Emerson 402 - Manual

Emerson 402

Emerson 402 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
Page: / 8

Table of Contents:

  • Page 2 – 02 and 402VP Instruction Manual; Installation and Retraction of Sensor; The sensor assembly includes items 8 through 11.; Figure 3. 402 Sensor with Retraction Assembly Kit
  • Page 3 – Instruction Manual; January 2014; Figure 4. 402 Sensor Installation
  • Page 4 – Retraction; working area for retraction.; Figure 5. 402 Sensor in Retracted Position
  • Page 5 – AY; Figure 6. Wiring for Sensor-Mounted Junction Box; Wiring; Note; Wire Color and Connections in Sensor; Wiring Diagrams; TB
  • Page 7 – Wiring through a Junction Box; Cleaning the Sensor; Calibration; on the Rosemount Analytical website.; Troubleshooting; Problem
  • Page 8 – Figure 15. Checking Temperature Element
Loading the manual

Instruction Manual

402 and 402VP

LIQ_MAN_ABR_402_402VP

January 2014

ENDURANCE

Conductivity Sensors

For additional information, please visit our website at www.rosemountanalytical.com

CAUTION
Sensor/Process Application Compatibility

The wetted sensor materials may not be compatible with process composition and operating conditions.
Application compatibility is entirely the responsibility of the user.

CAUTION

The 402 and 402VP sensors are retractable. BEFORE RETRACTING THE SENSOR, be absolutely certain the process pressure
is less than 64 psig (542 kPa abs) and the process temperature is at a safe level!

Sensor Specifications

Specifications

402

402VP

Electrode

-11, -12, -13 titanium

-11, -12, -13 titanium

Insulator

-11, -12, -13 PEEK (glass filled)

-11, -12, -13 PEEK (glass filled)

Other Wetted Materials

316 SST, EPDM, graphite, Neoprene

316 SST, EPDM, graphite, Neoprene

Temperature Range

32–212 °F (0–100 °C)

32–212 °F (0–100 °C)

Maximum Pressure

200 psig (1481 kPa abs)

200 psig (1481 kPa abs)

Maximum Retraction Pressure

64 psig (542 kPa abs)

64 psig (542 kPa abs)

Sensor Orientation

Keep ¼ in. (6 mm) clearance between electrodes and piping. The electrolytes must be completely submerged in the
process liquid, i.e., to the upper edge of the guard (item 10 in Figure 3).

Figure 1. Sensor Orientation

°

°

30°

30°

trapped air

trapped sludge

Figure 2. Recommended Installation

recommended

not

recommended

flow

flow

flow

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - 02 and 402VP Instruction Manual; Installation and Retraction of Sensor; The sensor assembly includes items 8 through 11.; Figure 3. 402 Sensor with Retraction Assembly Kit

2 402 and 402VP Instruction Manual January 2014 LIQ_MAN_ABR_402_402VP Installation and Retraction of Sensor Installation Retraction assembly kit (PN 23765-00) consists of items 1 through 7 shown in Figure 3 as well as a hex key and pipe tape. The sensor assembly includes items 8 through 11. IMPORTAN...

Page 3 - Instruction Manual; January 2014; Figure 4. 402 Sensor Installation

3 Instruction Manual 402 and 402VP LIQ_MAN_ABR_402_402VP January 2014 1. Make sure the system is shut down and there is no residual pressure. 2. Attach the ball valve (2) to the process piping using the 1–¼-inch FNPT port on the valve or the 1–¼ inch NPT nipple (1). Use pipe tape on male threads. 3....

Page 4 - Retraction; working area for retraction.; Figure 5. 402 Sensor in Retracted Position

4 402 and 402VP Instruction Manual January 2014 LIQ_MAN_ABR_402_402VP Retraction 1. Make sure the system pressure is at or below 542 kPa abs (64 psig) at the valve before proceeding. Provide adequate working area for retraction. 2. Unscrew the retainer (5) from the packing adapter (3). If it is diff...

Other Emerson Models