Page 2 - DANGER; ESSENTIAL INSTRUCTIONS; READ THIS PAGE BEFORE PROCEEDING!; About This Document; Date
CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor materials may not be com-patible with process composition and operat-ing conditions. Application compatibility isentirely the responsibility of the user. DANGER HAZARDOUS AREA INSTALLATION This sensor is not Intrinsically Safe. orExp...
Page 3 - TABLE OF CONTENTS
i MODEL 396/396VP/397/398/398VP TABLE OF CONTENTS TUpH MODELS 396/396VP/397/398/398VP pH/ORP SENSORS TABLE OF CONTENTS Section Title Page 1.0 DESCRIPTION AND SPECIFICATIONS ......................................................... 1 1.1 TUpH™ Features and Applications ..................................
Page 4 - LIST OF FIGURES; Number
MODEL 396/396VP/397/398/398VP TABLE OF CONTENTS LIST OF FIGURES Number Title Page 1-1 Cross Section Diagram of the TUpH Patented Reference Technology............. 2 2-1 Flow Through/Insertion Installation for Models 396, 396VP, 398, & 398VP .............. 10 2-2 Dimensional Drawing For Models 39...
Page 5 - DESCRIPTION AND SPECIFICATIONS; for maximum; Features and; pH bulb allows for ultimate chemical
SECTION 1.0 DESCRIPTION AND SPECIFICATIONS 1.1 TUpH™ FEATURES AND APPLICATIONS. Rosemount Analytical has achieved a new industrystandard for the life expectancy of pH sensors withthe TUpH Sensors. The low maintenance, disposableModels 396, 396VP, 397, 398, and 398VP TUpHSensors offer long life and h...
Page 7 - PERFORMANCE & PHYSICAL SPECIFICATIONS; SPECIFICATIONS; PERCENT LINEARITY
3 MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS 1.2 PERFORMANCE & PHYSICAL SPECIFICATIONS SPECIFICATIONS MODELS 396 &396VP MODELS 398 & 398VP MODEL 397 QUIK-LOC KIT Measurements pH: 0-14 pH: 0-14 pH: 0-14 — and Ranges ORP: -1500 to 1500 mv Available pH GPHT hem...
Page 8 - ORDERING INFORMATION; The Model 396 pH Sensor; features your selection of glass electrode type, either the; The Model 396VP insertion/submersion sensor; is the same as Model 396 except,
4 MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS 1.3 ORDERING INFORMATION MODEL 396 TUpH INSERTION/SUBMERSION STAINLESS STEEL pH SENSOR CODE ANALYZER/TC COMPATIBILITY (Required Selection) 50 For use with Models 1181, 1050, 1003 (3 K TC) 54 For use with Models 54/e, 1054A/B,...
Page 9 - The; The Model 398VP insertion/submersion sensor; is the same as Model 398, except
MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS The Model 398 Sensor, housed in a titanium tube and made with a patented Tefzel ref- erence junction, can be used with a 1 inch MNPT process connector (purchased sepa-rately). The sensor is available with a hemi bulb glass pH e...
Page 10 - The TUpH Model 397 pH Sensor
6 MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS The TUpH Model 397 pH Sensor is housed in a polypropylene body and is designed to be used with the Quik-Loc Kit. The sensor includes a large general purpose(GPHT) glass pH electrode and a large area polypropylene reference ju...
Page 13 - NOTE; NOTE; INSTALLATION
9 2.2.1 Flow Through and Insertion Mounting forModels 396, 396VP, 398, and 398VP. The Models 396, 396VP, 398, and 398VP Sensors can be used with a 1inch MNPT process connector at the front of the sensorfor mounting into a 1-1/2 inch tee or the process. SeeFigure 2-1 for installation configurations. ...
Page 14 - spring arm has been provided on one arm so that the; The sensor may obstruct flow through smaller pipes.; SAFETY WARNING
10 MODEL 396/396VP/397/398/398VP SECTION 2.0 INSTALLATION Wrap the pipe threads of the Twin-Kam Kamloc couplerwith Teflon tape before placing it into the process pipe. Thecoupler can be connected to any 1 in. process connectionand must be mounted within 80° of vertical, with the elec-trode facing do...
Page 15 - FIGURE 2-2. Dimensional Drawing For Models 396 and 398
11 MODEL 396/396VP/397/398/398VP SECTION 2.0 INSTALLATION FIGURE 2-2. Dimensional Drawing For Models 396 and 398 DWG. NO. REV. 40039601 J MILLIMETER INCH
Page 18 - FIGURE 2-5. Dimensional Drawings For Model 397 in the Quik-Loc Unit
14 MODEL 396/396VP/397/398/398VP SECTION 2.0 INSTALLATION FIGURE 2-5. Dimensional Drawings For Model 397 in the Quik-Loc Unit DWG. NO. REV. 40039701 C MILLIMETER INCH DWG. NO. REV. 40039702 C DWG. NO. REV. 40039704 B NOTE: The EP gasket (see drawing #40039702)provided with the coupler should be peri...
Page 19 - DO NOT CONNECT SENSOR CABLE TO POWER
15 MODEL 396/396VP/397/398/398VP SECTION 2.0 INSTALLATION 2.3 ELECTRICAL INSTALLATION. The Models 396, 396VP, 397, 398, and 398VP are for usewith a remote preamplifier. Each sensor comes with either aspecial 15 ft low noise coax cable or a Variopol (VP) con-nector, which is used with a mating Variop...
Page 35 - START UP AND CALIBRATION; CALIBRATION USING BUFFER
MODEL 396/396VP/397/398/398VP SECTION 3.0 START UP AND CALIBRATION SECTION 3.0 START UP AND CALIBRATION 3.1 SENSOR PREPARATION. Shake down the sensor to remove any air bubblesthat may be present at the tip of the pH glass bulb. Inmost cases, the pH sensor can simply be installed asshipped and readin...
Page 36 - MAINTENANCE
32 MODEL 396/396VP/397/398/398VP SECTION 4.0 MAINTENANCE 4.0 MAINTENANCE. The Model 396, 396VP, 397, 398, and 398VP Sensors are disposable type sensorsand therefore require minimum maintenance. Everysensor should be kept clean and free of debris andsediment at all times. The frequency of cleaning by...
Page 37 - Trouble; TROUBLESHOOTING
33 MODEL 396/396VP/397/398/398VP SECTION 5.0 TROUBLESHOOTING Trouble Probable Cause Remedy Meter reads off scale. (Display T.C. element shorted. Check T.C. element as instructed reads overrange.) in Section 4.2 andreplace sensor if defective. Sensor not in process or sample Make sure sensor is in pr...
Page 38 - RETURN OF MATERIAL; Call Rosemount Analytical for authorization.
34 MODEL 396/396VP/397/398/398VP SECTION 6.0 RETURN OF MATERIAL SECTION 6.0 RETURN OF MATERIAL 6.1 GENERAL. To expedite the repair and return of instruments, propercommunication between the customer and the factoryis important. Before returning a product for repair, call1-949-757-8500 for a Return M...
Page 39 - The shipping container should be marked:
WARRANTY Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufacturedor Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira-tion of the applicable warranty peri...
Page 40 - Specifications subject to change without notice.
Credit Cards for U.S. Purchases Only. The right people,the right answers,right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE http://www.raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca ParkwayIrvine, CA 92606 USATel: (949) 75...