Page 2 - WARNING
HomeFill ® Cylinder 2 Part No 1163156 WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS ...
Page 3 - TABLE OF CONTENTS
TABLE OF CONTENTS Part No 1163156 3 HomeFill ® Cylinder TABLE OF CONTENTS SPECIAL NOTES ............................................... 4RECEIVING INSPECTION .................................. 4INTENDED USE ................................................ 4SECTION 1—GENERAL GUIDELINES .................
Page 4 - SPECIAL NOTES; NOTICE; RECEIVING INSPECTION; SIGNAL WORD
SPECIAL NOTES HomeFill ® Cylinder 4 Part No 1163156 SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definit...
Page 5 - SECTION 1—GENERAL GUIDELINES; SECTION 1 - GENERAL GUIDELINES contains; This product is not intended as a life-sustaining or
SECTION 1—GENERAL GUIDELINES Part No 1163156 5 HomeFill ® Cylinder SECTION 1—GENERAL GUIDELINES WARNING SECTION 1 - GENERAL GUIDELINES contains important information for the safe operation and use of this product. DANGER This product is not intended as a life-sustaining or life-supporting device. WA...
Page 8 - SECTION 2—FEATURES; Product Diagram
SECTION 2—FEATURES HomeFill ® Cylinder 8 Part No 1163156 SECTION 2—FEATURES FIGURE 2.1 Product Diagram Cylinder Fill Port Cylinder Valve 540 Connection Port (with Protective Cap)
Page 9 - SECTION 3—TYPICAL PRODUCT PARAMETERS
SECTION 3—TYPICAL PRODUCT PARAMETERS Part No 1163156 9 HomeFill ® Cylinder SECTION 3—TYPICAL PRODUCT PARAMETERS DIMENSIONS: Weight 5.1 lbs (2.31 kg) Overall Length 13.25 in (33.6 cm) Diameter 4.4 in (11.1 cm) STORAGE CONDITIONS TEMPERATURE: -40°F (-40°C) to 140°F (60°C) MAXIMUM HUMIDITY: 95%
Page 10 - SECTION 4—INSPECTION AND COMPRESSOR INTERFACE; Cylinder Prefill Inspection; All cylinders MUST be inspected before attempting to fill; The Hydrostatic Testing Date; DO NOT fill cylinders that have not been tested in the
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 10 Part No 1163156 SECTION 4—INSPECTION AND COMPRESSOR INTERFACE NOTE: Refer to the HomeFill ® Compressor Owner’s Manual for compressor operating instructions. Cylinder Prefill Inspection WARNING All cylinders ...
Page 11 - External Examination
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1163156 11 HomeFill ® Cylinder External Examination 1. Examine the outside of the cylinder for the following conditions and replace the cylinder if they exist: • Dents or dings • Arc burns • Oil or grease...
Page 12 - Checking Cylinder Pressure; NEVER use tools of any kind to connect/disconnect the
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 12 Part No 1163156 Checking Cylinder Pressure WARNING NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur.DO NOT drop oxygen cylinders. Use two hands when...
Page 13 - Filling the Cylinder; Component Identification
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1163156 13 HomeFill ® Cylinder Filling the Cylinder NOTE: For this procedure, refer to FIGURE 4.1 thru FIGURE 4.6. 1. Check the cylinder pressure before attempting to fill cylinder. Refer to Checking Cy...
Page 14 - Fillport Covers
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 14 Part No 1163156 FIGURE 4.2 Fillport Covers 5. On the compressor unit, momentarily push DOWN on the outer ring (sleeve) of the connector fillport to reset the connector (FIGURE 4.3). Connect...
Page 15 - Resetting Connector Fillport
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1163156 15 HomeFill ® Cylinder NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing...
Page 16 - Connecting the Cylinder to the Compressor
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 16 Part No 1163156 NOTE: Refer to Cylinder Fill Times on page 20 for the length of time it will take to fill the cylinder. FIGURE 4.4 Connecting the Cylinder to the Compressor Cylinder Compressor ...
Page 17 - Disconnecting the Cylinder from the Compressor
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1163156 17 HomeFill ® Cylinder 11. After the cylinder is full, press the compressor power switch to the Off (O) position (FIGURE 4.5). 12. Grasp the cylinder assembly (FIGURE 4.5). 13. With the other ...
Page 18 - The fillport cover on the connector fillport and the post
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 18 Part No 1163156 WARNING The fillport cover on the connector fillport and the post valve fillport MUST be replaced after filling and whenever not in use. If either of the fillport covers are missing contact your healthcare provider ...
Page 19 - SECTION 5—MAINTENANCE; CAUTION
SECTION 5—MAINTENANCE Part No 1163156 19 HomeFill ® Cylinder SECTION 5—MAINTENANCE 1. After each use, clean exterior of the product with a dry, lint free cloth only. 2. Store product in a clean area free from grease, oil, and other sources of c...
Page 20 - SECTION 6—CYLINDER FILL TIMES; Cylinder Filling; PROBLEM
SECTION 6—CYLINDER FILL TIMES HomeFill ® Cylinder 20 Part No 1163156 SECTION 6—CYLINDER FILL TIMES NOTE: All filling times are approximate and may vary on environmental conditions. * Troubleshooting If post valve cylinder fails to function, try the following ...
Page 21 - NOTES
Page 22 - LIMITED WARRANTY
LIMITED WARRANTY HomeFill ® Cylinder 22 Part No 1163156 LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.This warranty is extended only to the original purchaser/user of our products.This warranty give...