Page 2 - Notices
ii Notices © Agilent Technologies, Inc. 2003 No part of this manual may be repro-duced in any form or by any means (including electronic storage and retrieval or translation into a foreign lan-guage) without prior agreement and writ-ten consent from Agilent Technologies, Inc. as governed by United S...
Page 3 - Contents; Using the Probe Adapter 4; Reducing ESD Damage 9; Specifications and Regulatory Information 10
Contents Contents 1 N1022A Active Probe Adapter 1 General Safety Considerations 2Instrument Markings 3 Using the Probe Adapter 4 Inspecting the N1022A 4Connecting the N1022A 5 Static-Safe Workstation 8 Reducing ESD Damage 9 Specifications and Regulatory Information 10 Specifications 10Regulatory Inf...
Page 5 - N1022A Active Probe Adapter; C A U T I O N; C A U T I O N
1 N1022A Active Probe Adapter N1022A Active Probe Adapter The N1022A Active Probe Adapter provides the interface to connect an Agilent Infiniimax Active Probe to an Agilent 86100-series Infiniium DCA. To use the N1022A requires that firmware revision A.03.05 or later be installed on the Infiniium DC...
Page 6 - General Safety Considerations; W A R N I N G
2 General Safety Considerations General Safety Considerations This product has been designed and tested in accordance with the standards listed in the Manufacturer’s Declaration of Conformity, and has been supplied in a safe condition. The product documentation contains information and warnings whic...
Page 7 - Instrument Markings
3 Instrument Markings Instrument Markings The instruction manual symbol. The product is marked with this warning symbol when it is necessary for the user to refer to the instructions in the manual. The laser radiation symbol. This warning symbol is marked on prod-ucts which have a laser output. The ...
Page 8 - Using the Probe Adapter; adapter through the plug-in module front panel connector.
4 Using the Probe Adapter Using the Probe Adapter There are a few things to consider when connecting the N1022A Active Probe Adapter to an active probe and the Infiniium DCA. • The dynamic range of the system will be 3.2 V (6.4 Vp-p), which with probe off- set, covers most digital technologies. • Th...
Page 9 - Technologies sales office.
5 Connecting the N1022A Technologies sales office. • If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier as well as your Agilent Technologies sales office. Keep the shipping materials for the carrier's inspection. The Agilent office will arrange...
Page 10 - Figure 1 on page 7; Probe ID
6 Connecting the N1022A panel. Connecting to an Aux Power connector prevents the probe from being automatically calibrated. 3 Connect the probe output to the probe adapter input. The probe tip should not be connected to a device under test at this time. C A U T I O N The probe power cable connector ...
Page 11 - Figure 1. Probe Channel Dialog Box; Click; Close; . You can now change the probe tips or probe attenuation. You can; Using modules with 2.4 mm electrical input
7 Connecting the N1022A Figure 1. Probe Channel Dialog Box 6 Click Close . You can now change the probe tips or probe attenuation. You can also perform a probe calibration. You can access information about probe attenuation and probe calibration in the Infiniium DCA Help system. Using modules with 2...
Page 12 - Static-Safe Workstation; Figure 2. Static Safe Workstation
8 Static-Safe Workstation Static-Safe Workstation Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe work-station. The following figure shows an example of a static-safe work station using two types of ESD...
Page 13 - Reducing ESD Damage; moving any assembly from the unit.
9 Reducing ESD Damage Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate ESD protection when used alone. To ensure user safety, the static-safe accessories must provide at least 1 M Ω of isolatio...
Page 14 - Specifications and Regulatory Information; Specifications; Characteristics
10 Specifications and Regulatory Information Specifications and Regulatory Information This section lists specifications and characteristics and regulatory information of the probe adapter. Specifications The distinction between specifications and characteristics is described as fol-lows: • Specific...
Page 17 - Regulatory Information; This ISM device complies with Canadian ICES-001.; Table 4. Notice for Germany: Noise Declaration
13 Regulatory Information Regulatory Information Electrostatic Discharge Immunity - IEC 61000-4-2: 1995 passes criterion B. According to IEC 61000-4-2: 1995, system measurement may be affected when electrostatic discharges of 2 kV or more are applied to the outer sur-faces of the N1022A probe adapte...
Page 18 - Declaration of Conformity
14 Declaration of Conformity Declaration of Conformity
Page 19 - Service; “Agilent Technologies Service Offices” on page 17; Checking Probe Power Outputs
15 Service Service Service for the N1022A Active Probe Adapter is by replacement only. To order a new N1022A, contact your local Agilent Technologies Sales Office. Return the failed probe adapter to your local Agilent service center. If the probe adapter is still under warranty or is covered by an A...
Page 20 - Table 5. Probe Power Outputs; Pin
16 Returning the N1022A Returning the N1022A When the probe adapter is returned to an Agilent service office for servicing, it must be adequately packaged and have a complete description of the failure symptoms attached. Please notify the service office before returning your probe adapter for ser-vi...
Page 21 - Seal the shipping container securely.; Agilent Technologies Service Offices; Call Center
17 Agilent Technologies Service Offices C A U T I O N Cover electrical connectors to protect sensitive components from electrostatic damage. C A U T I O N Use original packaging or comparable. Instrument damage can result from using packaging materials other than the original materials. Never use st...
Page 23 - Index
Index Index 1 A Accessories , 4 Agilent offices , 17 C CE mark , 3 cleaning , 2 , 4 , 17 CSA mark , 3 D Declaration of Conformity , 14 E EMC requirements , 13 environmental specifications , 10 ESD reducing damage caused by ESD , 9 static-safe work station , 9 I inspecting , 4 ISM1-A , 3 N noise decl...