Page 2 - Notices; C A U T I O N
Notices © Agilent Technologies, Inc. 2008 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla- tion into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United State...
Page 3 - Preface; Notation; • Softkey
N9360A Multi UE Tester Quick Reference Guide iii Preface Thank-you for purchasing the Agilent N9360A mobile communciations tester. • Before using the tester, the user is advised to read this manual carefully to ensure correct usage and also to fully utilize the tester capability. • This manual is a ...
Page 4 - THIS PAGE IS INTENTIONALLY LEFT BLANK.
iv N9360A Multi UE Tester Quick Reference Guide THIS PAGE IS INTENTIONALLY LEFT BLANK.
Page 5 - Contents; Legal Information; Caution and Safety Requirements; Safety Information; General Operation
N9360A Multi UE Tester Quick Reference Guide v Contents Notation I-iiiNotices I-iiiTrademarks I-iii 1 Legal Information Legal Information 1-2 Warranty 1-2 Technology Licenses 1-2 Restricted Rights Legend 1-2 Service And Support 1-3 Agilent On The Web 1-3 Agilent By Phone 1-3 2 Caution and Safety Req...
Page 7 - Agilent Technologies
1-1 Agilent Technologies 1 Legal Information Warranty 1-2 Technology Licenses 1-2 Restricted Rights Legend 1-2 Service And Support 1-3 Agilent On The Web 1-3 Agilent By Phone 1-3
Page 8 - Warranty
1-2 N9360A Multi UE Tester Quick Reference Guide 1 Legal Information Legal Information Warranty The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent discla...
Page 9 - Service And Support; Agilent On The Web; Double-click the link to; Agilent By Phone; Agilent Call Centers and Regional Headquarters; Manufacturing Address; Bayan Lepas Free Industrial Zone,
Legal Information 1 N9360A Multi UE Tester Quick Reference Guide 1-3 Service And Support Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer through your local Agilent Technologies Service Center. Agilent On The Web You can ...
Page 10 - THIS PAGE IS INTENTIONALLY LEFT BLANK
1-4 N9360A Multi UE Tester Quick Reference Guide 1 Legal Information THIS PAGE IS INTENTIONALLY LEFT BLANK
Page 11 - Caution and Safety Requirements; Safety Summary
2-1 Agilent Technologies 2 Caution and Safety Requirements Safety Summary 2-2 Safety Notices 2-2 Warning Label 2-2 General 2-3 When Operating The Tester 2-3
Page 12 - A warning label is stuck on the front panel of the Tester.; notice denotes a hazard. It calls attention to an; W A R N I N G
2-2 N9360A Multi UE Tester Quick Reference Guide 2 Caution and Safety Requirements Safety Information Safety Summary The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhe...
Page 13 - General; Make sure that the input signal level does not exceed the maximum; Do not turn off the Line switch on the rear panel of the Tester while
Caution and Safety Requirements 2 N9360A Multi UE Tester Quick Reference Guide 2-3 General When Operating The Tester W A R N I N G The protection provided by the N9360A tester may be impaired if the tester is used in a manner not specified by Agilent or the instructions on the display are not follow...
Page 15 - General Operation; Using a USB memory device
3-1 Agilent Technologies 3 General Operation Using a USB memory device 3-2 Using the Knob on the Front Panel 3-2 Preparation for Tests 3-3 GSM/W-CDMA/cdma2000 Mobile Phone Repair Process 3-4
Page 16 - Saving and Recalling Test Procedures:; To update firmware of the Tester, press the; FW Update; softkey to; Saving a Screen Image:; a screen image into a USB memory device, press the; Print Screen; Using the Knob on the Front Panel
3-2 N9360A Multi UE Tester Quick Reference Guide 3 General Operation Using a USB memory device A USB memory device can be used to save and recall the test procedures, to update the firmware of the Tester, and to save screen images as image files. Saving and Recalling Test Procedures: All settings fo...
Page 17 - Preparation for Tests; Typical Test Setup; Connecting RF Signal:; and enter the appropriate loss; Connecting a Printer:
General Operation 3 N9360A Multi UE Tester Quick Reference Guide 3-3 Preparation for Tests Figure 3-1 Typical Test Setup Connecting RF Signal: 1 Connect the mobile phone to the Tester using the user-supplied RF Cable, optional Antenna Coupler, or optional Shield Case. 2 On the [Loss] screen obtained...
Page 18 - Mobile Phone Repair Process
3-4 N9360A Multi UE Tester Quick Reference Guide 3 General Operation GSM/W-CDMA/cdma2000 Mobile Phone Repair Process Figure 3-2 Mobile Phone Repair Process A typical GSM, W-CDMA and cdma2000 mobile phone repair process at service centers is basically divided into two processes, Incoming Inspection a...
Page 19 - GSM System
4-1 Agilent Technologies 4 GSM System For Go/No-Go Testing 4-2 For Pinpointing Failures 4-3 For Troubleshooting 4-5 RF Channels 4-7 MS Power Control / Power Class 4-7 RX Level 4-9 RX Quality 4-10
Page 20 - Automatic Test; Press the; Start; After completing the test, check the results on the screen.
4-2 N9360A Multi UE Tester Quick Reference Guide 4 GSM System For Go/No-Go Testing Figure 4-1 [Automatic Test: Stand-by] Screen 1 Press the Automatic Test softkey on the [Initial] screen in GSM system. 2 Move the cursor to the “Procedure” field, and select a pre-defined test procedure. 3 Press the S...
Page 21 - For Pinpointing Failures
GSM System 4 N9360A Multi UE Tester Quick Reference Guide 4-3 For Pinpointing Failures Figure 4-2 [Automatic Test: Stand-by] Screen 1 Perform the steps described in “For Go/No-Go Testing.” 2 Press the Screen>> softkey to set screen mode to Detail or Value . Table 4-2 GSM Automatic test sequenc...
Page 22 - softkey to display the previous
4-4 N9360A Multi UE Tester Quick Reference Guide 4 GSM System 3 Rotate the knob to place the cursor in one of the Pass/Fail cells on the detail screen or test result cells on the value screen. Then, press the knob . 4 One of the following measurement result screens is displayed according to your sel...
Page 23 - For Troubleshooting; Manual Test; Turn the mobile phone on.; BS Call; softkey and respond to the call on the
GSM System 4 N9360A Multi UE Tester Quick Reference Guide 4-5 For Troubleshooting Figure 4-4 [Manual Test (GSM): Stand-by] Screen 1 Press the Manual Test softkey on the [Initial] screen. 2 Move the cursor to the “Procedure” field and select a pre-defined test procedure. 3 Turn the mobile phone on. 4...
Page 24 - Trigger
4-6 N9360A Multi UE Tester Quick Reference Guide 4 GSM System Figure 4-5 Various [Manual Test (GSM): Stand-by] Screens 7 While “Connection” is highlighted in the test f low, press the Trigger softkey to start single measurement or the Trigger Sing/Cont softkey to start continuous measurement. Press ...
Page 25 - Return; RF Channels; GSM RF channel frequencies and channel number; N O T E; GSM Power Control levels
GSM System 4 N9360A Multi UE Tester Quick Reference Guide 4-7 Press the Trigger Sing/Cont softkey again to terminate the continuous measurement. 9 In the graphical display, the marker function is available to read the data points by rotating the knob. On the power ramp graph display, the zoom functi...
Page 27 - RX levels
GSM System 4 N9360A Multi UE Tester Quick Reference Guide 4-9 RX Level Table 4-5 RX levels # RX Level # RX Level # RX Level 0 ≤− 110 dBm 22 –89 to –88 dBm 44 –67 to –66 dBm 1 –110 to –109 dBm 23 –88 to –87 dBm 45 –66 to –65 dBm 2 –109 to –108 dBm 24 –87 to –86 dBm 46 –65 to –64 dBm 3 –108 to –107 dB...
Page 28 - RX Quality
4-10 N9360A Multi UE Tester Quick Reference Guide 4 GSM System RX Quality Table 4-6 RX Quality # RX Quality # RX Quality 0 <0.2 % 4 1.6 to 3.2 % 1 0.2 to 0.4 % 5 3.2 to 6.4 % 2 0.4 to 0.8 % 6 6.4 to 12.8 % 3 0.8 to 1.6 % 7 >12.8 %
Page 33 - UARFCN; W-CDMA Frequency Band
W-CDMA System 5 N9360A Multi UE Tester Quick Reference Guide 5-5 UARFCN Mobile Phone Maximum Output Power Table 5-2 W-CDMA Frequency Band Band Uplink ; mobile phone transmit Downlink ; mobile phone receive General Additional General Additional I 9612 to 9888 — 10562 to 10838 — II 9262 to 9538 12, 37...
Page 36 - Print
6-2 N9360A Multi UE Tester Quick Reference Guide 6 cdma2000 System When "CDMA Mode" is set to MC-1x mode For Go/No-Go Testing Figure 6-1 [Auto Test (MC-1x): Stand-by] Screen 1 Press the Automatic Test softkey on the [Initial] screen in cdma2000 system 2 Move the cursor to the "Procedure&...
Page 37 - cdma2000 Automatic Test Sequence
cdma2000 System 6 N9360A Multi UE Tester Quick Reference Guide 6-3 Table 6-1 cdma2000 Automatic Test Sequence Step Action Needed Location Update Wait until the Location Update is complete. MS Call (Talk) Dial an arbitrary number and press the Off Hook key on the mobile phone. Talk Talk to the mobile...
Page 40 - Each test f low step is highlighted as it runs.
6-6 N9360A Multi UE Tester Quick Reference Guide 6 cdma2000 System When "CDMA Mode" is set to 1xEV-DO mode For Go/No-Go Testing Figure 6-4 [Auto Test (1xEV-DO): Stand-by] Screen 1 Press the Automatic Test softkey on the [Initial] screen in cdma2000 system 2 Move the cursor to the "Proced...
Page 43 - RF Channel; Radio System
cdma2000 System 6 N9360A Multi UE Tester Quick Reference Guide 6-9 RF Channel † The BAND15: AWS is available only when the Option C03 is installed in the Tester. Table 6-3 cdma2000 Frequency Band Radio System Channel Number BAND0: CEL US 1024 ~ 1323 991 ~ 1023 1 ~ 799 BAND1: PCS US 1024 ~ 1323 991 ~...