Page 2 - RECOMMENDATIONS FOR USERS
2 WARRANTY Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instruments. This manual will provide you with the necessary information for correct use of the instruments, as well as a precise idea of their versatility. If...
Page 6 - O P E R A T I O N A L G U I D E; pH ELECTRODES
6 INITIAL PREPARATION The meter is supplied with rechargeable batteries (located inside the meter - see page 14 for details).To prepare the instrument for use, connect the pH electrode and the temperature probe to the BNC and temperature sockets on the top of the instrument. The temperature probe ca...
Page 7 - pH CALIBRATION SOLUTIONS
7 The LCD will show the pH measurement and the temperature of the sample. In order to take more accurate pH measurements, make sure that the instrument is calibrated (see page 9 for details).The glass bulb and the junction on the pH electrode should always be moist, never allow it to dry out.If seve...
Page 8 - T R O U B L E S H O O T I N G G U I D E
8 Submerge the ORP electrode 4cm (1½") into the sample to be tested and wait a few minutes for the reading to stabilize. Measurements within the ±699.9 mV range are displayed with 0.1mV resolution, while outside this range the resolution automaticallyswitches to 1 mV.For accurate ORP measurement...
Page 9 - For refillable electrodes:
9 It is recommended to calibrate the instrument frequently, especially ifhigh accuracy is required.The pH range should be recalibrated:• Whenever the pH electrode or temperature probe is replaced.• At least once a week.• After testing aggressive chemicals.• When extreme accuracy is required. PREPARA...
Page 10 - PREPARATION PROCEDURE
10 • The LCD will flash the "WAIT NOT READY" message. • Once the reading is stable, if it is not close to the selected buffer, "WRONG " and "WRONG " will blink alternatively; if it is close to the selected buffer the display will change to "READY" andblinking "CFM...
Page 11 - ONE-POINT CALIBRATION
11 • If the reading is not close to the selected buffer, "WRONG " and "WRONG " will blink alternatively; • If the reading is close to the selected buffer and the reading is stable, the “READY” symbol is displayed and the “CFM” symbolstarts blinking on the LCD, asking for confirmation...
Page 13 - NiMH
13 The instrument allows the user to configure several parametersthrough the Setup Menu.To enter the Menu mode, while in measurement mode, press andhold the SETUP key for about 5 seconds.Once the menu is entered, each parameter can be changed by usingthe arrow keys; then pressing the CFM key will co...