Page 2 - Limited Warranty & Limitation of Liability
Limited Warranty & Limitation of Liability Each product from Fluke's Hart Scientific Division ("Hart") is warranted to be free from defects in mate- rial and workmanship under normal use and service. The warranty period is one year for the Surface Cal- ibrator. The warranty period begins...
Page 3 - Table of Contents
Table of Contents 1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 Warnings . . . . . . . . . . . . . . . . . . . . . ...
Page 5 - Symbols Used; Before You Start; Table 1
1 Before You Start 1.1 Symbols Used Table 1 lists the International Electrical Symbols. Some or all of these symbolsmay be used on the instrument or in this manual. Symbol Description AC (Alternating Current) AC-DC Battery CE Complies with European Union Directives DC Double Insulated Electric Shock...
Page 6 - Safety Information; Warnings
Symbol Description Canadian Standards Association OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re-fers to the level of Impulse Withstand Voltage protection provided. Equipment ofOVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied fromthe fixed installatio...
Page 8 - Cautions; DO use a ground fault interrupt device.; Authorized Service Centers; 99 E. Utah Valley Drive
1.2.2 Cautions • DO NOT plug the unit into 230V if the heater switches and fuse holderread 115V. This action will cause the fuses to blow and may damage theinstrument. • Components and heater lifetime can be shortened by continuous hightemperature operation. • DO NOT change the values of the calibra...
Page 11 - Introduction; Figure 1
2 Introduction Hart Scientific’s Model 3125 Detachable Hot Plate is controlled by Hart Scien-tific’s 2200 controller and uses a precision platinum RTD as a sensor with aheater to control the temperature. The 3125 and 2200 are a matched set. The controller display shows the temperature and also the s...
Page 13 - Specifications; Environmental Conditions
3 Specifications and EnvironmentalConditions 3.1 Specifications Temperature Range 35°C to 400°C (95°F to 752°F) Display Accuracy ±0.5°C to 200°C±1.0°C to 400°C Stability ±0.2°C to 300°C±0.3°C to 400°C Resolution 0.01° Uniformity ±0.3°C at 100°C±0.6°C at 200°C±0.9°C at 300°C±1.4°C at 400°C Heating Ti...
Page 15 - Troubleshooting
4 Troubleshooting 4.1 Troubleshooting Problems, Possible Causes,and Solutions In the event that the instrument appears to function abnormally, this sectionmay help to find and solve the problem. Several possible problem conditionsare described along with likely causes and solutions. If a problem ari...
Page 16 - Comments; EMC Directive
Problem Possible Causes and Solutions The display shows anyof the following: err 1 , err 2 , err 3 , err 4 , Err 5 , Err 6 , or Err 7 Controller problem. The error messages signify the following problems with the controller. Err 1 - a RAM error Err 2 - a NVRAM error Err 3 - a Structure error Err 4 -...