Page 2 - Manual Supplement; Replace pages 31 through 35, with the following:; Defibrillator Analyzer Specifications; Note
Impulse 6000D/7000DP Users Manual Supplement 4/08 1 Change #1, 46470, 46697, 47714, 48007, 49680 Replace pages 31 through 35, with the following: Defibrillator Analyzer Specifications Energy Output Measurement Compatible Defibrillator Waveshapes ................ Lown, Edmark, Trapezoidal, DC Bi-phas...
Page 3 - Voltage; Interphase delay (biphasic and pulsed biphasic); Sample rate
Manual Supplement Impulse 6000D/7000DP Users 2 8/08 Accuracy ............................................................. ± 0.1 ms Voltage Range ................................................................. 20 to 5000 V Accuracy ............................................................. ± (1...
Page 4 - Scope Output; Charge Time Measurement; Automated Defibrillator Test ECG Waves
Impulse 6000D/7000DP Users Manual Supplement 4/08 3 Scope Output Autorange ........................................................... 2000:1, 400:1 and 80:1: dependant on the range Waveform Playback Output ............................................................. BNC Output impedance..............
Page 5 - ECG General
Manual Supplement Impulse 6000D/7000DP Users 4 3/09 Ventricular Fibrillation ......................................... Coarse and fine Monomorphic Ventricular Tachycardia ............... 120 to 300 (by 5) BPM Polymorphic Ventricular Tachycardia ................. 5 types Asystole .......................
Page 6 - ECG Normal Sinus
Impulse 6000D/7000DP Users Manual Supplement 8/08 5 Normal Sinus waves Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 70 % 100 % 30 % 24 % 48 % 100 % 120 % 112 % 80 % Lead I reference Performance waves and R wave detection Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 100 % 150 % 50 % 100 % 100 % 100 % 1...
Page 7 - Noise Immunity
Manual Supplement Impulse 6000D/7000DP Users 6 4/08 Sine waves ......................................................... 0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz Pulse................................................................... 30 and 60 BPM, 60 ms pulse width R-Wave Detection Wavef...
Page 8 - Arrhythmia Selections
Impulse 6000D/7000DP Users Manual Supplement 4/08 7 Accuracy ......................................................... ±(10 % of setting + 0.2 mV) Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent) Lead II reference Lead # I II III V1 V2 V3 V4 V5 V6 Re...
Page 9 - On page 12, prior to; SCOPE OUTPUT; Table 3A. Synchronizing Marker Pulses; Defib fire
Manual Supplement Impulse 6000D/7000DP Users 8 4/08 Change #2 On page 12, prior to Analyzing Pacemakers (7000DP only) , add the following section: SCOPE OUTPUT Each time a defibrillator is fired into the Analyzer during an active defibrillator test, a delayed representation of the defibrillator puls...
Page 10 - On page 26, prior to; Setting the ECG Reference Lead; More
Impulse 6000D/7000DP Users Manual Supplement 4/08 9 Pulse playback +4 V, 1 ms pulse Start of defibrillator pulse playback [1] [1] Pulse playback includes 0.4 ms before the pulse was detected and therefore displays any early transitional signals. Use the synchronizing marker pulses to display specifi...