Page 2 - P R E L I M I N A R Y E X A M I N A T I O N

3 2 HI 9033 and HI 9034 multi-range conductivity and TDS meters, have been designed for extended use in wet, humid, dusty and muddy conditions. These meters utilize the latest in four-ring potentiometric technologywhich has been proven to provide higher accuracy than the commonamperometric method. H...

Page 4 - O P E R A T I O N A L G U I D E

7 6 INITIAL PREPARATION Each meter is supplied complete with a 9 V battery. Remove the backcover, unwrap the battery and install it while paying attention to itspolarity its (see "Battery Replacement" section). Connect the probe to the meter securely byaligning the pins with the socket, push...

Page 5 - calibration solutions see; C A L I B R A T I O N; TEMPERATURE EFFECT; conductivity probe features a built-in temperature sensor; AFTER MEASUREMENTS

9 8 For best results, choose a calibration solution with a conductivity orTDS value as close as possible to the sample to be measured.For example, for measurements in the 2 to 20 mS/cm range, use the HI 7030 (12.88 mS/cm) conductivity calibration solution. For accurate calibration, use two beakers f...