Page 3 - DANGER: HAZARDOUS AREA INSTALLATION; CAUTION: SENSOR/PROCESS APPLICATION COMPATIBILITY; Read this page before proceeding
DANGER: HAZARDOUS AREA INSTALLATION Emerson designs, manufactures, and tests its Rosemount Analytical products to meet many national and international standards. Because these instruments are sophisticated technical prod- ucts, you must properly install, use, and maintain them to ensure they continu...
Page 4 - About this document
II About This Document This manual contains instructions for installation and operation of the Model 1066 Smart Transmitter. The following list provides notes concerning all revisions of this document. Rev. Level Date Notes A 2/01 This is the initial release of the product manual. The manual has bee...
Page 5 - Contents
Table of Contents III 389VP Instruction Manual Table of Contents PN 51-389VP January 2013 Contents Section 1: Description and Specifications 1.1 Features and Applications ............................................................................................1 1.2 Performance and Physical Specif...
Page 6 - List of Tables; List of Figures
IV Table of Contents Table of Contents 389VP Instruction Manual January 2013 PN 51-389VP List of Tables No. Title 3-1 ORP of Saturated Quinhydrone Solution (In Millivolts)............................................................13 4-1 Ro and R1 Values for Temperature Compensation Elements ..........
Page 7 - Features and Applications; Section 1: Description and Specifications
Description and Specifications 1 389VP Instruction Manual Section 1: Description and specifications PN 51-389VP January 2013 Features and Applications 389VP triple-junction sensors are now offered with SMART capabilities. SMART option becomes enabled when use with the 1056, 1057, 1066, 56 instrument...
Page 8 - Performance and Physical Specifications
2 Description and Specifications tanks, or ponds. The 389VP is suitable for virtually all applications and is compatible with Rosemount Analytical and other manufacturers’ instruments. Installation is easily achieved through the wide variety of mounting configurations. 389VP features a 1 inch (MNPT)...
Page 9 - Ordering Information
Ordering Information The 389VP pH/ORP Sensor is offered with a Variopol (VP) connector and uses a mating VP cable (purchased separately). A remote preamplifier (j-box or instrument) must be used with this sensor, unless (-70) preamplifier option is selected. Model 389VP is SMART enable using option ...
Page 11 - Unpacking and inspection; Section 2: Installation; CAUTION; WARNING
Installation 5 389VP Instruction Manual Section 2: Installation PN 51-389VP January 2013 Unpacking and inspection Inspect the outside of the carton for any damage. If damage is detected, contact the carrier imme- diately. Inspect the hardware. Make sure all the items in the packing list are present ...
Page 12 - Submersion Mounting; Electrical Installation
6 Installation Section 2: Installation 389VP Instruction Manual January 2013 PN 51-389VP Submersion Mounting The 389 and 389VP Sensor has a 1 in. MNPT process connection at the back of the sensor. Utilizing a standard 1 in. union, the sensor may be mounted to a 1 in. SCH 80 CPVC or PVDF standpipe. T...
Page 23 - Sensor Preparation; Recommended two point buffer calibration procedure; Recommended pH Sensor Standardization
Start-up and Calibration 17 389VP Instruction Manual Section 3: Start-up and Calibration PN 51-389VP January 2013 Sensor Preparation Shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped an...
Page 24 - Calibration
18 Start-up and Calibration Section 3: Start-up and Calibration 389VP Instruction Manual January 2013 PN 51-389VP 3.3 3.3.1 2. Measure and record the pH of the process solution sample with a another temperature com- pensated, calibrated pH instrument. For best results, standardization should be perf...
Page 25 - Electrode Cleaning; Section 4: Maintenance
Maintenance 19 389VP Instruction Manual Section 4: Maintenance PN 51-389VP January 2013 The 389VP Sensor requires minimum maintenance. The sensor should be kept clean and free of debris and sediment at all times. The frequency of cleaning by wiping or brushing with a soft cloth or brush is determine...
Page 26 - Platinum Electrode Check
20 Maintenance Section 4: Maintenance 389VP Instruction Manual January 2013 PN 51-389VP 389VP ORP Platinum Electrode Check The platinum electrode may be checked as follows. There are two types of standard solutions which may be used to check the ORP electrode/transmitter system: Type 1: One type of ...
Page 27 - Cleaning Platinum Electrode
Maintenance 21 389VP Instruction Manual Section 4: Maintenance PN 51-389VP January 2013 476 +20 mV at 25°C when used with a saturated KCl/AgCl reference electrode and platinum meas- uring electrode. Some tolerance in mV values is to be expected due to the rather large liquid ref- erence junction pot...
Page 29 - Section 5: Troubleshooting
Troubleshooting 23 389VP Instruction Manual Section 5: Troubleshooting PN 51-389VP January 2013 Section 5: Troubleshooting y d e m e R e s u a C e l b a b o r P e l b u o r T Meter reads off scale (Display Defective preamplifier. Replace preamplifier (for code 02 reads overrange). sensors). For code...
Page 31 - Section 6: Return of Material; General; IMPORTANT; NOTE
Return of Material 25 389VP Instruction Manual Section 6: Return of Material PN 51-389VP January 2013 Section 6: Return of Material 6.1 General To expedite the repair and return of instruments, proper communication between the customer and the factory is important. A return material authorization nu...
Page 32 - RETURN OF MATERIALS REQUEST
26 Return of Material Section 6: Return of Material 389VP Instruction Manual January 2013 PN 51-389VP RETURNED BY: BILL TO: _____________________________ _____________________________ ____________________________ _____________________________ _____________________________ ___________________________...