Page 2 - CAUTION
2 VENTURE PART NO. 1093845 WARNING CAUTION Caution: Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the State in which he practices to use or order the use of this device. WARNING DO NOT install, maintain or operate this eq...
Page 3 - Invacare Venture Demand Oxygen; TABLE OF CONTENTS
3 PART NO. 1093845 VENTURE SAFETY SUMMARY ............................................. 4UNPACKING/HANDLING ..................................... 8INTRODUCTION ................................................ 10 Invacare Venture Demand Oxygen Delivery Device ...........................................
Page 4 - WARNING
4 VENTURE PART NO. 1093845 SAFETY SUMMARY WARNING This device should ONLY be used under the supervision of a physician. Verify your oxygen flowrate with your physician when you begin using your Demand Oxygen Delivery System.If you have any questions about the oxygen delivery system or its use, call...
Page 7 - DO NOT use a humidifier with the Demand Oxygen
7 PART NO. 1093845 VENTURE WARNING (CONTINUED) DO NOT use a humidifier with the Demand Oxygen Delivery Device.DO NOT expose the Demand Oxygen Delivery Device or system to water or any other liquids.RADIO FREQUENCY INTERFERENCE Most electronic equipment is influenced by Radio Frequency Interference ...
Page 8 - UNPACKING; Check for any obvious damage to the carton or its contents.; AND VENTURE PACK 212EX,
8 VENTURE PART NO. 1093845 UNPACKING / HANDLING UNPACKING 1. Check for any obvious damage to the carton or its contents. (If damage is evident, notify your local Invacare Home Health Care Dealer/Carrier.) 2. Remove all the loose packing from the carton.3. Carefully remove all the components from th...
Page 9 - HANDLING PROCEDURES; NOTE: Unless the Venture Demand Oxygen Delivery Device is to be; INSPECTION; scratches or other damages. Inspect all components.; STORAGE
9 PART NO. 1093845 VENTURE HANDLING PROCEDURES NOTE: Unless the Venture Demand Oxygen Delivery Device is to be used immediately, retain boxes, containers and packing materials for use in returning or storing the equipment until usage is required. INSPECTION 1. Examine exterior of the equipment for ...
Page 10 - INVACARE VENTURE DEMAND OXYGEN; NOTE: Actual Battery use time depends on your breathing rate.; INTRODUCTION
10 VENTURE PART NO. 1093845 INVACARE VENTURE DEMAND OXYGEN DELIVERY DEVICE Ordinary oxygen supply units deliver oxygen continuously. The normal breathing pattern is to inhale about one-third of the time, and exhale two-thirds of the time. This means that continuous flow units may be inefficient dur...
Page 11 - FEATURES; VENTURE DEMAND OXYGEN DELIVERY; NOTE: The OFF button does NOT stop flow delivery and is NOT; Oxygen OUTLET Fitting -
11 PART NO. 1093845 VENTURE FEATURES VENTURE DEMAND OXYGEN DELIVERY DEVICE COMPONENTS AND DEFINITION OF TERMS (FIGURE 1) ON/OFF Buttons - Determines flow selection method. The ON button turns Power ON for normal pulse operation ( ) and the OFF button disengages the electronics for continuous flow o...
Page 12 - Oxygen INLET Fitting; FIGURE 1 - VENTURE DODD; Oxygen OUTLET Fitting; êê; External Power; NOTE: Battery Compartment located on
12 VENTURE PART NO. 1093845 Oxygen INLET Fitting FIGURE 1 - VENTURE DODD FEATURES FEATURES ç Oxygen OUTLET Fitting Pulse LED (GREEN) Attention LED (YELLOW) Battery LED (RED) Low O 2 LED (GREEN) ê ê êê è External Power Input Jack ON(Pulse ModeOperation) OFF (Continuous Mode Operation) é NOTE: Batter...
Page 13 - VENTURE DODD SYSTEM COMPONENTS AND; Carrying Case with Padded Strap -
13 PART NO. 1093845 VENTURE FEATURES VENTURE DODD SYSTEM COMPONENTS AND DEFINITION OF TERMS (FIGURE 2) Cylinder - Lightweight aluminum cylinder provided. Regulator - Provides flowrate control of oxygen delivery with compressed gas cylinders. DISS Fitting - Connection point for direct connect to oxy...
Page 14 - SPECIFICATIONS; ALARMS
14 VENTURE PART NO. 1093845 SPECIFICATIONS OPERATING VOLTAGESEXTERNAL AC/DC POWER Output: 12 + 3 VDC, 100 mA ADAPTER: Connector: 5.5mm OD x 2.1mm ID x 12.0mm No. 2 Connector ALARMS Low Battery (Two [2] levels) Cannula Disconnect (Breath not sensed) Device Malfunction Low / No Gas Indicator INDICATO...
Page 15 - ENVIRONMENTAL; DEVICE DIMENSIONS
15 PART NO. 1093845 VENTURE SPECIFICATIONS ENVIRONMENTAL Operating Temperature: 40 ° to 104 ° F (5 ° to 40 ° C) Storage Temperature: - 40 ° to 158 ° F (-40 ° to 70 ° C) DEVICE DIMENSIONS Length: 5.4-inches (13.7cm) Thickness: 1.6-inches (4.0cm) Width: 3.6-inches (9.1cm) Weight: 16.0 ozs. (454gm)
Page 16 - THE USE OF OXYGEN; NOTE: ONLY a NASAL CANNULA can be used in the PULSE MODE to; SET-UP AND NORMAL PULSE OPERATION
16 VENTURE PART NO. 1093845 THE USE OF OXYGEN The type of oxygen administrating equipment to be used, as well as the flowrate and time of usage, must be prescribed by a physician. Your physician should verify your Venture DODD flowrate prescription. WARNING If at anytime the attendant or the end us...
Page 18 - If using an Oxygen Cylinder and Regulator:; NOTE: Remaining battery use times are dependent on battery type; OPERATION; OFF
18 VENTURE PART NO. 1093845 6. Set the prescribed flow rate on the oxygen source by performing one (1) of the following:A. If using an Oxygen Cylinder and Regulator: OPEN cylinder valve and adjust regulator to prescribed flow rate. B. If using a Liquid Oxygen Reservoir: Set flow control knob on res...
Page 19 - NOTE: Whenever the pulse mode is used, the battery is monitored.; FIGURE 3
19 PART NO. 1093845 VENTURE NOTE: Whenever the pulse mode is used, the battery is monitored. If the SECOND BATT. LED illuminates, either replace the battery or switch to continuous mode if oxygen supply is adequate. When in continuous mode, the battery is NOT used.NOTE: When in the continuous flow ...
Page 20 - PROLONGING THE LIFE OF YOUR BATTERY; The Venture DODD battery is powered by a single D-cell battery.
20 VENTURE PART NO. 1093845 PROLONGING THE LIFE OF YOUR BATTERY The Venture DODD battery is powered by a single D-cell battery. The Alkaline battery type is recommended for longest life. NOTE: The Alkaline battery should last for approximately 80 hours with normal use, but will last even longer if ...
Page 21 - CARE OF YOUR VENTURE DEMAND; CLEANING THE OUTER CASE; Clean outer case with damp cloth. If necessary, a mild dish; CARRYING CASE; The carrying case can be hand washed in warm water. Air dry; MAINTENANCE
21 PART NO. 1093845 VENTURE CARE OF YOUR VENTURE DEMAND OXYGEN DELIVERY DEVICE AND SYSTEM CLEANING THE OUTER CASE WARNING DO NOT submerge in water to clean.DO NOT open outer case. If outer case is removed by anyone EXCEPT an Invacare Dealer the warranty will be voided. CAUTION Use no oil. Never app...
Page 22 - NOTE: Rechargeable batteries can be used to power Venture DODD,; ACCESSORIES AND SPARE PARTS
22 VENTURE PART NO. 1093845 M6 Aluminum Cylinder, 164L Capacity M9 Aluminum Cylinder, 248L Capacity Piston Regulator, 0.5-6 L/Min., 20 psig, CGA 870 Shoulder Bag, Hunter Green (Dual size for Cylinder diameters of 3 and 4-inches) Venture Device Pouch 12-inch Connection Tubing Barb Fitting, 1/4-28 UN...
Page 23 - TROUBLESHOOTING; PROBLEM
23 PART NO. 1093845 VENTURE TROUBLESHOOTING This is normal operation. Turn unit OFF, check cylinder pressure is above refill limit (500 psig) and post valve is open, check regulator is set to correct flowrate. If problem still exists, call your Home Health Care Dealer or Service Representative IMME...
Page 25 - please contact your Invacare Home Health Care Dealer for service.
25 PART NO. 1093845 VENTURE TROUBLESHOOTING PROBLEM POSSIBLE CAUSE SOLUTIONS External input will not power device (Cont.). "Low" battery light comes on too soon after battery replacement. External power adapter plug does not go into jack in unit. Unit pulses at times other than inhalation. ...
Page 26 - OXYGEN CONSERVED AND BATTERY; COMPATIBILITY; Gaseous or liquid oxygen systems that meet the; CONSERVED
26 VENTURE PART NO. 1093845 OXYGEN CONSERVED AND BATTERY USE TIME BASED ON BREATH RATE i.e. 77% Oxygen conservation over Continuous Flow achieved with Venture DODD in Variable Breath Mode @ 20 BPM Rate. Oxygen supply lasts 4.3-times longer. Venture DODD responds to individual breath rates. Therefor...
Page 27 - LIMITED WARRANTY
27 PART NO. 1093845 VENTURE PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.This warranty is extended only to the original purchaser/user of our products.This warranty gives you specific legal rights and you may also...