Page 2 - DANGER; HAZARDOUS AREA INSTALLATION; ESSENTIAL INSTRUCTIONS; READ THIS PAGE BEFORE PROCEEDING!; About This Document; Date; RETRACTABLE SENSORS; CAUTION; SENSOR/PROCESS APPLICATION COMPATIBILITY; WARNING; Special Conditions for safe use; to a build up of an electrostatic charge.
DANGER HAZARDOUS AREA INSTALLATION Installations near flammable liquids or in hazardousarea locations must be carefully evaluated by qualifiedon site safety personnel. This sensor is not IntrinsicallySafe or Explosion Proof. To secure and maintain an intrinsically safe installa-tion, the certified s...
Page 3 - RETRACTABLE pH/ORP SENSORS; TABLE OF CONTENTS; Section; LIST OF TABLES; Title
MODEL 396R pH/ORP TABLE OF CONTENTS MODEL 396R AND 396RVP RETRACTABLE pH/ORP SENSORS TABLE OF CONTENTS Section Title Page 1.0 DESCRIPTION AND SPECIFICATIONS........................................................... 1 1.1 Features and Applications .......................................................
Page 4 - LIST OF FIGURES
MODEL 396R pH/ORP TABLE OF CONTENTS LIST OF FIGURES Figure No. Title Page 1-1 Cross Section Diagram of the TUpH Reference Technology ............................ 1 2-1 Exploded View of Ball Valve Kit PN 23240-00 used with process connector .PN 23166-00 (or PN 23166-01) .................................
Page 5 - FEATURES AND APPLICATIONS; DESCRIPTION AND SPECIFICATIONS; FIGURE 1-1. Cross Section Diagram of the TUpH Reference Technology
1.1 FEATURES AND APPLICATIONS The Model 396R and 396RVP Sensors are specifically designed for improved life in harsh, dirty applications wherea separate sample stream is difficult to provide and greaterinsertion depths are required. Model 396R is designed foruse with a 1-1/4 in. or 1-1/2 in. ball va...
Page 6 - PERFORMANCE AND PHYSICAL SPECIFICATIONS
MODEL 396R pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS MODEL 396R Measured Range: ORP: -1500 to 1500mV pH: 0 to 14 Percent Linearity Over pH Range: Wetted Materials: Polypropylene, EPDM, titanium, glass, (platinum: ORP only) Process Connections: 1-1/2 or 1-1/4 in. with ball valve, 1 in. withou...
Page 7 - ORDERING INFORMATION; The; The Model 396RVP ball valve retractable sensor; features a gel-filled electrolyte solution with the large area,
3 MODEL 396R pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS 1.3 ORDERING INFORMATION The Model 396R Sensor is housed in a titanium tube, with a polypropylene reference junction and titanium solu- tion ground for use with a ball valve (order separately) for hot tap applications. The sensor is avai...
Page 8 - Retractable Mounting
4 TABLE 1-2. OTHER ACCESSORIES FOR MODEL 396R TABLE 1-1. COMMONLY USED ACCESSORIES FOR MODEL 396R PART DESCRIPTION 22698-00 Preamplifier plug-in for J-box, for Model 1003, 22698-02 Preamplifier plug-in for J-box, for Models 1181/1050 22698-03 Preamplifier plug-in for J-box, for Models 1054A/B, 2054,...
Page 9 - TABLE 1-4. OTHER ACCESSORIES FOR MODEL 396RVP; TABLE 1-3. COMMONLY USED ACCESSORIES FOR MODEL 396RVP
TABLE 1-4. OTHER ACCESSORIES FOR MODEL 396RVP PART DESCRIPTION 22698-00 Preamplifier plug-in for junction box, for Model 1003, 22698-02 Preamplifier plug-in for junction box, for Models 1181, 1050 22698-03 Preamplifier plug-in for junction box, for Models 1054B, 2081 22743-01 Pt100 preamplifier for ...
Page 10 - INSTALLATION; Inspect the outside; NOTE
6 MODEL 396R pH/ORP SECTION 2.0 INSTALLATION SECTION 2.0 INSTALLATION 2.1 UNPACKING AND INSPECTION. Inspect the outside of the carton for any damage. If damage is detected,contact the carrier immediately. Inspect the instrumentand hardware. Make sure all items in the packing listare present and in g...
Page 11 - Over tightening the hex nut may damage the ferrule.
6. While holding the sensor in position, tighten the hexnut of the male connector to firmly secure the sensorin place. When the hex nut is tightened, the Teflonferrule inside the compression fitting clamps the sen-sor tube. NOTE A stainless steel ferrule is available if theTeflon ferrule does not in...
Page 12 - FIGURE 2-2. Typical Mounting Configurations for Model 396R
8 MODEL 396R pH/ORP SECTION 2.0 INSTALLATION JUNCTION BOX IS OPTIONAL FIGURE 2-2. Typical Mounting Configurations for Model 396R FIGURE 2-3. Typical Mounting Configurations for Model 396RVP
Page 31 - FIGURE 4-15. Model 54 Wiring through Remote Junction Box
MODEL 396R pH/ORP SECTION 4.0 WIRING MODEL 396RVP FIGURE 4-14. Model 54/54e Wiring FIGURE 4-15. Model 54 Wiring through Remote Junction Box FIGURE 4-16. Model 2700 Wiring FIGURE 4-17. Model SCL-(P/Q) Wiring 27
Page 32 - FIGURE 4-20. Model 1055pH/pH Wiring through Remote Junction Boxes
28 MODEL 396R pH/ORP SECTION 4.0 WIRING MODEL 396RVP FIGURE 4-19. Model 1055pH/pH Wiring FIGURE 4-18. Model Xmt-P-XX-10 Wiring FIGURE 4-20. Model 1055pH/pH Wiring through Remote Junction Boxes
Page 34 - Recommended pH Sensor Standardization; START UP AND CALIBRATION
30 5.1 START UP. To obtain best accuracy, the sensor must be calibrated as a loop with the analyzer. Pleaserefer to the respective analyzer/transmitter instructionmanual for proper calibration procedures. 5.2 396R pH CALIBRATION 1. After a temporary connection is established between the sensor and t...
Page 35 - An ORP loop is best; Therefore, the sensor should be
5.3 396R ORP CALIBRATION. An ORP loop is best calibrated using an ORP standard solution. Most indus-trial applications have a number of ORP reactionsoccurring in sequence or simultaneously. There can beseveral components that are oxidized or reduced bythe reagents that are used. Theoretically, the O...
Page 36 - MAINTENANCE
32 6.1 Maintenance. The Model 396R Sensor is a dis- posal type sensor and therefore requires minimummaintenance. The sensor should be removed from theprocess periodically and checked in buffer solutions. Ifthe sensor will not calibrate, refer to youranalyzer/transmitters instruction manual for prope...
Page 38 - Pull back hard on the sensor assembly, as if trying
34 4. Discard used O-ring from male connector body(A). Coat new O-ring with a thin film of the O-ringlubricant provided. Position it in the machined O-ring groove in place of the discarded O-ring. CAUTION Make sure lubricant does not contact any part of thesensor tip particularly the glass bulb. 5. ...
Page 40 - DIAGNOSTICS AND TROUBLESHOOTING; DESCRIPTION OF PROBLEM; DBMJcsB; TABLE 7-1 Troubleshooting with Diagnostics
36 MODEL 396R pH/ORP SECTION 7.0 DIAGNOSTICS AND TROUBLESHOOTING SECTION 7.0 DIAGNOSTICS AND TROUBLESHOOTING 7.1 DIAGNOSTICS AND TROUBLESHOOTING WITH MODEL 54/81/3081 pH/ORP. The Model 54 and 54e Analyzers and Models 81 and 3081 pH Transmitters automatically search for fault conditions that would ca...
Page 41 - Problem
37 MODEL 396R pH/ORP SECTION 7.0 DIAGNOSTICS AND TROUBLESHOOTING Problem Probable Cause Remedy Meter reads off scale. (Display Defective preamplifier. Replace preamplifier (for code 02 reads overrange). sensors). For code 01, replace sensor. T.C. element shorted. Check T.C. element as instructed in ...
Page 42 - RETURN OF MATERIAL; Call Rosemount Analytical for authorization.
38 Model 396R pH/ORP SECTION 8.0 RETURN OF MATERIAL SECTION 8.0 RETURN OF MATERIAL 8.1 GENERAL. To expedite the repair and return of instruments, pleasec a l l 1 - 9 4 9 - 7 5 7 - 8 5 0 0 f o r a R e t u r n M a t e r i a l sAuthorization (RMA) number. 8.2 WARRANTY REPAIR. The following is the proce...
Page 43 - Emerson Process Management; The shipping container should be marked:
WARRANTY Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufacturedor Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira-tion of the applicable warranty peri...
Page 44 - Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only. The right people,the right answers,right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE http://www.raihome.com Specifications subject to change without notice. 8 Emerson Process Management 2400 Barranca ParkwayIrvine, CA 92606 USATel: (949) 757-8500Fax: (94...