio...
Page 39 - CHANGING ALARM PARAMETERS; Alarm 1 control
MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.7 CHANGING ALARM PARAMETERS 3. Press any key to enter the main menu. Move the cursor to "Program"and press Enter (F4). 4. Move the cursor to "Configure" and press Enter (F4). 5. Move the cursor to "Alarms" and press Enter (F4). ...
Page 40 - Activate
34 MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.7 CHANGING ALARM PARAMETERS (continued) c. To change the activation method, press Edit (F4). Use the é key to scroll through the choices: "Process", "Temperature", and “pH”(if pH was enabled). "Process" means the measurement ...
Page 42 - Feed limit; F e e d l i m i t a l a r m 1
36 Feed limit : Disable Timeout: 3600 sec Exit Edit 1.000 ppm 26.2°C 12.0mA F e e d l i m i t a l a r m 1 MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.7 CHANGING ALARM PARAMETERS (continued) Example: The setpoint is 3.00 ppm. The URV is +5.00 ppm and the LRV is 0.00 ppm. The time period is 30 sec...
Page 43 - Timer; Interval timer
MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.7 CHANGING ALARM PARAMETERS (continued) 11. Interval timer setup: Alarm 1, 2, or 3 can be used as an interval timer. The selected relay willopen and close at time intervals programmed by the user. The intervaltimer is useful for automatic cleaning of s...
Page 44 - CONFIGURING THE pH MEASUREMENT; pH sensor
38 MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.8 CONFIGURING THE pH MEASUREMENT 1. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 2. Move the cursor to "Configure" and press Enter (F4). 3. Move the cursor to “pH” and press Enter (F4). 4. ...
Page 45 - Diagnostics
MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.8 CONFIGURING THE pH MEASUREMENT (continued) key to change "Off" to "On". Press Save (F4). NOTE Choosing "On" means the controller will display pH diagnos-tic warning messages and fault alarms. Choosing "Off"means the me...
Page 46 - Temp coeff; Temperature coeff
40 MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.8 CONFIGURING THE pH MEASUREMENT (continued) Temp coeff : 0.000 Operate iso: 7.00 pHSensor iso: 7.00 pH Exit Enter Stabilize pH and Stabilize time: For the controller to accept cal- ibration data, the pH must remain within a specified range for aspe...
Page 47 - TEMPERATURE COMPENSATION AND TEMPERATURE UNITS; pH
Temp comp : Auto Temp units: °C Exit Edit Temp comp: ManualTemp units: °C Temperature : 25.0°C Exit Edit MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.9 TEMPERATURE COMPENSATION AND TEMPERATURE UNITS 1. Refer to Section 6.1 for a discussion of the ways in which temperature affects amperometric and...
Page 48 - Noise rejection; Temperature
42 Noise rejection : 60 Hz Exit Edit MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.10 NOISE REDUCTION 1. For maximum noise reduction the frequency of the ac power must be entered into the analyzer. Temperature Noise rejection Main sensor cal Exit Enter 2. Press any key to enter the main menu. Move...
Page 49 - MAIN SENSOR CALIBRATION PARAMETERS
Stabilize : 0.050 ppm Stabilize time: 10 sec Exit Edit Stabilize: 0.050 ppmStabilize time: 10 sec Dual range cal : Disable Exit Edit Stabilize: 0.050 ppmStabilize time: 10 sec Salinity : 0.0 o/oo Exit Edit MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.11 MAIN SENSOR CALIBRATION PARAMETERS 1. Main ...
Page 50 - Bar meas; BAROMETRIC PRESSURE
44 Bar meas : Auto Bar units: mm Hg Exit Edit MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.12 BAROMETRIC PRESSURE 1. The barometric pressure menu header appears only if an oxygen sensor (ppm or ppb level) is being used. Barometric pressure is used during air calibration. Noise rejection Barometri...
Page 51 - Lock all
MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.13 SECURITY 1. The controller can be programmed to require a password for access to menus. There are three levels: Level 1: A level 1 user can 1. Zero and calibrate the main (amperometric) sensor 2. Calibrate the barometric pressure sensor 3. Calibrate...
Page 52 - CONTROLLER MODE PRIORITY; TABLE 5-2. Controller Mode Priority Chart
46 MODEL 54eA SECTION 5.0 SOFTWARE CONFIGURATION 5.14 CONTROLLER MODE PRIORITY The Model 54eA controller can function in differentmodes depending on both how it is configured, whatprocess conditions exist, and actions an operator mayhave made. To reconcile these possible modes, there isa set priorit...
Page 53 - CALIBRATION - TEMPERATURE
47 MODEL 54eA SECTION 6.0 CALIBRATION - TEMPERATURE SECTION 6.0 CALIBRATION - TEMPERATURE 6.1 INTRODUCTION All four amperometric sensors (oxygen, ozone, free chlorine, total chlorine, and monochloramine) are membrane-covered sensors. As the sensor operates, the analyte (the substance to be determine...
Page 54 - TEMPERATURE CALIBRATION; Adjust temp
48 6.2 TEMPERATURE CALIBRATION MODEL 54eA SECTION 6.0 CALIBRATION - TEMPERATURE 2. From the main display, press any key. With the cursor on “Calibrate,”press Enter (F4). NOTE If Hold was enabled in Section 5.6, the hold screen will appear.To activate hold, refer to Section 5.6, step 11. 3. Press the...
Page 55 - CALIBRATION - DISSOLVED OXYGEN
MODEL 54eA SECTION 7.0 CALIBRATION - DISSOLVED OXYGEN SECTION 7.0 CALIBRATION - DISSOLVED OXYGEN 7.1 INTRODUCTION As Figure 7-1 shows, oxygen sensors generate a current directly proportional to the concentration of dissolvedoxygen in the sample. Calibrating the sensor requires exposing it to a solut...
Page 56 - ZEROING THE SENSOR; Calibrate main sensor
50 1. Place the sensor in a fresh solution of 5% sodium sulfite (Na 2 SO 3 ) in water. Be sure air bubbles are not trapped against the membrane. The current will drop rapidly at first and then gradually reach a stable zerovalue. To monitor the sensor current, go to the main display. Press any key to...
Page 57 - CALIBRATING THE SENSOR IN AIR
1. Remove the sensor from the process liquid. Use a soft tissue and a stream of water from a wash bottle to clean the membrane. Blot dry. The membrane must be dry during air calibration. 2. Pour some water in a beaker and suspend the sensor with the membrane about 0.5 inch (1 cm) above the water sur...
Page 59 - CALIBRATING THE SENSOR AGAINST A STANDARD INSTRUMENT; Calibrate
The analyzer and sensor can be calibrated against a standard instrument. For oxygen sensors installed in aera-tion basins in waste treatment plants, calibration against a second instrument is often preferred. For an accuratecalibration be sure that… 1. The standard instrument has been zeroed and cal...
Page 60 - CALIBRATING BAROMETRIC PRESSURE; mm Hg; Standardize
54 1. If the barometric pressure measured by the controller does not agree with the local barometric pressure, cal- ibrate the pressure sensor. A pressure error of 3 mm Hg introduces an error of about 0.5% in the final meas-urement. When calibrating the pressure reading, be sure to use the actual ba...
Page 62 - ZEROING THE SENSOR
56 1. Place the sensor in the zero standard (see Section 8.1). Be sure no air bubbles are trapped against the mem- brane. The sensor current will drop rapidly at first and then gradually reach a stable zero value. To monitor thesensor current, go to the main display. Press any key to obtain the main...
Page 63 - FULL SCALE CALIBRATION
1. Place the sensor in the process liquid. If automatic pH correction is being used, calibrate the pH sensor (see Section 11.0) and place it in the process liquid. If manual pH correction is being used, measure the pH of the process liquid and enterthe value. See Section 5.8. Adjust the sample flow ...
Page 64 - DUAL SLOPE CALIBRATION; Adjust temperature; Calibrate low point
58 MODEL 54eA SECTION 8.0 CALIBRATION - FREE CHLORINE (499ACL-01) Figure 8-2 shows the principle of dual slope cali-bration. Between zero and concentration C1, thesensor response is linear. When the concentrationof chlorine becomes greater than C1, the responseis non-linear. In spite of the non-line...
Page 65 - Low point
MODEL 54eA SECTION 8.0 CALIBRATION - FREE CHLORINE (499 ACL-01+3.) 7. Once the reading is stable, the screen at left appears. Sample the process liquid. Make a note of the reading before takingthe sample. Immediately determine free chlorine. Note the controllerreading again. If the present reading (...
Page 67 - ZEROING THE SENSOR
MODEL 54eA SECTION 9.0 CALIBRATION - FREE CHLORINE (498CL-01) 61 1. BEFORE ZEROING THE SENSOR, PLACE IT IN THE FLOW CELL AND ALLOW THE SENSOR TO OPER-ATE IN A FLOWING CHLORINATED SAMPLE FOR AT LEAST TWO HOURS. 2. After two hours, remove the sensor from the flow cell and place it in the zero standard...
Page 68 - FULL SCALE CALIBRATION
MODEL 54eA SECTION 9.0 CALIBRATION - FREE CHLORINE (498CL-01) 1. Place the sensor in the process liquid. Be sure the controller has been programmed for manual pH correction and the pH is 7.00. Adjust the sample flow until it is in the range recommended for the sensor. Refer to the sensor instruction...
Page 69 - DUAL SLOPE CALIBRATION
MODEL 54eA SECTION 9.0 CALIBRATION - FREE CHLORINE (498CL-01) 63 Figure 9-2 shows the principle of dual slope cali-bration. Between zero and concentration C1, thesensor response is linear. When the concentrationof chlorine becomes greater than C1, the responseis non-linear. In spite of the non-linea...
Page 71 - CALIBRATION - TOTAL CHLORINE; See TCL manual for calibration procedure.
MODEL 54eA SECTION 10.0 CALIBRATION - TOTAL CHLORINE SECTION 10.0 CALIBRATION - TOTAL CHLORINE See TCL manual for calibration procedure. 65
Page 72 - This page intentionally left blank.
Page 73 - CALIBRATION - MONOCHLORAMINE
MODEL 54eA SECTION 11.0 CALIBRATION - MONOCHLORAMINE SECTION 11.0 CALIBRATION - MONOCHLORAMINE 67 11.1 INTRODUCTION As Figure 11-1 shows, a monochloramine sensor generates a current directly proportional to the concentration ofmonochloramine in the sample. Calibrating the sensor requires exposing it...
Page 74 - ZEROING THE SENSOR
68 MODEL 54eA SECTION 11.0 CALIBRATION - MONOCHLORAMINE 1. Place the sensor in the zero standard. Be sure no air bubbles are trapped against the membrane. The sensor current will drop rapidly at first and then gradually reach a stable zero value. To monitor the sensor current, goto the main display....
Page 75 - FULL SCALE CALIBRATION
MODEL 54eA SECTION 11.0 CALIBRATION - MONOCHLORAMINE 1. Place the sensor in the process liquid. Adjust the sample flow until it is within the range recommended for the sensor. Refer to the sensor instruction sheet. 2. Adjust the monochloramine concentration until it is near the upper end of the cont...
Page 77 - ZEROING THE SENSOR
1. Place the sensor in the zero standard. Be sure no air bubbles are trapped against the membrane. The sensor current will drop rapidly at first and then gradually reach a stable zero value. To monitor the sensor current, goto the main display. Press any key to obtain the main menu. Press the ê key ...
Page 78 - FULL SCALE CALIBRATION
72 1. Place the sensor in the process liquid. Adjust the sample flow until it is within the range recommended for the sensor. Refer to the sensor instruction sheet. 2. Adjust the ozone concentration until it is near the upper end of the control range. Wait until the controller read-ing is stable bef...
Page 80 - Autocal buffer 1
74 4.02 pH Buf1done: 4.00 @25°C Abort Cont 4.02 pH Autocal buffer 1 Abort Cont Adjust temperature Calibrate pH Output trim Exit Enter Buffer calibration Standardize pHpH slope Exit Enter 1. Be sure the pH feature has been enabled. See Section 5.8. 2. Obtain two buffer solutions. Ideally the buffer p...
Page 81 - Autocal buffer 2
MODEL 54eA SECTION 13.0 CALIBRATION - pH 9. Remove the sensor from buffer 1, rinse it with water, and place it inbuffer 2. Swirl the sensor. Press Cont (F3). "Wait" flashes until thereading is stable. 10. The screen at left appears once the reading is stable. Use the é or ê key to change the...
Page 82 - Calibrate point 1
76 1. Be sure the pH feature has been enabled. See Section 5.8. 2. The controller comes from the factory set for automatic pH calibration. To do a manual calibration, the factory default setting must be changed. Refer to Section 5.8. 3. Obtain two buffer solutions. Ideally the buffer pH values shoul...
Page 83 - Calibrate point 2
MODEL 54eA SECTION 13.0 CALIBRATION - pH 10. The screen at left appears if the pH reading in step 9 was changed. Press Pt2 (F3). Go to step 11. 11. Rinse the sensor and thermometer with water and place them in buffer 2. Be sure the glass bulb and junction are completely sub-merged. Swirl the sensor....
Page 84 - Standardize pH
78 1. The pH measured by the controller can be changed to match the reading from a second or referee instrument. The process of making the two reading agree is called standardization, or one-point calibration. 2. During standardization, the difference between the two pH values is converted to the eq...
Page 85 - pH SLOPE ADJUSTMENT
1. If the electrode slope is known from other measurements, it can be entered directly into the controller. The slope must be entered as the slope at 25°C. To calculate the slope at 25°C from the slope at temperature t°C,use the equation: slope at 25°C = (slope at t°C) Changing the slope overrides t...
Page 86 - CALIBRATION - CURRENT OUTPUTS; TRIMMING THE OUTPUTS; Output trim; Trim output 1; Cal point
80 MODEL 54eA SECTION 14.0 CALIBRATION - CURRENT OUTPUTS SECTION 14.0 CALIBRATION - CURRENT OUTPUTS 14.1 INTRODUCTION Although the controller outputs are calibrated at the factory, they can be trimmed in the field to match the readingfrom a standard current meter. Both the low output (0 or 4 mA) and...
Page 87 - PID AND TPC CONTROL
81 MODEL 54eA SECTION 15.0 PID AND TPC CONTROL SECTION 15.0 PID AND TPC CONTROL 15.1 PID CONTROL (CODE -20) PID Control The Model 54eA current outputs can be programmedfor PID control. PID control is used with a controldevice that is capable of varying its output from 0 to100 percent in response to ...
Page 91 - TIME PROPORTIONAL CONTROL
MODEL 54eA SECTION 15.0 CALIBRATION - CONTROL 15.2 TIME PROPORTIONAL CONTROL (TPC) MODE (Code -20) In the TPC mode, you must establish the followingparameters which will determine how the Model54eA controller responds to your system (seeSection 5.7): • Setpoint • Time period • URV point (or 100% on)...
Page 92 - TROUBLESHOOTING; TROUBLESHOOTING WHEN A FAULT MESSAGE IS SHOWING
86 MODEL 54eA SECTION 16.0 TROUBLESHOOTING SECTION 16.0 TROUBLESHOOTING 16.1 OVERVIEW The 54eA controller continuously monitors itself and the sensor for faults. When the controller detects a fault in theamperometric or pH sensor or in the instrument itself it displays a fault message . If alarm 4 w...
Page 94 - TROUBLESHOOTING WHEN NO FAULT MESSAGE IS SHOWING - OXYGEN
88 MODEL 54eA SECTION 16.0 TROUBLESHOOTING 16.2.8 Sense line open Most Rosemount Analytical sensors use a Pt100 or a Pt1000 in a three-wire configuration (see Figure 16-4). Thein and return leads connect the RTD to the measuring circuit in the analyzer. A third wire, called the sense line, isconnect...
Page 100 - TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING —
MODEL 54eA SECTION 16.0 TROUBLESHOOTING 16.7 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — MONOCHLORAMINE 16.7.1 Zero current is too high A. Is the sensor properly wired to the analyzer? See Section 3.3. B. Is the zero solution monochloramine-free? Take a sample of the solution and test it for ...
Page 103 - TROUBLESHOOTING WHEN NO FAULT MESSAGE IS SHOWING - OZONE
MODEL 54eA SECTION 16.0 TROUBLESHOOTING 16.8 TROUBLESHOOTING WHEN NO FAULT MESSAGE IS SHOWING - OZONE 16.8.1 Zero current is too high A. Is the sensor properly wired to the controller. See Section 3.3. B. Is the zero solution ozone free? Test the zero solution for ozone level. The concentration shou...
Page 105 - TROUBLESHOOTING WHEN NO FAULT MESSAGE IS SHOWING - pH
MODEL 54eA SECTION 16.0 TROUBLESHOOTING 16.9 TROUBLESHOOTING WHEN NO FAULT MESSAGE IS SHOWING - pH 16.9.1 Warning or error message during two-point calibration. Once the two-point (manual or automatic) calibration is complete, the controller automatically calculates the sen-sor slope (at 25°C). If t...
Page 108 - TROUBLESHOOTING NOT RELATED TO MEASUREMENT PROBLEMS
102 MODEL 54eA SECTION 16.0 TROUBLESHOOTING 16.10 TROUBLESHOOTING NOT RELATED TO MEASUREMENT PROBLEMS Problem Action Display segments missing Replace display board Alarm relays are chattering 1. Check alarm setpoints.2. Increase hysteresis time delay settings (see Section 5.7) Incorrect current outp...
Page 109 - SIMULATING INPUTS - OTHER AMPEROMETRIC MEASUREMENTS
MODEL 54eA SECTION 16.0 TROUBLESHOOTING 16.12 SIMULATING INPUTS - OTHER AMPEROMETRIC MEASUREMENTS To check the performance of the controller, use a decade box and a battery to simulate the current from the sen-sor. The battery, which opposes the polarizing voltage, is necessary to ensure that the se...
Page 111 - SIMULATING TEMPERATURE
MODEL 54eA SECTION 16.0 TROUBLESHOOTING 16.14 SIMULATING TEMPERATURE 16.14.1 General. The 54eA controller accepts either a Pt100 RTD (for pH,499ADO, 499ATrDO, 499ACL-01, 499ACL-02, 499ACL-03,and 499AOZ sensors) or a 22k NTC thermistor (for Hx438and Gx448 DO sensors and most steam-sterilizable DOsens...
Page 112 - MEASURING REFERENCE VOLTAGE
106 MODEL 54eA SECTION 16.0 TROUBLESHOOTING 16.15 MEASURING REFERENCE VOLTAGE Some processes contain substances that poison or shiftthe potential of the reference electrode. Sulfide is a goodexample. Prolonged exposure to sulfide converts the ref-erence electrode from a silver/silver chloride electr...
Page 113 - MAINTENANCE
MODEL 54eA SECTION 17.0 MAINTENANCE SECTION 17.0 MAINTENANCE REPLACEMENT PARTS PART NUMBER DESCRIPTION 23540-05 Enclosure, Front with Keyboard 23848-00 Power Supply Circuit Board Shield 23849-00 Half Shield, Power Supply 23969-02 PCB, CPU and power supply, calibrated, 115/230 Vac 23969-06 PCB, CPU a...
Page 114 - RETURN OF MATERIAL
MODEL 54eA SECTION 18.0 RETURN OF MATERIAL SECTION 18.0 RETURN OF MATERIAL 18.1 GENERAL. To expedite the repair and return of instruments, proper communication between the customer and the factory isimportant. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authoriz...