Instruction Manual
403 and 403VP
LIQ-MAN-ABR-403-403VP
February 2015
PURSense
™
Conductivity Sensors
CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY
The wetted sensor materials may not be compatible with process com position and operating conditions.
Application compat ibility is entirely the responsibility of the user.
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process
temperature is lowered to a safe level!
Specifications - Sensor
Specifications
403 and 403VP
Wetted Materials
titanium, PCTFE, 316 SS, EP
Temperature Range
32–221 °F (0–105 °C) Sensors tolerate
steam sterilization to 135 °C
Maximum Pressure
250 psig (1825 kPa abs)
NOTE
Elastomers and fluorocarbon resins are compatible with 21CFR177. Elastomers also meet the requirements of USP
Class VI. Stainless steel contains <5 % delta ferrite. All surfaces in have 16 microinch (0.4 micrometer) Ra finish.
Installation
Depending on the option selected, the sensor can be installed in either a 1 ½-inch or 2-inch Tri-Clamp tee. The gasket,
clamp, and tee must be supplied by the user. The electrodes must be completely submerged in the process liquid, i.e., up
to the inside surface of the flange.
If the sensor is installed in a side stream with the sample draining to open atmosphere, bubbles may accumulate on the
electrodes. Trapped bubbles will cause errors. Normally, as bubbles accumulate the conductivity reading drifts down. To
control bubble formation, apply a small amount of back pressure to the sensor.
For additional information, please visit our website at www.rosemountanalytical.com.
Figure 1. Sensor Orientation
°
°
Figure 2. Recommended Installation
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
2 403 and 403VP Instruction Manual February 2015 LIQ-MAN-ABR-403-403VP Wiring NOTE For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or the Manual DVD enclosed with each product. 1056, 1057, 56, 5081, 60...
3 Instruction Manual 403 and 403VP LIQ-MAN-ABR-403-403VP February 2015 Figure 8. Wiring for 5081-C Figure 9. Wiring for 6081-C Figure 5. Wiring for 1066 Figure 6. Wiring (Panel) for Xmt-C-10 Figure 7. Wiring(Pipe or Wall) for Xmt-C-11 WHEN INCH AND METRIC DIMS ARE GIVEN INCH MILLIMETER WIRING MODEL ...
4 Wiring through a Junction Box If wiring connections are made through a remote junction box (PN 23550-00), wire point-to-point. Use cable23747-00 (factory-terminated) or 9200275 (no terminations). Pin Out Diagram for 400VP Figure 10. VP pin-out (viewed from connector end of sensor, looking down) Cl...