Emerson 141 - Manuals

Emerson 141 – Manual in PDF format online.

Manuals:

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12

Summary

Page 2 - INSTALLATION; sensor into the process pipe.; Sensor Section on page 6.; SIDE VIEW

Keep 1/4 inch (6 mm) clearance between electrodesand piping. The electrodes must be completely sub-merged in the process liquid, i.e., to the level of thethreaded connection. See Figure 1 for recommendedorientation. If the sensor is installed in a side stream with thesample draining to open atmosphe...

Page 3 - FIGURE 3. Installing model 141 sensor; INSTALLATION – 141 SENSOR; fitting until the sensor is correctly positioned.; FIGURE 4. Installing model 142 sensor

3 MODEL 140,141, and 142 sensors INSTALLATION FIGURE 3. Installing model 141 sensor DWG. NO. REV. 40014204 B INSTALLATION – 141 SENSOR 1. Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee.2. Remove the plastic shipping cap from the sensor.3. Screw the sensor into the fitting. Use...

Page 4 - WIRING; WIRE COLOR AND CONNECTIONS IN SENSOR; COLOR FUNCTION; FIGURE 6. Model 54eC Wiring; WIRING DIAGRAMS

RTD RTD in RTD sense RTD return MODEL 140,141, and 142 sensors WIRING WIRING All 140 series sensors have a junction box mounted on the back of the sensor. Wiring connections in the junction boxare shown in Figure 5. Figure 5. Sensor junction box wiring. Terminals in thejunction box are not numbered....