Page 2 - May be covered by one or more pat. - see
Allegra X-30 Series Centrifuges Instructions For Use PN B05529AF (March 2022) ©2022 Beckman Coulter, Inc. All rights reserved Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc. in the...
Page 3 - iii; Revision History; For labeling updates, go to
PN B05529AF iii Revision History For labeling updates, go to www.beckman.com/techdocs and download the latest version of the manual or system help for your instrument. Issue AD, 11/2015 Changes were made to: Introduction, CE Mark. Issue AE, 09/2018 Changes were made to: CHAPTER 1, Available Rotors ....
Page 4 - iv
Page 5 - Safety Notice; to how to proceed in any situation,; DANGER; will result in death or serious injury.; WARNING; could result in death or serious injury.; CAUTION; IMPORTANT; Safety During Installation and/or Maintenance
PN B05529AF v Safety Notice Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to operate the instrument. Do not attempt to perform any procedure before carefully reading all instructions. Always follow product labeling and manufacturer’s recommendations. I...
Page 6 - Electrical Safety; the instrument and damage electrical or mechanical components.; Safety Against Risk of Fire; For safe operation of the equipment, observe the following:
PN B05529AF vi Safety Notice Electrical Safety Electrical Safety To reduce the risk of electrical shock, this equipment uses a three-wire electrical cord and plug to connect this equipment to earth-ground. To preserve this safety feature: • Make sure that the matching wall outlet receptacle is prope...
Page 7 - vii; Chemical and Biological Safety; safety precautions are taken.
PN B05529AF vii Safety Notice Chemical and Biological Safety • Maintain a 7.6-cm (3-in.) clearance envelope around the centrifuge while it is running. During operation you should come within the envelope only to adjust instrument controls, if necessary. Never bring any flammable substances within th...
Page 8 - viii
PN B05529AF viii Safety Notice Chemical and Biological Safety
Page 9 - ix; Contents
ix Contents Revision History , iii Safety Notice , v Alerts for Danger, Warning, Caution, Important, and Note , v Safety During Installation and/or Maintenance , v Electrical Safety , vi Safety Against Risk of Fire , vi Mechanical Safety , vi Chemical and Biological Safety , vii Introduction , xv In...
Page 10 - Operation
x Contents Name Rating Plate , 1-8 Nonrefrigerated Model Specifications , 1-8 Refrigerated Model Specifications , 1-9 Available Rotors , 1-10 CHAPTER 2: Operation , 2-1 Introduction , 2-1 Run Procedure , 2-2 Preparation and Loading , 2-2 Entering Run Parameters , 2-5 Starting a Run , 2-9 Pulse Funct...
Page 11 - xi; Installation; Introduction; Allegra X-30 Series Centrifuge Warranty
xi Contents APPENDIX A: Installation , A-1 Introduction , A-1 Installing the Instrument , A-1 Transportation Safety Devices , A-2 Electrical Requirements , A-2 Test Run , A-4 Beckman Coulter, Inc. Allegra X-30 Series Centrifuge Warranty
Page 12 - xii; Illustrations
xii Illustrations Illustrations 1.1 Interior View of the Rotor Chamber , 1-3 1.2 The Power Switch , 1-4 1.3 System Panels , 1-4 1.4 System Keys , 1-5 1.5 Program Keys , 1-6 1.6 Digital Displays , 1-7 2.1 Tapered Sleeve Position , 2-2 2.2 Default Parameters , 2-6 3.1 Screw Locations , 3-4 3.2 Interna...
Page 13 - xiii; Tables; Error Message Chart
xiii Tables Tables 2.1 Acceleration/Deceleration Times (in minutes:seconds) , 2-9 3.1 Error Message Chart , 3-2 3.2 Troubleshooting Chart , 3-3
Page 14 - xiv
Page 15 - xv; Intended Use; In Vitro Diagnostic; Certification
PN B05529AF xv Introduction Intended Use For In Vitro Diagnostic Use. The Allegra X-30 Series centrifuges are intended to separate human samples, including blood, urine, and other bodily fluids to prepare samples for downstream in vitro diagnostic procedures, which may include molecular diagnostic, ...
Page 16 - xvi; NOTE; Conventions; Typographic Conventions; display names; Cursor keys; CFC-Free Centrifugation
PN B05529AF xvi Introduction Conventions • APPENDIX A, Installation contains instructions for installing and connecting the centrifuge. NOTE If the centrifuge is used in a manner other than specified in this manual, the safety and performance of this equipment could be impaired. Further, the use of ...
Page 17 - xvii; CE Mark
PN B05529AF xvii Introduction CE Mark It is very important that customers understand and follow all laws regarding the proper decontamination and safe disposal of electrical equipment. For Beckman Coulter products bearing this label please contact your dealer or local Beckman Coulter office for deta...
Page 18 - xviii
Page 19 - CHAPTER 1; Description; Centrifuge Function and Safety Features; Centrifuge Function; in; Models; Nonrefrigerated
PN B05529AF 1-1 CHAPTER 1 Description Introduction This chapter provides a brief physical and functional description of the Beckman Coulter Allegra X-30 Series centrifuges. The operating controls and indicators are also described; instructions for their use are in CHAPTER 2, Operation . Chemical com...
Page 20 - Safety Features; Housing; Troubleshooting; Rotor Chamber
PN B05529AF 1-2 Description Chassis Safety Features Allegra X-30 Series centrifuges have been designed and tested to operate safely indoors at altitudes up to 2000 m (6562 ft). Instrument safety features include: • The door has an electromechanical door-locking mechanism to prevent operator contact ...
Page 21 - Drive; Temperature Sensing and Control (refrigerated models only)
PN B05529AF 1-3 Description Temperature Sensing and Control (refrigerated models only) 1 Figure 1.1 Interior View of the Rotor Chamber Drive The asynchronous three-phase direct-drive motor is brushless for clean, quiet operation. A tie-down screw is used to attach the rotor to the drive shaft. The r...
Page 22 - Controls and Indicators; Power Switch; The power switch is located on the centrifuge right side panel (see; Control Panel; programming keys, and digital displays (see; light
PN B05529AF 1-4 Description Controls and Indicators Controls and Indicators Power Switch The power switch is located on the centrifuge right side panel (see Figure 1.2 ). This two-position rocker switch ( I , on; O , off) controls electrical power to the centrifuge. Figure 1.2 The Power Switch NOTE ...
Page 23 - System Keys
PN B05529AF 1-5 Description Controls and Indicators 1 System Keys The centrifuge operation is controlled through the system keys (see Figure 1.4 ). Each key (except the PULSE key) has an LED in the upper left corner that lights to indicate that the key can be activated. Figure 1.4 System Keys Progra...
Page 25 - Digital Displays; represent selected acceleration and deceleration profiles (see
PN B05529AF 1-7 Description Controls and Indicators 1 Digital Displays Digital displays indicate rotor speed, run time, rotor chamber temperature, and numbers that represent selected acceleration and deceleration profiles (see Figure 1.6 ). When the power is turned on, they show the operating parame...
Page 26 - Name Rating Plate
PN B05529AF 1-8 Description Name Rating Plate Name Rating Plate The name rating plate is affixed to the rear of the centrifuge. Check that the line voltage agrees with the voltage listed on this name rating plate before connecting the centrifuge. Always mention the serial number and the model number...
Page 27 - Refrigerated Model Specifications
PN B05529AF 1-9 Description Refrigerated Model Specifications 1 Refrigerated Model Specifications Only values with tolerances or limits are guaranteed data. Values without tolerances are informative data, without guarantee. Noise level 0.91 m (3 ft) in front of instrument (approx.) £68 dBa Installat...
Page 28 - Available Rotors
PN B05529AF 1-10 Description Available Rotors Available Rotors See the applicable rotor manual for information on rotor use, care and maintenance, and rotor accessories. Electrical supply Class I Maximum heat dissipation into room under steady - state conditions 3311 Btu/h (0.97 kW) Noise level 0.91...
Page 31 - CHAPTER 2; turned off; identified in the World Health Organization; Laboratory Biosafety Manual; on the centrifuge while it is operating.
PN B05529AF 2-1 CHAPTER 2 Operation Introduction This section contains operating procedures for the centrifuge, using any of the Beckman Coulter rotors designed for use in these centrifuges. Refer to the applicable rotor manual for instructions on preparing the rotor for centrifugation. To prevent c...
Page 32 - Run Procedure; Preparation and Loading; Press the power switch to on (; key and lift the door up it remains in the open position.
PN B05529AF 2-2 Operation Run Procedure Run Procedure The following detailed operating procedures are summarized at the end of this section. If you are an experienced user of this centrifuge, you can turn to the summary for a quick review of operating steps. Preparation and Loading For fast temperat...
Page 33 - shaft and carefully lowering it straight down.; Lower the yoke straight down onto the drive shaft.
PN B05529AF 2-3 Operation Run Procedure 2 CAUTION Do not drop the rotor onto the drive shaft. The shaft can be damaged if the rotor is forced sideways or dropped onto it. Install the rotor by centering it over the shaft and carefully lowering it straight down. 6 Install the rotor according to the in...
Page 34 - Seat the filled buckets onto the yoke pins.; For a fixed angle rotor, the steps include the following:; Lower the rotor straight down onto the drive shaft.
PN B05529AF 2-4 Operation Run Procedure c. Seat the filled buckets onto the yoke pins. • Be sure to fill all positions on the yoke with buckets. For a fixed angle rotor, the steps include the following: a. Lower the rotor straight down onto the drive shaft. b. Tighten the tie-down screw clockwise on...
Page 35 - Always run any type of rotor with a balanced load.; the drive shaft and the tie-down screw is tight before each run.; Entering Run Parameters; can then be used for the current run or altered as required.
PN B05529AF 2-5 Operation Run Procedure 2 c. Attach the lid and tighten it. Some rotors have a knob and others have a bolt for the T-handle wrench. NOTE Fixed-angle rotors can be centrifuged without lids attached when longer tubes are used. CAUTION Always run any type of rotor with a balanced load. ...
Page 36 - Selecting a Rotor Number; Setting Run Speed
PN B05529AF 2-6 Operation Run Procedure Figure 2.2 Default Parameters Selecting a Rotor Number Each rotor that can be run in the centrifuge has a rotor number; the number is engraved on the rotor (and on the lid of fixed angle rotors). The centrifuge internal memory contains a list of all rotors tha...
Page 37 - Setting RPM; Setting Run Time; Timed Run
PN B05529AF 2-7 Operation Run Procedure 2 Setting RPM 1 Press the RPM key. • The last digit on the SPEED display ( 0 ) flashes, indicating that the RPM can be entered (in 100-RPM increments) with the cursor keys. 2 Press the ▲ or ▼ cursor key until the required RPM is displayed. • The corresponding ...
Page 38 - Setting Run Temperature (refrigerated models only); Setting Acceleration Rate
PN B05529AF 2-8 Operation Run Procedure The last digit on the TIME display flashes, indicating that the time can be entered with the cursor keys. 2 Press the ▲ or ▼ cursor key until the required run duration is displayed. Setting Run Temperature (refrigerated models only) Run temperature can be set ...
Page 39 - Setting Deceleration Rate; Starting a Run
PN B05529AF 2-9 Operation Run Procedure 2 Setting Deceleration Rate The DECEL key is used to select deceleration rates that maintain optimum separation. When the DECEL key is pressed, the ACC/DEC display flashes, indicating that one of the ten preset deceleration rates can be entered with the cursor...
Page 40 - on the centrifuge during operation.; Pulse Function
PN B05529AF 2-10 Operation Run Procedure 1 Check that all parameters are correct and the door is shut and latched. 2 Press the ENTER key, then the START key. • As the run begins, the instrument rotor identification system compares the rotor in the chamber to the entered rotor number and the speed en...
Page 41 - Changing Parameters During a Run; Stopping a Run; Disabling the Audible Tones
PN B05529AF 2-11 Operation Run Procedure 2 • If the PULSE key is pressed while the rotor is running at set speed, the rotor continues running at speed until the PULSE key is released. When the PULSE key is released, the rotor begins decelerating to 0rpm using maximum deceleration. • The centrifuge m...
Page 42 - After the rotor stops spinning; Unloading; centrifuge and accessories.
PN B05529AF 2-12 Operation Unloading Press and hold the STOP key for deceleration at the maximum rate (see Table 2.1 ). NOTE If you hold the STOP key for more than a full second, the deceleration process cannot be interrupted; the centrifuge cannot be restarted until the rotor comes to a complete st...
Page 43 - Summary of Run Procedures; Always run the rotor with a balanced load; Do not lift or move the centrifuge while the rotor is spinning.
PN B05529AF 2-13 Operation Summary of Run Procedures 2 Summary of Run Procedures For runs at other than room temperature, refrigerate or warm the rotor beforehand for fast equilibration. For low-temperature runs, precool the centrifuge by running a 30-minute cycle (with a precooled rotor installed) ...
Page 44 - light comes on, press the; key to
PN B05529AF 2-14 Operation Summary of Run Procedures 8 After the rotor stops spinning and the OPEN DOOR light comes on, press the OPEN DOOR key to release the door latch; open the door. 9 Unload the rotor according to instructions in the applicable rotor manual. CAUTION If disassembly reveals eviden...
Page 45 - CHAPTER 3; required. Maintenance procedures are contained in; User Messages; memory and critically interfere with future operation.; to
PN B05529AF 3-1 CHAPTER 3 Troubleshooting Introduction This section lists possible malfunctions, together with probable causes and corrective actions required. Maintenance procedures are contained in CHAPTER 4, Care and Maintenance . For any problems not covered here contact Beckman Coulter Field Se...
Page 47 - Other Possible Problems
PN B05529AF 3-3 Troubleshooting Other Possible Problems 3 Other Possible Problems Possible malfunctions that may not be indicated by diagnostic messages are described in Table 3.2 , along with probable causes and corrective actions required. Possible causes for each problem are listed in the probabl...
Page 48 - Turn the power switch to off (; Make sure that the rotor is not spinning.
PN B05529AF 3-4 Troubleshooting Emergency Access Emergency Access If the facility power fails only momentarily, the centrifuge resumes operation when power is restored and the rotor returns to set speed. However, if the rotor comes to a complete stop you have to restart the run when the power is res...
Page 49 - Set the covers aside.; grounding wires or the wires will be damaged.
PN B05529AF 3-5 Troubleshooting Emergency Access 3 4 Set the covers aside. 5 Insert a Phillips-head screwdriver through one of the exposed holes in the centrifuge front panel and loosen, but do not remove, the screw. a. Repeat for the other screw. 6 Move the centrifuge forward on the bench until the...
Page 50 - Never try to slow or stop the rotor by hand.
PN B05529AF 3-6 Troubleshooting Emergency Access DANGER Never try to slow or stop the rotor by hand. 11 After removing the rotor, replace the front panel, screws, and covers.
Page 51 - CHAPTER 4; Care and Maintenance; discussed in; Maintenance; Preventive Maintenance; long service life of the centrifuge.
PN B05529AF 4-1 CHAPTER 4 Care and Maintenance Introduction For maintenance not covered in this manual, contact Beckman Coulter Field Service. User messages are discussed in CHAPTER 3, Troubleshooting . Refer to the applicable rotor manual and to Chemical Resistances (publication IN-175) for instruc...
Page 52 - Cleaning; Chemical Resistances
PN B05529AF 4-2 Care and Maintenance Cleaning 2 Regularly check the air intake and exhaust vents for obstructions. a. Keep vents clear and clean. 3 Use a sponge to wipe condensation out of the chamber bowl between runs to prevent chamber icing (refrigerated models). a. If chamber icing occurs, defro...
Page 53 - Tube Breakage; clean the interior of the chamber bowl.; glass fragments may be embedded in their surfaces.; Carefully remove any glass particles that may remain.; Decontamination; perform appropriate decontamination procedures. Refer to; Sterilization and Disinfection; may be used on this surface. See
PN B05529AF 4-3 Care and Maintenance Tube Breakage 4 4 Clean the centrifuge case and door by wiping with a cloth dampened with diluted Solution 555. Do not use acetone or other solvents. Tube Breakage If a glass tube breaks, and all the glass is not contained in the bucket or rotor, you must thoroug...
Page 54 - Storage and Transport; Storage; described in the specifications given in; Returning a Centrifuge; packaging and shipping instructions.; Supply List; Replacement Parts
PN B05529AF 4-4 Care and Maintenance Storage and Transport Storage and Transport Storage Before storing a centrifuge for an extended period, return it to its original shipping container to protect it from dust and dirt. Reinsert the shipping foam (removed at time of centrifuge installation) into the...
Page 57 - APPENDIX A; clearances and electrical power are available.; without assistance from a lifting device or another person.; Installing the Instrument; system and be ignited by the motor.; area surrounding the centrifuge.; with the help of another person
PN B05529AF A-1 APPENDIX A Installation Introduction This chapter contains instructions for installing and connecting the centrifuge. Check that required clearances and electrical power are available. WARNING These centrifuges each weigh 48 kg/106 lb (non-refrigerated model) or 78 kg/172 lb (refrige...
Page 58 - Transportation Safety Devices; Nonrefrigerated Model Specifications
PN B05529AF A-2 Installation Transportation Safety Devices a. Make sure that the centrifuge front feet are fully supported on the table. b. Locate the centrifuge in an area with sufficient ventilation to allow for heat dissipation. c. Check that there are 7.6-cm (3-in.) clearances at the sides of th...
Page 60 - the name rating plate affixed to the centrifuge.; Test Run; condition following shipment. See
PN B05529AF A-4 Installation Test Run WARNING To reduce the risk of electrical shock, this equipment uses a three-wire electrical cord and plug to connect the centrifuge to earth-ground. To preserve this safety feature: • Make sure that the matching wall outlet receptacle is properly wired and earth...
Page 64 - Related Documents
© 2022 Beckman Coulter, Inc. All Rights Reserved www.beckman.com Related Documents Chemical Resistances for Beckman Coulter Centrifugation Products PN IN-175 Available in hard copy or electronic pdf by request. Available at www.beckman.com