Pressure and Temperature:

Wetted materials:

Unfilled PEEK, 316L stainless steel, EP (option 22 only). PEEK meets 21CFR177.2415. EP meets

21CFR177.2600.

Surface finish:

16 microinch (0.4 micron) Ra (except electrode surfaces)

Steam sterilization:

tolerates SIP to 284°F (140°C)

Model 410VP

Four-electrode Conductivity Sensor

Instruction Sheet

PN 51A-410VP/rev.D
February 2013

CAUTION

SENSOR/PROCESS APPLICATION COMPATIBILITY

The wetted sensor materials may not be compatible
with process com position and operating conditions.
Application compat ibility is entirely the responsibility
of the user.

INSTALLATION

SENSOR SPECIFICATIONS

The 410VP sensor is available with four process connections: 1½ inch Tri-Clamp

1

, 2-inch Tri-Clamp, G 1¼, and Varivent N

2

.

Gaskets and clamps for the Tri-Clamp and Varivent connections must be supplied by the user.

Install the sensor so that the electrodes are completely immersed in the process liquid. Avoid installing the sensor in places
where air bubbles are likely to get trapped or sediment is likely to accumulate on the electrodes. Generally, mounting the
sensor in a vertical pipe run is best. If the sensor must be installed in a horizontal pipe, place the sensor in the 3 o’clock
position. Keep at least 1.0 inch (25 mm) clearance between the end of the sensor and the opposite pipe wall.

To keep response time as fast as possible, do not install the sensor in dead legs or areas where circulation is poor.

1Tri-Clamp is a registered trademark of Alfa Laval, Inc.
2Varivent is a registered trademark of Tuchenhagen, GmbH.

WARNING

Before removing the sensor, be absolutely certain
that the process pressure is reduced to 0 psig and
the process temperature is lowered to a safe level!

CAUTION

For additional information, please refer to the Instruction Manuals CD shipped with this
product, or visit our website at

www/rosemountanalytical

.

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - Current is; WIRING

Figure 1 . Wire color and functions. Current is injected through the gray and orange wires. Voltageis measured across the white and black wires. Figure 2. Model 1056 and Model 56 wiring. WIRING MODEL 410VP WIRING 2 Figure 3.Model 1066 wiring Figure 4. Model 6081-C wiring The terminal end of the sens...

Page 3 - CALIBRATION; SETUP AND CALIBRATION; PROBLEM

3 MODEL 410VP CALIBRATION The sensor is calibrated at the factory and does not require initial user-calibration. Simply configure the analyzerto accept a four-electrode sensor and enter the cell constant and calibration factor printed on the label. After a period of service, the sensor may require c...

Page 4 - ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE; Specifications subject to change without notice.; Emerson Process Management

Credit Cards for U.S. Purchases Only. The right people,the right answers,right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE http://www.rosemountanalytical.com Specifications subject to change without notice. 8 Emerson Process Management 2400 Barranca ParkwayIrvine, CA 92606 USATel: (949) 757-...