Page 2 - Installation & User Manual; Table of contents; MAS2600 Ordering Information
Installation & User Manual Ed 19UK June 2013 Page 2 of 28 Table of contents Description and Operation ............................................................................................. 3 MAS2600 Tank Contents Transmitter ...................................................................
Page 3 - Page; Description and Operation; High temperature gauge version.; Application
Installation & User Manual June 2013 Ed 19UK Page 3 of 28 CERTIFICATE REF.: 4-20 mA Lint = 20 uH x MARINE AUTOMATION DENMARK S/N: Umax = 33 V T Cint = 10 nF C i EE X a 90-C.95948 T5 amb.: -40 to +85 °C I 45 mA K = < 2600 TWO WIRE TRANSMITTER 2600 SERIES S/N 46374 1.5 bar g PDCR-941-1 MAS 2600...
Page 4 - Technical Specifications; Programmable Measuring Ranges:; mA; Operating Temperature Ranges:; High temperature cable:
Installation & User Manual Ed 19UK June 2013 Page 4 of 28 0 100 200 300 400 500 600 700 17 20 23 26 29 32 Load r es is tanc e (ohm ) Power supply (V) Max. load resistance at various power supply Technical Specifications Transducer Ranges: Gauge: 0-3.5 / 0-7 / 0-16 / 0-35 m H 2 O. Absolute: 0-20....
Page 6 - Type; Transducer type; NOT
Installation & User Manual Ed 19UK June 2013 Page 6 of 28 Type MAS2600-XXX-XX-X/XX A: Absolute Transducer. The absolute transducer has no breather pipe, because it refers to absolute vacuum. An atmospheric pressure sensor is used to compensate for the atmospheric pressure. The absolute transduce...
Page 7 - Transducer type 7; Selecting the Measuring Range
Installation & User Manual June 2013 Ed 19UK Page 7 of 28 MAS 2600 Amplifier box Flange mountedG022P013 Sensor pos. abovelowest point Tankheight Vent pipe height TWO WIRE TRANSMITTER 2600 SERIES 90-C.95948 T5 Cint = 10 nF C CERTIFICATE REF.: 4-20 mA Lint = 20 uH x i EE X a MARINE AUTOMATION DENM...
Page 8 - Basic Rules; tank height + vent pipe height; Forepeak Tank
Installation & User Manual Ed 19UK June 2013 Page 8 of 28 Absolute Transducers : The Absolute Transmitter is typically used if the amplifier is exposed to water, e.g. on open deck. The Absolute Transmitter is available in 2 ranges: 0-20 and 0-35 mH 2 O. In practice this means 10 mH 2 O less than...
Page 9 - Temperature Sensor; Without temperature sensor.; Cable; C. The transducer cable is delivered in whole meters.
Installation & User Manual June 2013 Ed 19UK Page 9 of 28 Class approved Sea valve TWO WIRE TRANSMITTER 2600 SERIES 90-C.95948 T5 Cint = 10 nF C CERTIFICATE REF.: 4-20 mA Lint = 20 uH x i EE X a MARINE AUTOMATION DENMARK S/N: Umax = 33 V Tamb.: -40 to +85 °C I 45 mA K= < 2600 Weather deck MAS...
Page 10 - Brackets for Internal Mounting - drawing no. G022P010
Installation & User Manual Ed 19UK June 2013 Page 10 of 28 Brackets for Internal Mounting - drawing no. G022P010 : Bulkhead Tank bottom TWO WIRE TRANSMITTER 2600 SERIES 90-C.95948 T5 Cint = 10 nF C CERTIFICATE REF.: 4-20 mA Lint = 20 uH x i EE X a MARINE AUTOMATION DENMARK S/N: Umax = 33 V Tamb....
Page 12 - Flexible rubber tube mounting 80
Installation & User Manual Ed 19UK June 2013 Page 12 of 28 Tank bottom H Bulkhead Lint = 20 uH CERTIFICATE REF.: 4-20 mA Tank top x S/N: MARINE AUTOMATION DENMARK Umax = 33 V 90-C.95948 TWO WIRE TRANSMITTER Cint = 10 nF X EE i a T5 C 2600 SERIES I 45 mA amb.: -40 to +85 °C < = K T 2600 Bulkhe...
Page 14 - Amplifier Box; Without amplifier box.
Installation & User Manual Ed 19UK June 2013 Page 14 of 28 Amplifier Box MAS2600-XXX-XX-X/XX The MAS2600 amplifier box is manufactured in polyester, which has the strength and stability of aluminium, insulation and resistance properties of plastics. Colour code RAL 7000. The amplifier box is pro...
Page 15 - S and M
Installation & User Manual June 2013 Ed 19UK Page 15 of 28 Amplifier PCB MAS2600-XXX-XX-X/XX The PCB is available in different types for the different transducers, with or without terminals for Pt100 temperature sensor. 0: Without amplifier PCB. P or S: Programmable output range. The amplifier c...
Page 16 - Handling; lift by the cable if transducer is mounted with fittings.
Installation & User Manual Ed 19UK June 2013 Page 16 of 28 Handling The MAS2600 transducer is a highly sensitive electronic piece of equipment and proper handling is most important. To avoid damaging the transducers : • NEVER lift by the cable if transducer is mounted with fittings. • NEVER remo...
Page 17 - Installation; not
Installation & User Manual June 2013 Ed 19UK Page 17 of 28 Installation Prior to installation please check that: • the serial numbers on the transducer, amplifier and lid of amplifier box agree. The serial numbers are to be found on the inside of the lid of the amplifier box on the bottom line -...
Page 18 - -20 mA signal cable
Installation & User Manual Ed 19UK June 2013 Page 18 of 28 If the amplifier box is installed in an area with over pressure or negative pressure in relation to the atmospheric pressure in the tank, the breather tube must be directed to an area with atmospheric pressure equalling the pressure insi...
Page 19 - Intrinsically Safe Installation
Installation & User Manual June 2013 Ed 19UK Page 19 of 28 Intrinsically Safe Installation The MAS2600 is designed for use in an approved intrinsically safe installation. The amplifier is fitted with a label describing the IS category: S/N: XXXXXX-YYYY DEMKO 11 ATEX 146506X IECEx ULD 10.0013X Ui...
Page 20 - in the lid should be tightened fully.
Installation & User Manual Ed 19UK June 2013 Page 20 of 28 Before installing intrinsically safe units within an approved installation, it is essential to ensure that there are no deviations from the conditions of the order or the requirements of the approved block schematic diagram and the follo...
Page 21 - Hazardous area
Installation & User Manual June 2013 Ed 19UK Page 21 of 28 4-20 mA Hazardous area Safe area Any 28V / 300RShunt zener diode saftybarrier or Ex repeatercertified to Ex Ia IIC. Transducer Amplifier box MARINE AUTOMATION DENMARK Cint = 10 nF 2600 SERIES 4-20 mA CERTIFICATE REF.: Lint = 20 uH x EE X...
Page 22 - Testing and Recalibrating; Testing; Set the calibrator to pressure mode.
Installation & User Manual Ed 19UK June 2013 Page 22 of 28 Testing and Recalibrating Please observe that testing and recalibration of sensors mounted in hazardous arias shall be carried out of Ex qualified personnel. Testing Before dispatch, the MAS2600 transmitter is calibrated in accordance wi...
Page 24 - Transducer range; Trd range; Atm press
Installation & User Manual Ed 19UK June 2013 Page 24 of 28 Recalibrating the Absolute Transmitter using Test Cup type G022S100 When recalibrating the MAS2600 Absolute Transmitter a pressure calibrator with a range of -1 to 3.5 bar should be used. It should have an accuracy of 0.05% FS or better,...
Page 26 - Maintenance and Trouble Shooting; Maintenance; Visual inspection of the cable, transducer and amplifier.
Installation & User Manual Ed 19UK June 2013 Page 26 of 28 Maintenance and Trouble Shooting Maintenance Damcos recommends that the following to be checked annually: 1. Visual inspection of the cable, transducer and amplifier. 2. Compare signal with e.g. hand sounding, and if necessary perform a ...
Page 27 - Emerson Process Management; The tank is
Installation & User Manual June 2013 Ed 19UK Page 27 of 28 MAS2600 Test sheet ATT.: Emerson Process Management Marine Solutions Damcos A/S, Service Department FAX.: +45 5578 7272 Name: Ship name: Yard: NB No: MAS2600: Transducer serial no.: Amplifier serial no.: Tank name: Liquid in tank HFO: DO...