2 MODEL 499A CL-01 INSTALLATION FIGURE 4. Low Flow Cell (PN 24091-00 or 24091-01) INSTALLATION Install the sensor in a flowing sample. Keep the sample flowas constant as possible at a value within the limits below. Flow through 1 to 5 gpm (3.8 to 19 L/min) Open channel 1 ft/sec (0.3 m/sec) Low flow ...
MODEL 499A CL-01 WIRING WIRING 3 FIGURE 5. Wiring Sensor with Standard Cable toModel 1056 and Model 56 Analyzer FIGURE 6. Wiring Sensor with Optimum EMI/RFI orVariopol Cable to Model 1056 and Model 56 Analyzer FIGURE 7. Wiring Sensor with Standard Cable toModel 54eA Analyzer FIGURE 8. Wiring Sensor ...
FIGURE 13. Wiring Sensor with Standard Cable toModel 1066 Transmitter. FIGURE 14. Wiring Sensor with Optimum EMI/RFI orVariopol Cable to Model 1066 Transmitter. CALIBRATION Zero point: Even in the absence of free chlorine, the 499ACL-01 sensor generates a small signal called thezero current. Failing...