TABLE OF CONTENTS Page I © 2007 DH Instruments, a Fluke Company T T A A B B L L E E O O F F C C O O N N T T E E N N T T S S T A B L E O F C O N T E N T S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I T A B L E S . . . ...
ABOUT THIS MANUAL Page VII © 2007 DH Instruments, a Fluke Company A A B B O O U U T T T T H H I I S S M M A A N N U U A A L L This manual is intended to provide the user with the basic information necessary to operate an RPM4/HPMS A70M/A20M reference pressure monitor with high pressure mounting syst...
1. INTRODUCTION Page 1 © 2007 DH Instruments, a Fluke Company 1 1 . . I I N N T T R R O O D D U U C C T T I I O O N N 1 . 1 P R O D U C T O V E RV I E W RPM4/HPMS A70M/A20M-AF is the combination of an RPM4 A70M/A20M-AF reference pressure monitor and an HPMS High Pressure Mounting System. RPM4 A70M/A...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 2 1 . 2 S P E C I F I C AT I O N S 1 . 2 . 1 G E N E R A L S P E C I F I C A T I O N S Power Requirements: 85 to 264 VAC, 50/60 Hz, 25 VA max consumption and 12VDC, 1.2 A Operating Temperature Range: ...
1. INTRODUCTION Page 3 © 2007 DH Instruments, a Fluke Company 1 . 2 . 2 P R E S S U R E M E A S U R E M E N T S P E C I F I C A T I O N S 1.2.2.1 QUARTZ REFERENCE PRESSURE TRANSDUCERS (Q-RPT) RPM4 A70M/A20M is configured with two quartz reference pressure transducers (Q-RPT) modules to measure press...
2. INSTALLATION Page 5 © 2007 DH Instruments, a Fluke Company 2 2 . . I I N N S S T T A A L L L L A A T T I I O O N N 2 . 1 U N PA C K I N G A N D I N S P E C T I O N RPM4/HPMS A70M/A20M-AF is usually delivered as part of an PGC-10000-AF system which includes a GPC1-10000-AF gas pressure controller ...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 6 2 . 1 . 2 I N S P E C T I N G C O N T E N T S Check that all items are present and have no visible damage. If damage is noted, report it to your Receiving Department for appropriate action. A new RP...
2. INSTALLATION Page 7 © 2007 DH Instruments, a Fluke Company Each of the three components of the PGC-10000-AF system has its own Operation and Maintenance Manual and individual setup and start up instructions. Figure 1 shows the typical setup configuration of a complete PGC-10000-AF system. Figure ...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 8 2 . 4 S E T U P 2 . 4 . 1 P R E P A R I N G F O R O P E R A T I O N To prepare RPM4 for check out and operation: Remove the plastic plug from the RPM4/HPMS rear panel TEST connection. Remove the pro...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 10 2.4.2.3 RPM4/HPMS SIDE VIEW 1. Lo Q-RPT shut off valve 2. RPM4/HPMS A70M/A20M Reference Pressure Monitor 3. Pressure relief valve (3 300 psi) 4. Foot switch (remote [ENT]) cable connector 5. TEST p...
2. INSTALLATION Page 11 © 2007 DH Instruments, a Fluke Company 2.4.3.2 RPM4 REAR PANEL 1. 12VDC Power Supply Connection 2. Accessories connector (to HPMS rear panel FOOTSWITCH and LEDs) 3. IEEE-488 Connector 4. COM2 Connector 5. COM1 Connector 6. Power Switch 7. Fuse 8. Electrical Power Connector (I...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 12 USE THE CORRECT PRESSURE CONNECTORS: The RPM4/HPMS TEST port fitting is a DH500 F (see Section 1.2.1). It is NOT a 1/8 in. NPT F. Never use a fitting other than the corresponding male fitting in th...
2. INSTALLATION Page 13 © 2007 DH Instruments, a Fluke Company 2 . 5 P O W E R - U P A N D V E R I F I C AT I O N 2 . 5 . 1 S W I T C H P O W E R O N Actuate the power switch on the RPM4 rear panel. Observe the front panel display as RPM4 initializes, error checks and goes to the MAIN RUN screen (se...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 14 It is normal for RPM4 to indicate a value other than zero when vented when gauge mode is first entered or ranges are changed, especially if AutoZ is OFF, RPM4 has been OFF for some time or its loca...
3. OPERATION Page 15 © 2007 DH Instruments, a Fluke Company 3 3 . . O O P P E E R R A A T T I I O O N N 3 . 1 U S E R I N T E R FA C E RPM4/HPMS is designed to offer the optimum balance between simple, straight forward operation and the availability of a wide variety of functions with a high level o...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 16 P R E S S U R E 1 U N I T M h z R R D D I S P L A Y F U C N I O N n n / n n DISPLAY FIELD NAME PURPOSE CONTENTS SECTION 1. PRESSURE1 Measured pressure Displays pressure measured by the active Q-RPT...
3. OPERATION Page 17 © 2007 DH Instruments, a Fluke Company 3 . 1 . 2 F U N C T I O N / D A T A K E Y P A D L A Y O U T A N D P R O T O C O L The RPM4 has a function/data keypad for local operator access to direct functions, function menus and for data entry. 1. The Function/Data keys allow very com...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 18 3 . 1 . 3 R E M O T E [ E N T ] ( E N T E R ) F O O T S W I T C H Operating the footswitch is the equivalent of pressing [ENT] on the front panel. The remote ENTER feature can be particularly conve...
3. OPERATION Page 19 © 2007 DH Instruments, a Fluke Company 3 . 2 . 2 G A U G E M O D E , D Y N A M I C C O M P E N S A T I O N F O R A T M O S P H E R I C P R E S S U R E The RPM4 A70M/A20M-AF Q-RPTs are intrinsically absolute but they are also used for gauge measurement mode (see Section 3.3.3, ...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 20 The ranges available on the RPM4 are accessed using [RANGE] (see Section 3.3.1) and/or created using [AutoRange] (see Section 3.3.4). Most settings made in an RPM4 range, such as unit of measure, m...
3. OPERATION Page 21 © 2007 DH Instruments, a Fluke Company The Lo Q-RPT isolation valve (3) isolates and protects the Lo Q-RPT from the TEST port when it is closed (knob fully CCW) and opens the Lo Q-RPT to the TEST port when it is open (knob fully CW). The Valve Position LEDs (2, 4) indicate the p...
3. OPERATION Page 23 © 2007 DH Instruments, a Fluke Company 1. HPMS TEST port (DH500 F) 2. Connection to RPM4 Lo Q-RPT (A20M) (3 000 psi) 3. Lo Q-RPT pressure relief valve (3 300 psi) 4. Lo Q-RPT shut off valve 5. RPM4 A70M/A20M-AF Reference Pressure Monitor 6. Connection to RPM4 Hi Q-RPT (A70M) (10...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 24 3 . 2 . 6 D I R E C T F U N C T I O N K E Y S S U M M A R Y Table 3 provides a brief summary of RPM4 direct function key operation. It may be useful to keep a copy of this summary near the RPM4/HPM...
3. OPERATION Page 25 © 2007 DH Instruments, a Fluke Company Ranges created using AutoRange or AutoTest may be saved, with their settings, for reactivation (see Section 3.4.1). The [RANGE] function key allows the available RPM4 pressure ranges to be viewed and selected. OPERATION Pressing the [RANG...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 26 3 . 3 . 2 [ U N I T ] PURPOSE To select the unit of measure in which RPM4 displays pressure values. To select the measurement mode (absolute, gauge), use [MODE] (see Section 3.3.3). PRINCIPLE R...
3. OPERATION Page 27 © 2007 DH Instruments, a Fluke Company 3 . 3 . 3 [ M O D E ] PURPOSE To set the measurement mode (absolute, gauge) for the active range. To select the unit of measure, use [UNIT] (see Section 3.3.2). PRINCIPLE RPM4 A70M/A20M-AF supports simple, one-step switching between two...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 28 PRINCIPLE RPM4 A70M/A20M-AF is designed to support the calibration and test of a very wide variety of test ranges in absolute and gauge measurement modes. The AutoRange function simplifies the ta...
3. OPERATION Page 29 © 2007 DH Instruments, a Fluke Company OPERATION To use the AutoRange function, press [AutoRange] and respond to the RPM4’s prompts. Select AutoRange measurement mode: The screen is identical in appearance and function to the [MODE] screen (see Section 3.3.3). Select pressure ...
3. OPERATION Page 31 © 2007 DH Instruments, a Fluke Company Leak check is range specific in the sense that a leak check is run using the active range. However, only one set of leak check results is maintained in memory and each leak test completed overwrites the memory. View leak check always shows ...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 32 OPERATION To set the DISPLAY function press [DISPLAY] from the main run screen. The display is: 1avg 2rate 3dev 4RPT 5HiLo 6freeze 7clean The cursor is on the active DISPLAY function. Selecting a...
3. OPERATION Page 33 © 2007 DH Instruments, a Fluke Company The Average DISPLAY has a second screen that allows the instantaneous pressure readings to be viewed while an averaging cycle is running. The instantaneous Average screen is: 1. Instantaneous pressure values at RPM4’s normal update rate. 2....
3. OPERATION Page 35 © 2007 DH Instruments, a Fluke Company With the Deviation DISPLAY active the main run screen is: 1. Deviation of current pressure from the target value. 2. Target value. 8998.2 psi a zH3 D -1.8 T9000.0 Pressing [ENT] from the main run screen when the Deviation DISPLAY is active ...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 36 With the RPT DISPLAY active the main run screen is: 1. Active Q-RPT display. 2. Inactive Q-RPT display with Ready/Not Ready indicator on far left ( <*> indicates Ready ). 2010.66 psi a zHi *1...
3. OPERATION Page 37 © 2007 DH Instruments, a Fluke Company 3.3.6.6 Freeze PURPOSE To activate the Freeze DISPLAY. See Section 3.3.6, Principle. OPERATION To activate the Freeze DISPLAY press [DISPLAY] , <6Freeze> . Pressing <6Freeze> returns to the main run screen with the Freeze ...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 38 3 . 3 . 7 [ H E A D ] PURPOSE To cause a pressure fluid head correction to be added to or subtracted from the pressure measured by the RPM4 reference pressure transducer in order to predict the p...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 40 Before running AutoZ in gauge mode, ensure that the pressure applied to the Q-RPT is truly zero gauge (atmospheric pressure). If running AutoZ in gauge mode results in a zero offset that RPM4 consi...
3. OPERATION Page 41 © 2007 DH Instruments, a Fluke Company If running AutoZ results in a value of P offset that is greater than ± 0.005 % FS of the span of the Q-RPT that is being AutoZeroed, the Q-RPT and/or the reference used as the source of P std,0 may be out of tolerance or the AutoZ process m...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 42 Run AutoZ by COM2 AutoZ by COM2 allows the value of P std,0 (see Section 3.5.1 Principle) to be read automatically from a DHI RPMx connected to the RPM4 COM2 communications port. The RPMx is read...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 44 • Stability test for Ready/Not Ready indication(see Section 3.2.1): The DUT tolerance divided by 10, or 2 ppm of Q-RPT span, whichever is greater. • Upper limit (UL): The DUT full scale plus 5 %, o...
3. OPERATION Page 45 © 2007 DH Instruments, a Fluke Company • <DUT tolerance:>: Enter the DUT measurement uncertainty, accuracy or precision specification as desired in ± % FS (percent of full scale). This value will be used to determine appropriate RPM4 resolution and stability test settings....
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 46 The display is: 0.0 psi g Lo Exercise DUT & [ENT] The top line is the standard RPM4 run screen indicating the current measured pressure (see Section 3.1.1). Exercise the DUT if desired using th...
3. OPERATION Page 47 © 2007 DH Instruments, a Fluke Company 3 . 4 [ S E T U P ] PURPOSE [SETUP] accesses a menu of functions and features commonly used in setting up and using RPM4. OPERATION To access the SETUP menu, press [SETUP] from the MAIN RUN screen. The display is: 1range 2res 3stab 4UL ...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 48 Press [ENT] to save the range or [ESC] to return to the MAIN RUN screen without saving the range. Default (<DF>) Q-RPT ranges cannot be saved. They are permanent. Duplicate AutoRange ranges c...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 50 3 . 4 . 4 < 4 U L > ( U P P E R L I M I T ) PURPOSE To set the upper pressure limit value for a pressure range and measurement mode. PRINCIPLE The UPPER LIMIT function provides the user w...
3. OPERATION Page 51 © 2007 DH Instruments, a Fluke Company The over pressure function causes the measured pressure display to flash. The overpressure function also logs the time and date of the overpressure condition in both user and factory logs to assist in incident diagnosis (see Section 3.5.7.5...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 52 1. Full scale of the AutoTest with pressure unit of measure and measurement mode. 2. Number of test points executed/total number of points in the AutoTest. 3. Date and time that the last point in t...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 56 being AutoZeroed. Given the high pressure range of the RPM4 Q-RPTs, adequate uncertainty is easily available from a variety of digital barometers or, for the Hi Q-RPT, by the Lo Q-RPT. In absolute ...
3. OPERATION Page 57 © 2007 DH Instruments, a Fluke Company Recommendations for the Use of the AutoZ Function The AutoZ function provides a powerful and easy to use tool for improving the stability over time of RPM4 Q-RPTs and maximizing the recalibration interval by compensating for zero drift betw...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 58 3.5.1.1 EDIT AUTOZ The edit AutoZ function should be used only with great care as entering inappropriate values may result in out of tolerance measurements. In normal operation, the value of the Au...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 60 The display prompts you to connect COM1 to COM2 using a standard pin-to-pin DB-9F to DB-9M RS232 cable (see Section 4.2.1.1, 4.2.1.3). Once the cable has been installed, press [ENT] to run the self...
3. OPERATION Page 61 © 2007 DH Instruments, a Fluke Company To access the PREFS menu press [SPECIAL] , <5prefs> . The display is: 1ScrSvr 2sound 3time 4ID 5level The PREFS menu includes: <1ScrSvr> View and change the screen saver activation time (see Section 3.5.5.1). <2sound> View...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 62 The sound function only affects the valid key press tone. The invalid key press tone and other RPM4 sounds cannot be adjusted or suppressed. 3.5.5.3 <3TIME> PURPOSE To view and edit the RPM...
3. OPERATION Page 63 © 2007 DH Instruments, a Fluke Company 3.5.5.5 <5LEVEL> (SECURITY) PURPOSE To set user protection levels to restrict access to certain functions and to edit the password required for changing user levels. PRINCIPLE RPM4’s front panel user interface provides the means t...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 64 The security levels are structured to support typical levels of operation as shown in Table 6. Specifically, the security levels prevent execution of the functions accessed by the key strokes marke...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 66 customizes the [UNIT] function key to support his/her most commonly used units. Up to six units can be included in the [UNIT] screen. OPERATION To customize the [UNIT] function key, press [SPECIA...
3. OPERATION Page 67 © 2007 DH Instruments, a Fluke Company The INTERNAL menu choices include: <1baro> View the real time output of the RPM4 on-board barometer (see Section 3.5.7.1). <2ReadRt> Turn ON and OFF RPM4’s automated, rate of change dependent, reading and integration and display...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 68 When the automated read rate function is ON, three pressure rate of change dependent read rates are used. The result is three display update rates: Table 8. READRT - display update rates PRESSURE R...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 70 Table 9. Reset – Sets RESET RESULT SEE SECTION [UNIT] Pressure unit of measure to first of the six available (psi) 3.3.2 [MODE] Measurement mode to the native mode of the Q-RPT (absolute) 3.3.3 [RA...
3. OPERATION Page 71 © 2007 DH Instruments, a Fluke Company 3.5.9.4 <4 CAL> PURPOSE The Reset - Cal function will reset the Q-RPT and barometer calibration coefficients and reset AutoZ values to zero. This will change the RPM4 calibration and could cause it to make out of tolerance measureme...
4. REMOTE OPERATION Page 73 © 2007 DH Instruments, a Fluke Company 4 4 . . R R E E M M O O T T E E O O P P E E R R A A T T I I O O N N 4 . 1 O V E RV I E W Most of the RPM4 front panel functions can also be executed by commands from a remote computer. The host computer can communicate to the RPM4 us...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 74 4.2.1.2 IEEE-488 The RPM4 IEEE-488 interface is located on the back of the unit. The physical and electrical interface conforms to IEEE Std 488.1-1987 Subset E2 and IEEE Std. 488.2-1992. You should...
4. REMOTE OPERATION Page 75 © 2007 DH Instruments, a Fluke Company 4 . 3 . 1 C L A S S I C P R O G R A M M E S S A G E F O R M A T Each program message sent is also a query. You can only send one program message to the RPM4 at time. After sending any program message, you must wait for the RPM4 to re...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 76 Examples: • IEEE-488 enhanced mode command series using query operator: “*CLS?” (Clear the error queue. Wait for reply) “UNIT? KPA” (Generates a reply. User must wait for reply before continuing) “...
4. REMOTE OPERATION Page 77 © 2007 DH Instruments, a Fluke Company 4 . 4 C O M M A N D S 4 . 4 . 1 P R O G R A M M I N G M E S S A G E S Some commands accept an optional suffix. This suffix can be used to specify the explicit Q-RPT to address. In most cases, each Q-RPT in the RPM4 operates as an ind...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 78 4 . 4 . 2 E R R O R M E S S A G E S S Table 15. Error #s and descriptions REPLY DESCRIPTION ERR# 0 “OK” ERR# 2 “Text argument is too long” ERR# 3 “Arguments cannot be 0” ERR# 4 “External device not...
4. REMOTE OPERATION Page 79 © 2007 DH Instruments, a Fluke Company 4 . 4 . 3 P R O G R A M M E S S A G E D E S C R I P T I O N O V E R V I E W Each program message description is separated into the following sections: Purpose A brief description of the programs message’s function. Command This is th...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 80 4 . 4 . 4 P R O G R A M M E S S A G E D E S C R I P T I O N S # Purpose To allow the host PC to communicate with a device connected to the RPM4 COM2 port. Classic “#xx” Arguments xx: The string to ...
4. REMOTE OPERATION Page 97 © 2007 DH Instruments, a Fluke Company ZOFFSETn Purpose Read or set the AutoZ pressure offset (P offset ) for the specified or active Q-RPT in the current measurement mode. Command Query “ZOFFSET n GaOffset , AbsOffset , DifOffset ” “ZOFFSET n ?” Classic “ZOFFSET n = GaOf...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 98 4 . 5 . 2 S T A T U S B Y T E R E G I S T E R The RPM4 contains an 8 bit Status Byte Register that reflects the general status of the RPM4. Table 16. 8 Bit status byte register OPER (128) RQS/MSS (...
4. REMOTE OPERATION Page 99 © 2007 DH Instruments, a Fluke Company MAV Message Available Bit 4 (16) Indicates that at least one reply message is waiting in the RPM4 IEEE-488 output queue. ERROR Error Queue Not Empty Bit 2 (4) Indicates that at least one command error message is waiting in the RPM4 I...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 100 4 . 5 . 4 R E A D Y S T A T U S R E G I S T E R The RPM4 contains an 8 bit Ready Status Register that reflects specific RPM4 Q-RPT measurement ready events. Enabled events in this register will se...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 105 © 2007 DH Instruments, a Fluke Company 5 5 . . M M A A I I N N T T E E N N A A N N C C E E , , A A D D J J U U S S T T M M E E N N T T S S A A N N D D C C A A L L I I B B R R A A T T I I O O N N 5 . 1 O V E RV I E W RPM4 A70M/A20M-AF was designed...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 106 If the RPM4 is used only in gauge mode, Autozeroing in absolute mode is not necessary and vice-versa. Gauge and absolute mode Autozeroing are completely independent one from the other. See Section...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 107 © 2007 DH Instruments, a Fluke Company Never apply a pressure greater than 16 psi (110 kPa) to the barometer port. Overpressure and possible damage may result. 5 . 3 R P M 4 Q - R P T C A L I B R AT I O N 5 . 3 . 1 P R I N C I P L E RPM4 A70M/A20...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 108 As editing PA and PM values changes Q-RPT calibration, they should only be edited by qualified personnel as part of the calibration process. Caution should be taken to avoid accidental editing and...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 109 © 2007 DH Instruments, a Fluke Company pressures. Absolute pressures are be defined by adding atmospheric pressure measured by a barometer to gauge pressure defined by the piston gauge. If, after calibration, the RPM4 will not be used in absolute...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 110 5 . 3 . 4 R E C O M M E N D E D C A L I B R A T I O N P O I N T S E Q U E N C E Calibration adjustments to RPM4 Q-RPTs are made by altering calibration coefficients, PA and PM (see Section 5.3.1.1...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 111 © 2007 DH Instruments, a Fluke Company OPERATION See Section 5.3.7 for more detailed information on editing and viewing Q-RPT calibration information. Turning absolute mode operation ON and OFF for a Q-RPT occurs in the same area in which calib...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 114 5 . 4 R P M 4 R E PA I R 5 . 4 . 1 R E M O V I N G R P M 4 F R O M H P M S The RPM4/HPMS is an integrated assembly. The RPM4 should not be removed from the HPMS unless it is necessary for RPM4 and...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 115 © 2007 DH Instruments, a Fluke Company 5 . 4 . 2 O P E N I N G A N D C L O S I N G T H E R P M 4 E N C L O S U R E To open the RPM4 case, first remove the RPM4 from the HPMS (see Section 5.4.1). Once the RPM4 is removed from the HPMS, proceed as ...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 116 web or have difficulty downloading or loading software, contact your DHI representative or a DHI Authorized Service Provider (see Table 23) for assistance. If you believe you have discovered an er...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 117 © 2007 DH Instruments, a Fluke Company 5.4.4.2 POWER SUPPLY MODULE + 12 V DC (± 2 %) @ 2.1 Amps 5.4.4.3 DRIVER BOARD The driver board is controlled by the mini micro board (see Section 5.4.4.1). It supports: • Frequency counters (2) for reading Q...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 118 5.4.4.7 COOLING FAN The cooling fan is mounted on the RPM4 rear panel and connected to the main board. It is a 12V brushless DC fan. 5 . 4 . 5 Q - R P T M O D U L E P N E U M A T I C S C H E M A T...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 119 © 2007 DH Instruments, a Fluke Company Table 22. RPM4 Rear Panel Illustrated Parts Breakdown (see Figure 16) # REQUIRED DESCRIPTION PART NO. 1 Rear panel 124154 8 Screw .30C100MXFS 1 Fan D2510M12B 1 Fan guard JGW25 2 Screw .25C160MXPS 1 Power ent...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 120 Table 23. RPM4 Illustrated Parts Breakdown of Display Assembly (see Figure 17) # REQUIRED DESCRIPTION PART NO. 1 Display assembly 103122 1 Ribbon cable 401963 1 Panel, front 123566 2 Standoff, M3 ...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 121 © 2007 DH Instruments, a Fluke Company Table 24. RPM4 Illustrated Parts Breakdown of Internal Assembly (see Figure 18) # REQUIRED DESCRIPTION PART NO. 1 Bracket, barometer 123549 1 Barometer assembly 401898-06 1 Cable assembly, barometer 401966 8...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 123 © 2007 DH Instruments, a Fluke Company Table 26. RPM4 Illustrated Parts Breakdown of Electronic Chassis (see Figure 20) # REQUIRED DESCRIPTION PART NO. 1 Micro card 401365 4 Standoff SS-5172-200-00 10 Screw .30C80KCSS 11 Lock washer .30NLOCS 1 Dr...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 124 5 . 6 H P M S M A I N T E N A N C E There are no HPMS maintenance procedures other than the overhaul procedure. 5 . 6 . 1 H P M S O V E R H A U L RPM4/HPMS A70M/A20M-AF is generally overhauled as ...
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS Page 125 © 2007 DH Instruments, a Fluke Company 5 . 7 H P M S I L L U S T R AT E D PA R T S B R E A K D O W N Table 27. HPMS Assembly Illustrated Parts Breakdown (see Figure 21 and 22) # REQUIRED DESCRIPTION PART NO. 1 HPMS bracket 124142 1 RPM4 mounting ...
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL © 2007 DH Instruments, a Fluke Company Page 126 Figure 21. HPMS Schematic
6. TROUBLESHOOTING Page 129 © 2007 DH Instruments, a Fluke Company 6 6 . . T T R R O O U U B B L L E E S S H H O O O O T T I I N N G G RPM4 is a sophisticated pressure setting and measuring instrument with advanced on-board features and functions. Before assuming that unexpected behavior is caused b...
7. APPENDIX Page 133 © 2007 DH Instruments, a Fluke Company 7 7 . . A A P P P P E E N N D D I I X X 7 . 1 U N I T C O N V E R S I O N 7 . 1 . 1 P R E S S U R E RPM4 performs all internal calculations in SI units. Numerical values input or output in other units are converted to SI immediately after e...
8. WARRANTY Page 135 © 2007 DH Instruments, a Fluke Company 8 8 . . W W A A R R R R A A N N T T Y Y 8 . 1 O V E RV I E W Except to the extent limited or otherwise provided herein, DH Instruments, a Fluke Company warrants for one year from purchase, each new product sold by it or one of its authorize...
Fluke Manuals
-
Fluke 54IIB
Manual
-
Fluke 1750
Manual
-
Fluke 1502A
Manual
-
Fluke 187
Manual
-
Fluke 99B
Manual
-
Fluke Thermometer 576
Manual
-
Fluke PM8907
Manual
-
Fluke 2686A
Manual
-
Fluke 574
Manual
-
Fluke 23
Manual
-
Fluke 196C
Manual
-
Fluke 80T-150U
Manual
-
Fluke 80K-40
Manual
-
Fluke 335D
Manual
-
Fluke 68X Series
Manual
-
Fluke PM3370B
Manual
-
Fluke 199
Manual
-
Fluke 8010A
Manual
-
Fluke 6062a
Manual
-
Fluke 5626/5628
Manual