Page 2 - Table of Contents
2 3 5 5 6 7 8 9 9 9 10 11 11 11 11 12 13 15 15 16 17 17 19 Table of Contents Vector network analyzer (VNA) overview Agilent 8510 family of products Agilent 8510 network analyzer Agilent 8510 firmware upgrade kits Test sets Sources Upgrade considerations Upgrading individual components Upgrading the ...
Page 3 - Figure 1. Device characterization; Figure 2. Network analyzer configuration; information on a CRT; DUT; Transmitted
3 Figure 1. Device characterization DUT Processor/Display Receiver/Detector Signal separation Source Incident Reflected Transmitted Incident Reflected Transmitted Figure 2. Network analyzer configuration Vector network analyzer (VNA) overview Vector network analyzer (VNA) measurement systems are use...
Page 4 - Source
4 8510 VNA system Source 8360 Test set 851X Receiver/detector 8510 processor/display network analyzer Signal separation Stimulus = + + Figure 3. 8510 measurement systems are made up of modular components, the 8510 network analyzer, a test set and a compatible source. The Agilent Technologies 8510 ne...
Page 5 - Agilent 8510 network analyzer; Table 1. Hardware revision and measurement capability; Figure 4. 8510 family of system solutions; Agilent 8510 family of products
5 Agilent 8510 network analyzer Tables 1 and 2 show the different features of the 8510A, B, and C, hardwareand firmware, respectively. Table 1. Hardware revision and measurement capability Product Firmware revisions Hardware features 8510A A.02.00 • Tape drive 8510B B.03.00 to B.06.54 • Tape drive 8...
Page 6 - Table 2. Firmware revision and measurement capability; Agilent 8510 Firmware upgrade kits; Figure 5. 8510 firmware upgrade kits
6 Table 2. Firmware revision and measurement capability Revision Time domain 1 Pulse 2 Power domain Other added key features limit lines A.02.00 B.03.00 • Adapter removal calibration• Direct control of two sources B.03.11 • Direct control of up to 4 test sets B.04.00 • 40 GHz test set capability• Fr...
Page 7 - Test Sets; Table 3. Family of test sets
Test Sets The test set in a VNA measurementsystem is used to separate the inci-dent, the reflected, and the transmittedsignals. It is also used to convert the RF signal to IF (intermediate fre- quency) signal and to pass the IFto the receiver. Once signals are separated, their individual magnitude a...
Page 8 - Sources; Table 5. 8360 general purpose synthesizers
8 Sources The RF or microwave signal source in a VNA measurement system providesthe incident signal used to stimulate the device-under-test (DUT). The DUTresponds by ref lecting part of the incident and transmitting the remainingpart. By sweeping the frequency of the source, the frequency response o...
Page 9 - Upgrade considerations; Upgrading individual components
9 Upgrade considerations There are numerous reasons for upgrading the Agilent 8510 system. The 8510network analyzer maintains excellent performance while adapting to differentmeasurement requirements such as basic component testing, on-wafer probing,pulsed device characterization, antenna and RCS (R...
Page 11 - Test set upgrades; multiple; Expanding measurement capabilities; Adds pulsed-RF measurement capability (Option 008) to the 8510C; Adding time domain capabilities
11 Test set upgrades A single 8510 network analyzer can be configured to control up to four testsets. The operator can switch between test sets, without reconnections, usingfront panel controls. This arrangement is often referred to as the multiple test set configuration. Each test set must be equip...
Page 12 - Upgrading systems; Figure 8. 8510 family of system solutions; Necessary accessories such as calibration kits,
12 Upgrading systems Figure 8 shows the different 8510 systems and their components. This figure is an excellent guide when upgrading systems. Figure 8. 8510 family of system solutions To upgrade from one system to another, simply order the pieces that are neededin the new system that are currently ...
Page 13 - What test set do I need?; type are critical when selecting the appropriate test set.; What source do I need?; System upgrades
13 Here is a checklist of items to consider when upgrading systems: 1. Which 8510 do I have? The 8510C is recommended for the latest available measurement capabilities.• To upgrade an 8510A to an 8510C, order 85103E.• To upgrade an 8510B to an 8510C, order 85103F.• Make sure the appropriate options ...
Page 15 - current system
15 Examples of upgrade paths Figure 10. Upgrading an 85107B 50 GHz system to an 85106D waveguide system to operate in V-band (or W-band) Measurement accessories There are measurement accessories for seven device connector types: 7 mm,3.5 mm, 2.92 mm, 2.4 mm, 1.85 mm, 1 mm and Type-N. Calibration kit...
Page 16 - Calibration kits
16 Calibration kits Error correction requires that the systematic errors in the measurement system be characterized by measuring known devices (standards) over the frequency range of interest with the process of calibration. All calibration kitscontain standards used for this purpose. The standards ...
Page 17 - Verification kits; airline, mismatched airline and fixed; Test port cables and adapters; This cable arrangement is for applications where the device
17 Verification kits Verification kits are used to verify the performance specifications of an 8510system. All kits include a precision Z o airline, mismatched airline and fixed attenuators. Measured data and uncertainties traceable to the U.S. NationalInstitute of Standards and Technology (NIST) ar...
Page 18 - For 85110L Test sets (7 mm rugged test port connectors); For 8510XF systems (1 mm test port connectors)
18 Configuration B. This cable arrangement is for applications where the device under test is connected between cable ends. This setup offers more f lexibilitywhen connecting to the device under test. For 8514B/8515A/85110A test sets (3.5 mm rugged test port connectors) Cables/adapter Connector type...
Page 19 - Related literature; Color Brochure; Product overviews
19 Related literature Pub. number Agilent 8510 Network Analyzer, Color Brochure 5091-8970E Agilent 8510 System Solutions 5965-8837E Agilent 8510 Family Network Analyzer, Configuration Guide 5091-8967E Agilent 8510 Family Network Analyzer, Data Sheet 5091-8484E Agilent 85103C/D Upgrade Packages to th...