Model 499A OZ
Ozone Sensor
Instruction Sheet
PN 51A-499AOZ/rev.H
September 2010
SPECIFICATIONS - SENSOR
Pressure:
0 to 65 psig (101 to 549 kPa abs)
Temperature:
32 to 122°F (0 to 50°C)
Process Connection:
1 inch MNPT
Wetted Parts:
Polysulfone, Viton
1
, Teflon
2
TFE, silicone
Cathode:
gold
Type
PN
Wetted Materials
Process Connection Maximum Temperature Maximum Pressure
1-1/2 in. tee
23567-00
CPVC, Buna N
1-1/2 in. socket
122°F (50°C)
65 psig (549 kPa abs)
915240-03
3/4 in. NFPT
2 in. tee
915240-04
PVC, Buna N
1 in. NFPT
120°F (49°C)
60 psig (515 kPa abs)
915240-05
1-1/2 in. NFPT
Low Flow*
24091-00
Polyester/polycarbonate, Compression fitting for
158°F (70°C)
90 psig (722 kPa abs)
316SS, silicone
1/4 in. OD tubing
1 Viton is a registered trademark of E.I. duPont de Nemours & Co.
2 Teflon is a registered trademark of E.I. duPont de Nemours & Co.
SPECIFICATIONS - FLOW CELLS
For additional information, please visit our website
at
www.emersonprocess.com/raihome/liquid/
.
* Pressure and temperature specifications for the low flow cell exceed the temperature and pressure specifications for the sensor.
CAUTION
Do not exceed pressure and temperature specifications.
Pressure: 65 psig max (549 kPa abs max)
Temperature: 32 to 122°F (0 to 50°C)
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
The wetted sensor materials may not be compatible
with process com
position and operating conditions.
Application compat ibility is entirely the responsibility of
the user.
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
2 MODEL 499A OZ INSTALLATION INSTALLATION FIGURE 1. Sensor Orientation FIGURE 2. Flow-Through 2 in. Tee FIGURE 3. Flow-Through 1-1/2 in. Tee Flow through 1 to 5 gpm (3.8 to 19 L/min) Open channel 1 ft/sec (0.3 m/sec) Low flow cell 2 to 5 gph (7.6 to 19 L/hr) The Model 499AOZ sensor is NOT recommende...
MODEL 499A OZ WIRING WIRING 3 FIGURE 5. Wiring Sensor with Standard Cable toModel 54eA Analyzer FIGURE 6. Wiring Sensor with OptimumEMI/RFI orVariopol Cable to Model 54eA Analyzer FIGURE 7. Wiring Sensor with Standard Cable toModel 1056 and Model 56 Analyzer FIGURE 8. Wiring Sensor with Optimum EMI/...
FIGURE 13. Wiring Sensor with Standard Cable toModel 1066 Transmitter. FIGURE 14. Wiring Sensor with Optimum EMI/RFIor Variopol Cable to Model 1066 Transmitter. CALIBRATION Zero point: Even in the absence of ozone, the 499AOZ sensor generates a small signal called the zero current.Failing to correct...