Emerson 402VP - Manual
Emerson 402VP – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.
Table of Contents:
- Page 2 – 02 and 402VP Instruction Manual; Installation and Retraction of Sensor; The sensor assembly includes items 8 through 11.; Figure 3. 402 Sensor with Retraction Assembly Kit
- Page 3 – Instruction Manual; January 2014; Figure 4. 402 Sensor Installation
- Page 4 – Retraction; working area for retraction.; Figure 5. 402 Sensor in Retracted Position
- Page 5 – AY; Figure 6. Wiring for Sensor-Mounted Junction Box; Wiring; Note; Wire Color and Connections in Sensor; Wiring Diagrams; TB
- Page 7 – Wiring through a Junction Box; Cleaning the Sensor; Calibration; on the Rosemount Analytical website.; Troubleshooting; Problem
- Page 8 – Figure 15. Checking Temperature Element
Instruction Manual
402 and 402VP
LIQ_MAN_ABR_402_402VP
January 2014
ENDURANCE
™
Conductivity Sensors
For additional information, please visit our website at www.rosemountanalytical.com
CAUTION
Sensor/Process Application Compatibility
The wetted sensor materials may not be compatible with process composition and operating conditions.
Application compatibility is entirely the responsibility of the user.
CAUTION
The 402 and 402VP sensors are retractable. BEFORE RETRACTING THE SENSOR, be absolutely certain the process pressure
is less than 64 psig (542 kPa abs) and the process temperature is at a safe level!
Sensor Specifications
Specifications
402
402VP
Electrode
-11, -12, -13 titanium
-11, -12, -13 titanium
Insulator
-11, -12, -13 PEEK (glass filled)
-11, -12, -13 PEEK (glass filled)
Other Wetted Materials
316 SST, EPDM, graphite, Neoprene
316 SST, EPDM, graphite, Neoprene
Temperature Range
32–212 °F (0–100 °C)
32–212 °F (0–100 °C)
Maximum Pressure
200 psig (1481 kPa abs)
200 psig (1481 kPa abs)
Maximum Retraction Pressure
64 psig (542 kPa abs)
64 psig (542 kPa abs)
Sensor Orientation
Keep ¼ in. (6 mm) clearance between electrodes and piping. The electrolytes must be completely submerged in the
process liquid, i.e., to the upper edge of the guard (item 10 in Figure 3).
Figure 1. Sensor Orientation
°
°
30°
30°
trapped air
trapped sludge
Figure 2. Recommended Installation
recommended
not
recommended
flow
flow
flow
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
2 402 and 402VP Instruction Manual January 2014 LIQ_MAN_ABR_402_402VP Installation and Retraction of Sensor Installation Retraction assembly kit (PN 23765-00) consists of items 1 through 7 shown in Figure 3 as well as a hex key and pipe tape. The sensor assembly includes items 8 through 11. IMPORTAN...
3 Instruction Manual 402 and 402VP LIQ_MAN_ABR_402_402VP January 2014 1. Make sure the system is shut down and there is no residual pressure. 2. Attach the ball valve (2) to the process piping using the 1–¼-inch FNPT port on the valve or the 1–¼ inch NPT nipple (1). Use pipe tape on male threads. 3....
4 402 and 402VP Instruction Manual January 2014 LIQ_MAN_ABR_402_402VP Retraction 1. Make sure the system pressure is at or below 542 kPa abs (64 psig) at the valve before proceeding. Provide adequate working area for retraction. 2. Unscrew the retainer (5) from the packing adapter (3). If it is diff...