Emerson 369P - Manual

Emerson 369P

Emerson 369P – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12
13 Page 13
14 Page 14
15 Page 15
16 Page 16
17 Page 17
18 Page 18
19 Page 19
20 Page 20
21 Page 21
22 Page 22
23 Page 23
24 Page 24
25 Page 25
26 Page 26
27 Page 27
28 Page 28
29 Page 29
30 Page 30
31 Page 31
32 Page 32
33 Page 33
34 Page 34
35 Page 35
36 Page 36
37 Page 37
38 Page 38
39 Page 39
40 Page 40
Page: / 40

Table of Contents:

  • Page 2 – Instruction Sheet; A pH glass electrode does have a limited shelf life of one year.; Note
  • Page 3 – July 2013; ORP Calibration; Remove the sensor from the buffer, rinse, and install in the process.; Maintenance for ORP Electrode; pH 4 Solution; Installation; Figure 1. Dimensional Drawing
  • Page 4 – Figure 2. Flow Through and Insertion Installation
  • Page 5 – Figure 5. Wiring 81 through Remote Junction Box
  • Page 7 – Figure 10. Wiring 2081 through Remote Junction Box
  • Page 9 – Figure 14. Wiring 54 through Remote Junction Box
  • Page 36 – Figure 67. Wiring Details of 396PVP Gray Cable for use with 6081
  • Page 39 – Part; Accessories
Loading the manual

Instruction Sheet

TUpH 396P and 396PVP Sensors

LIQ_MAN_ABR_396P_396PVP

July 2013

Submersion/Insertion pH/ORP Sensors

Specifications

Sensor Type:

TUpH 396P, 396 PVP; GPLR hemi or

flat glass

Range:

*pH: 0–14

ORP:

-1500 to 1500 mv

Percent Linearity*

pH Range

GPLR hemi bulb

GPLR flat bulb

0 – 2 pH

94%

2 – 12 pH

97%

98%

12 – 13 pH

98%

95%

13 – 14 pH

98%

Pressure Range-Hemi bulb:

100–1136 kPa [abs]

(0–150 psig)

Temperature Range

: 0–100 °C (32-212 °F)

Pressure Range-Flat bulb:

100–790 kPa [abs]

(0–100 psig)

Minimum Conductivity:

75 µS/cm, min; 100 µS/cm

nominal

Flow Rate:

up to 2 ft/sec

Wetted Materials:

Titanium, Polypropylene, EPDM,

glass; platinum (ORP only)

Process Connection:

1 in. MNPT front and rear

facing threads

Integral Cable:

396P: (-01) -25 ft; (02) -15 ft coaxial

VP8 Cable:

use PN 24281-00 15ft VP8 cable

Preamplifier Options:

396P: Remote or Integral SMART (-01)
396PVP: Remote or Integral SMART (-70)

Weight/Shipping Weight:

0.45 kg/0.9 kg (1 lb/2 lb)

WARNING

Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process
temperature is lowered to a safe level!

WARNING

Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.

CAUTION
SENSOR/PROCESS APPLICATION COMPATIBILITY

The wetted sensor materials may not be compatible with process com position and operating conditions. Application
compat ibility is entirely the responsibility of the user.

CAUTION

The buffer in the protective boot may cause skin or eye irritation.

ATEX DIRECTIVE

Special Conditions for safe use
1. All pH/ORP sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the danger due to a

build up of an electrostatic charge.

2. All pH/ORP sensor Models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. a.c. test

to earth. This must be taken into consideration at installation.

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - Instruction Sheet; A pH glass electrode does have a limited shelf life of one year.; Note

2 TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP Storage 1. It is recommended that electrodes be stored in their original shipping containers until needed. 2. Do not store at temperatures below -5 °C (23 °F). 3. Electrodes should be stored with a protective cap cont...

Page 3 - July 2013; ORP Calibration; Remove the sensor from the buffer, rinse, and install in the process.; Maintenance for ORP Electrode; pH 4 Solution; Installation; Figure 1. Dimensional Drawing

3 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 ORP Calibration 1. After making an electrical connection between the sensor and the instrument, obtain a standard solution of saturated quinhydrone. This can also be made quite simply by adding a few crystals of quinh...

Page 4 - Figure 2. Flow Through and Insertion Installation

4 TUpH 396P and 396PVP Sensors Instruction Sheet July 2013 LIQ_MAN_ABR_396P_396PVP Figure 2. Flow Through and Insertion Installation Figure 3. Wiring for 81 Figure 4. Wiring for 1181

Other Emerson Models