Page 2 - WARNING; WARNING; Invacare products are specifically designed and; DANGER; The information contained in this document is subject to
WARNING Oxygen Conserving Cylinder 2 Part No 1143247 WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR A QUALIFIED TECHNICIAN BEFORE ATTEMPT...
Page 3 - TABLE OF CONTENTS
TABLE OF CONTENTS Part No 1143247 3 Oxygen Conserving Cylinder TABLE OF CONTENTS SPECIAL NOTES .............................................. 4RECEIVING INSPECTION .................................. 4INTENDED USE ................................................ 4SECTION 1—GENERAL GUIDELINES ...........
Page 4 - SPECIAL NOTES; SIGNAL WORD
SPECIAL NOTES Oxygen Conserving Cylinder 4 Part No 1143247 SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words. RECEIVING INSPECTION Remo...
Page 5 - SECTION 1—GENERAL GUIDELINES; “Caution: Federal law restricts this device to sale by or on the; Only personnel instructed and trained in the use of the
SECTION 1—GENERAL GUIDELINES Part No 1143247 5 Oxygen Conserving Cylinder SECTION 1—GENERAL GUIDELINES WARNING SECTION 1 - IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first com...
Page 6 - Cylinders MUST be used only by the oxygen patient and are
SECTION 1—GENERAL GUIDELINES Oxygen Conserving Cylinder 6 Part No 1143247 WARNING DO NOT attempt to repair product. DO NOT immerse product in any kind of liquid. Store product in a clean area when not in use. Only use Invacare Corporation HF2PCM2 carrying bag.NEVER smoke in an area where oxygen is b...
Page 8 - SECTION 2—FEATURES; This product is not intended as a life-sustaining or
SECTION 2—FEATURES Oxygen Conserving Cylinder 8 Part No 1143247 SECTION 2—FEATURES DANGER This product is not intended as a life-sustaining or life-supporting device. FIGURE 2.1 Product Diagram Outlet Connection Cannula Flow Dial Cylinder Gauge Indicator Pointer Cylinder Fill Port
Page 9 - SECTION 3—TYPICAL PRODUCT PARAMETERS; SECTION 3—TYPICAL PRODUCT
SECTION 3—TYPICAL PRODUCT PARAMETERS Part No 1143247 9 Oxygen Conserving Cylinder SECTION 3—TYPICAL PRODUCT PARAMETERS CONSERVER S/N RANGE: 75,000+ INLET PRESSURE RANGE: 300 - 2100 psig INTERNAL REGULATED PRESSURE: 50 psig DIMENSIONS: HF2PCM2KIT: Weight 2.0 lbs (.90 kg) HF2PCM2KIT: Overall Length 8....
Page 10 - SECTION 4—INSPECTION AND COMPRESSOR INTERFACE; SECTION 4—INSPECTION AND; NOTE: Refer to the HomeFill Compressor Owner’s Manual for; Cylinder Prefill Inspection; All cylinders MUST be inspected before attempting to fill; The Hydrostatic Testing Date; Aluminum cylinders MUST undergo testing every five years.; DO NOT fill cylinders that have not been tested in the
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conserving Cylinder 10Part No 1143247 SECTION 4—INSPECTION AND COMPRESSOR INTERFACE NOTE: Refer to the HomeFill Compressor Owner’s Manual for compressor operating instructions. Cylinder Prefill Inspection WARNING All cylinders MUST be inspected be...
Page 11 - External Examination
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 11 Oxygen Conserving Cylinder External Examination 1. Examine the outside of the cylinder for the following conditions, and replace the cylinder if they exist:• Dents or dings• Arc Burns• Oil or Grease• Any other signs of damage that migh...
Page 13 - Connecting/Disconnecting the Cylinder; Connecting the Cylinder to the Compressor; NEVER use tools of any kind to connect/disconnect the; CAUTION; DO NOT connect the outlet (cannula) connection to the
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 13 Oxygen Conserving Cylinder Connecting/Disconnecting the Cylinder to/from the Compressor Connecting the Cylinder to the Compressor WARNING NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, sev...
Page 14 - be able to accept the cylinder fillport. Pushing DOWN; Resetting Connector Fillport
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conserving Cylinder 14 Part No 1143247 4. Momentarily push DOWN on the outer ring (sleeve) of the connector fillport to reset the connector as shown in FIGURE 4.2. NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), th...
Page 15 - cover are not shown for clarity.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 15 Oxygen Conserving Cylinder FIGURE 4.3 Connecting the Cylinder to the Compressor Cylinder/Regulator Assembly Compressor Cylinder Gauge Cylinder Fillport Connector Fillport PULL UP NOTE: Cylinder bag, cylinder fillport cover and connecto...
Page 16 - Disconnecting the Cylinder from the Compressor
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conserving Cylinder 16 Part No 1143247 Disconnecting the Cylinder from the Compressor WARNING NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. NOTE: For this proce...
Page 17 - The fillport cover on the connector fillport and the
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 17 Oxygen Conserving Cylinder 5. When the cylinder fillport is disconnected from the connector fillport, release the outer ring (sleeve) of the connector fillport and use two hands to remove the cylinder from the compressor cradle. WARNIN...
Page 18 - Changing the Carrying Position of the; when cleaning. The cylinder MUST be in the bag to ensure the
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conserving Cylinder 18 Part No 1143247 Changing the Carrying Position of the Cylinder NOTE: For this procedure, refer to FIGURE 4.6 on page 19.The cylinder MUST be kept in the cylinder bag at all times except when cleaning. The cylinder MUST be in...
Page 19 - ii. Thread the end of the belt that has no buckle and
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 19 Oxygen Conserving Cylinder ii. Thread the end of the belt that has no buckle and retaining loop through the two loops on the cylinder bag. See Detail “B”. NOTE: The web material on the belt should be facing away from the cylinder bag. ...
Page 20 - Cleaning the Cylinder Bag; NOTE: Skip STEP one if the belt is to be cleaned with the cylinder
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conserving Cylinder 20Part No 1143247 Cleaning the Cylinder Bag NOTE: Skip STEP one if the belt is to be cleaned with the cylinder bag. 1. Unhook and remove the cylinder bag and belt from body.2. Re-attach the buckle. NOTE: Either end of the belt ...
Page 21 - SECTION 5—OPERATING INSTRUCTIONS; SECTION 5—OPERATING; NEVER smoke in an area where oxygen is being; DO NOT direct flow of oxygen at any person, or
SECTION 5—OPERATING INSTRUCTIONS Part No 1143247 21 Oxygen Conserving Cylinder SECTION 5—OPERATING INSTRUCTIONS WARNING NEVER smoke in an area where oxygen is being administered.NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere. NOTE: For this procedure, refer ...
Page 22 - Cannula; When pointer on contents gauge enters red zone, it is
SECTION 5—OPERATING INSTRUCTIONS Oxygen Conserving Cylinder 22 Part No 1143247 5. Place the cannula over your ears and position the prongs in your nose as instructed by your Health Care provider or cannula manufacturer. Refer to FIGURE 5.1. 6. Turn dial until indicating pointer is aligned with the p...
Page 23 - SECTION 6—MAINTENANCE; After each use, clean exterior of the product with a dry, lint
SECTION 6—MAINTENANCE Part No 1143247 23 Oxygen Conserving Cylinder SECTION 6—MAINTENANCE 1. After each use, clean exterior of the product with a dry, lint free cloth only. 2. Store product in a clean area free from grease, oil, and other sources of contamination. CAUTION DO NOT use cleaning solutio...
Page 24 - SECTION 7—CYLINDER FILL TIMES; NOTE: All filling times are approximate and may vary on; Troubleshooting; If Conserving Cylinder fails to function, consult the; PROBLEM
SECTION 7—CYLINDER FILL TIMES Oxygen Conserving Cylinder 24 Part No 1143247 SECTION 7—CYLINDER FILL TIMES NOTE: All filling times are approximate and may vary on environmental conditions. * Troubleshooting If Conserving Cylinder fails to function, consult the Troubleshooting guide. If problem cannot...
Page 25 - NOTES
SECTION 7—CYLINDER FILL TIMES Part No 1143247 25 Oxygen Conserving Cylinder NOTES
Page 26 - LIMITED WARRANTY
LIMITED WARRANTY Oxygen Conserving Cylinder 26 Part No 1143247 LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.This warranty is extended only to the original purchaser/user of our products.This warran...