Page 2 - WARNING
HomeFill ™ 2 Part No 1148114 WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSIONAL, DEALER OR A QUALIFIED TECHNICIAN BEFORE ATTEMPTING TO USE THIS EQUIPMEN...
Page 3 - TABLE OF CONTENTS
TABLE OF CONTENTS Part No 1148114 3 HomeFill ™ TABLE OF CONTENTS SPECIAL NOTES ............................................... 4RECEIVING INSPECTION .................................. 4INTENDED USE ................................................ 4SECTION 1—GENERAL GUIDELINES .................. 5SEC...
Page 4 - SPECIAL NOTES; SIGNAL WORD
SPECIAL NOTES HomeFill ™ 4 Part No 1148114 SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words. RECEIVING INSPECTION Remove product from ...
Page 5 - SECTION 1—GENERAL GUIDELINES; Be sure all connections are tight and leak free.
SECTION 1—GENERAL GUIDELINES Part No 1148114 5 HomeFill ™ SECTION 1—GENERAL GUIDELINES WARNING SECTION 1 - IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product. DO NOT use this product or any available optional equipment without first completely reading ...
Page 6 - DO NOT attempt to repair product.
SECTION 1—GENERAL GUIDELINES HomeFill ™ 6 Part No 1148114 WARNING DO NOT attempt to repair product. DO NOT immerse product in any kind of liquid. Store product in a clean area when not in use. Only use an Invacare Corporation HF2POST6BAG or HF2POST9BAG carrying bag. NEVER smoke in an area where oxyg...
Page 7 - DO NOT store cylinders near sources of heat or flame.
SECTION 1—GENERAL GUIDELINES Part No 1148114 7 HomeFill ™ WARNING NEVER oil or lubricate the cylinder fittings. NEVER use any penetrating oil or lubricant such as WD-40 or 3 in One. If connection difficulty is experienced verify that the cylinder fitting is being inserted straight down into the comp...
Page 8 - SECTION 2—FEATURES; Product Diagram
SECTION 2—FEATURES HomeFill ™ 8 Part No 1148114 SECTION 2—FEATURES FIGURE 2.1 Product Diagram Cylinder Fill Port Toggle Valve
Page 9 - SECTION 3—TYPICAL PRODUCT PARAMETERS
SECTION 3—TYPICAL PRODUCT PARAMETERS Part No 1148114 9 HomeFill ™ SECTION 3—TYPICAL PRODUCT PARAMETERS DIMENSIONS: HF2POST6: Weight 4.0 lbs (1.81 kg) HF2POST6: Overall Length 11.1 in. (2.8 cm) HF2POST6: Diameter 4.4 in. (11.1 cm) HF2POST9: Weight 5.0 lbs (2.27 kg) HF2POST9: Overall Length 14 in (35....
Page 10 - SECTION 4—INSPECTION AND COMPRESSOR INTERFACE; NOTE: Refer to the HomeFill; Cylinder Prefill Inspection; All cylinders MUST be inspected before attempting to fill; The Hydrostatic Testing Date; Aluminum cylinders MUST undergo testing every five years.; DO NOT fill cylinders that have not been tested in the
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 10 Part No 1148114 SECTION 4—INSPECTION AND COMPRESSOR INTERFACE NOTE: Refer to the HomeFill ™ Compressor Owner’s Manual for compressor operating instructions. Cylinder Prefill Inspection WARNING All cylinders MUST be inspected before attempti...
Page 11 - External Examination
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1148114 11 HomeFill ™ External Examination NOTE: For this procedure, refer to FIGURE 4.1 on page 12. 1. Examine the outside of the cylinder for the following conditions and replace the cylinder if they exist: • Dents or dings • Arc burns • Oil or...
Page 12 - Checking Cylinder Pressure; NEVER use tools of any kind to connect/disconnect the; Install a conserver or regulator onto the cylinder according
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 12 Part No 1148114 FIGURE 4.1 External Examination Checking Cylinder Pressure WARNING NEVER use tools of any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur.DO NOT drop oxygen cylind...
Page 13 - less; Filling the Cylinder
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1148114 13 HomeFill ™ 2. Ensure toggle valve is open. Open the toggle valve by rotating counterclockwise. 3. Examine the gauge on the connected conserver or regulator. Perform one of the following: • If the cylinder pressure is less than 1500 psi...
Page 14 - Component Identification
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 14 Part No 1148114 FIGURE 4.2 Component Identification 4. Remove the post valve cover (located on the cylinder) and the connector fillport covers (located on the compressor) (FIGURE 4.3). FIGURE 4.3 Fillport Covers 5. On the compressor unit, m...
Page 15 - Resetting Connector Fillport
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1148114 15 HomeFill ™ NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing DOWN momentarily will reset the connector fillport (GREEN dots...
Page 16 - Connecting the Cylinder to the Compressor
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 16 Part No 1148114 NOTE: Refer to Cylinder Fill Times on page 20 for the length of time it will take to fill the cylinder. FIGURE 4.5 Connecting the Cylinder to the Compressor Cylinder Compressor Cradle Outer Ring (Sleeve) Connector Fillport P...
Page 17 - After the cylinder is full, press the compressor power
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1148114 17 HomeFill ™ 11. After the cylinder is full, press the compressor power switch to the Off (O) position (FIGURE 4.6). 12. Grasp the cylinder assembly (FIGURE 4.6). 13. With the other hand, grasp the outer ring (sleeve) of the connector fi...
Page 18 - The fillport cover on the connector fillport and the post
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 18 Part No 1148114 WARNING The fillport cover on the connector fillport and the post valve fillport MUST be replaced after filling and whenever not in use. If either of the fillport covers are missing contact your healthcare provider before us...
Page 19 - SECTION 5—MAINTENANCE; After each use, clean exterior of the product with a dry, lint; CAUTION
SECTION 5—MAINTENANCE Part No 1148114 19 HomeFill ™ SECTION 5—MAINTENANCE 1. After each use, clean exterior of the product with a dry, lint free cloth only. 2. Store product in a clean area free from grease, oil, and other sources of contamination. CAUTION DO NOT use cleaning solutions. DO NOT immer...
Page 20 - SECTION 6—CYLINDER FILL TIMES; HF2POST6BAG or HF2POST9BAG Carrying Bag.; Cylinder; PROBLEM
SECTION 6—CYLINDER FILL TIMES HomeFill ™ 20 Part No 1148114 SECTION 6—CYLINDER FILL TIMES NOTE: All filling times are approximate and may vary on environmental conditions. * Troubleshooting If post valve cylinder fails to function, try the following solutions. If problem cannot be corrected, consult...
Page 21 - NOTES
Page 22 - LIMITED WARRANTY
LIMITED WARRANTY HomeFill ™ 22 Part No 1148114 LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.This warranty is extended only to the original purchaser/user of our products.This warranty gives you spe...