Page 2 - Notices; Manual Printing History
2 Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prio...
Page 3 - Certification
3 • Ground The Instrument To avoid electric shock hazard, the instrument chassis and cabinet must be connected to a safety earth ground by the supplied power cable with earth blade. • DO NOT Operate In An Explosive Atmosphere Do not operate the instrument in the presence of flammable gasses or fumes...
Page 4 - Exclusive Remedies
4 Warranty This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period corresponding to the individual warranty periods of its component products. Instruments are warranted for a period of one year. Fixtures and adapters are warranted for a peri...
Page 5 - Assistance; Safety Symbol
5 Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Office. Addresses are provided at the back of this manual. Safety Symbol General defin...
Page 7 - Contents
Contents 7 1. Operation Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Typical Characteristics . . . . . . . ....
Page 9 - Operation
9 1 Operation This operating note provides complete information on the 16048A Test Leads. The 16048A is shown pictorially in Figure 1-1, its physical dimensions are given in Table 1-1, and typical characteristics related to offset error are given in Table 1-2. To order additional copies of this oper...
Page 10 - Product Description
10 Chapter 1 Operation Product Description Product Description The 16048A consists of a direct attachment, 4-terminal pair type fixture which is equipped with four BNC (m) terminated-coaxial test leads. These test leads are used to attach user-fabricated test fixtures. DC bias levels of up to ±40V c...
Page 11 - Specifications
Chapter 1 11 Operation Specifications Specifications Table 1-1 Specifications of the 16048A Function: For use with 4 terminal-pair LCR Meters and Impedance Analyzers. Connector Type: BNC male Cable Length: Approximately 1m Maximum Voltage: ± 40V peak max. (AC+DC) Weight: 315 g Safety Standards: EN61...
Page 12 - Typical Characteristics
12 Chapter 1 Operation Typical Characteristics Typical Characteristics Table 1-2 Typical Characteristics NOTE f is the measurement frequency in megahertz. The incremental errors calculated from the equation in the table for measurements at frequencies above 1 MHz are additive. Figure 1-2 Incremental...
Page 13 - Compensation for Fixture Residual Impedance Error
Chapter 1 13 Operation Compensation for Fixture Residual Impedance Error Compensation for Fixture Residual Impedance Error The 16048A has inherent stray capacitance, residual inductance, and residual resistance that affect the accuracy of measured values. To compensate for, or negate, these residual...
Page 15 - Service
Page 16 - Maintenance
16 Chapter 2 Service Maintenance Maintenance An exploded view of the 16048A (for parts identification) is shown in Figure 2-1. Do not disassemble any further than shown. Maintenance consists principally of cleaning contacts and replacing worn or damaged parts. Take special care when cleaning contact...
Page 21 - REGIONAL SALES AND SUPPORT OFFICES
REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, andfor a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact oneof the following centers and ask for a test ...