Page 3 - PRECAUTIONS FOR SAFE OPERATION; WARNING
2 PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the u...
Page 5 - CAUTION
4 PRECAUTIONS FOR SAFE OPERATION Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electric shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance. These may be harmfu...
Page 7 - CAUTIONS FOR USAGE
6 CAUTIONS FOR USAGE 1. If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes. This protects the compressor. 2. This inner cabinet is refrigerated by the forced circulation of cooled air inside the chamber. Ensure that the intake and exhaust...
Page 8 - Model with CE mark; ENVIRONMENTAL CONDITIONS
7 CAUTIONS FOR USAGE Model with CE mark This unit is equipped with a circuit breaker on the back. Make sure this breaker is switched ON before the unit starts to run. When the operation of the unit is stopped by this breaker, contact with dealer or service station after disconnected the power supply...
Page 9 - NAME AND FUNCTION OF PARTS
8 NAME AND FUNCTION OF PARTS 1. Glass window: Water can sometimes condense on the glass in areas of high humidity. Wipe off the condensation with a dry soft cloth. 2. Magnetic door gasket: Always keep clean. 3. Control panel: The panel is divided into two portion; Upper portion is an indicator and l...
Page 10 - Control panel
9 NAME AND FUNCTION OF PARTS Control panel 1. Alarm buzzer stop key (BZ): Press this key to silence the buzzer in the event that the alarm operates and buzzer sounds. 2. Display select key (REF./FREEZ.): By pressing this key, internal temperature display of refrigerator or freezer and each setting i...
Page 11 - BEFORE COMMENCING OPERATION; Installation site
10 BEFORE COMMENCING OPERATION Installation site To operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions: 1. A location not subjected to direct sunlight Installation in a location subjected to direct sunlight may lead to inadequat...
Page 12 - Installation
11 BEFORE COMMENCING OPERATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a neutral detergent and wipe it up with a wet cloth. 2. Adjust the ...
Page 13 - OPERATING INSTRUCTIONS
12 OPERATING INSTRUCTIONS Control panel 1. Setup of chamber temperature Table 1 shows the basic operation method. Perform key operation in the sequence indicated in the table. The example in the table is based on the assumption that the refrigerator temperature is 4 o C and the freezer temperature i...
Page 15 - Alarm and safety functions
14 OPERATING INSTRUCTIONS Alarm and safety functions This unit has the alarm and safety functions shown in Table 3, and also a self diagnostic function. Kind of alarm or safety Situation Indication Buzzer Safety operation High temperature alarm ・ Refrigerator If the chamber temperature deviates from...
Page 16 - Defrost cycles
15 OPERATING INSTRUCTIONS Defrost cycles Refrigerator There is no need for routine defrosting of the unit as this occurs automatically as follows: 1. Cycle defrost To keep the temperature stable inside the chamber, the refrigeration compressor is cycled on and off. During “off” periods any frost whi...
Page 17 - TEMPERATURE RECORDER; Automatic temperature recorder
16 TEMPERATURE RECORDER Automatic temperature recorder To record the chamber temperature, an optional automatic temperature recorder is available. Please consult with our sales department or agency for recorder installation. For the proper usage of temperature recorder, refer to an instruction manua...
Page 18 - Attachment of temperature recorder
17 TEMPERATURE RECORDER Attachment of temperature recorder For refrigerator 1. Attach the fixture to the temperature recorder with two screws. (Fig. 1) 2. Remove the screws at the right corner on the top of the unit and fix the recorder to the position. (Fig. 2) 3. Remove a rubber cap and insulation...
Page 19 - REMOTE ALARM TERMINAL; Connection of remote alarm terminal; MAINTENANCE; Cleaning
18 REMOTE ALARM TERMINAL WARING Always disconnect the power supply to the unit before connection of alarm equipment to remote alarm terminals to prevent electric shock or injury. The terminal of the remote alarm is installed in the terminal cover at rear of the cabinet. The maximum contact rating is...
Page 21 - SPECIFICATIONS
20 SPECIFICATIONS Name Pharmaceutical Refrigerator with Freezer Model MPR-411F/MPR-411FR Refrigerator Freezer External dimensions W800 x D600 x H1805 (mm) Internal dimensions W720 x D495x H1425 (mm) W317 x D440x H576 (mm) Effective capacity 340 L 82 L Exterior Acrylic finish baked on...
Page 22 - PERFORMANCE
21 PERFORMANCE Refrigerator Freezer Control range 2 o C to 14 o C (Ambient temp.; -5 o C to +35 o C, No load) -10 o C to -30 o C (Ambient temp.; -5 o C to +35 o C, No load) Noise level 39 dB (A scale) Maximum pressure 1705 kPa 1901 kPa Rated voltage AC 110 V AC 115 V AC 220 V AC 220 V AC 230...
Page 23 - Safety check sheet
22 CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety check sheet 1. Refrigerator contents : Risk of infection: Risk of toxicity: Risk from radioactive sources: (List all potentially hazardous materials that have bee...