Page 3 - Amplitude Specifications; Amplitude Range; Average Continuous Power; Gain Compression; >10 MHz; Displayed Average Noise Level; kHz RBW; Nominal Dynamic Range

3 Amplitude Specifications Amplitude specifications do not apply for Analog+mode and negative peak detector mode except asnoted in “Amplitude Characteristics.” Amplitude Range Displayed average noise levelto +30 dBm 8591E (Opt. 001) Displayed average noise level to+72 dBmV Maximum Safe Input Level (...

Page 5 - Audible Noise

5 Option Specifications Option 010 and 011 Tracking GeneratorFrequency Range 8591E 100 kHz to 1.8 GHz (Opt. 011, 75 Ω ) 1 MHz to 1.8 GHz 8593E, 94E, 95E, 96E 9 kHz to 2.9 GHz Output Level Range 8591E 0 to –70 dBm 8591E (Opt. 011) +42.8 to –27.2 dBmV 8593E, 94E, 95E, 96E –1 to –66 dBm Resolution 0.1 ...

Page 6 - Power Requirements; User Program Memory; Weight

6 Power Requirements ON (Line 1) 90 to 132 V rms, 47 to 440 Hz195 to 250 V rms, 47 to 66 HzPower consumption <500 VA; <180W Standby (Line 0) Power consumption <7 W User Program Memory 238 Kbytes non-volatile RAM Data Storage (nominal) Internal 24 traces or 32 states External 50 traces, 8 st...