Agilent 8904A - Manual

Agilent 8904A

Agilent 8904A – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
Page: / 2
Loading the manual

8904A-04

SUPERSEDES: None

Continued

DATE: 01 January 1992

8904A Multifunction Synthesizer (Option 006/H16)

Serial Numbers:

2948A04593 / 2948A04693

Replacement of Audio Transformer to Correct Inverted Waveform

Parts Required:

9100-4833, Audio Transformer

Situation:

Some 8904A Multifunction Synthesizer (Option 006/H16) with serial numbers listed
above may have inverted OUTPUT 1 waveforms. This is because the primary of Audio
Transformer T1 was wound backwards, resulting in a 180 degree phase shift from the pri-
mary to the secondary. OUTPUT 2 waveforms are not affected.

The most noticeable waveform is the ramp because it is asymmetrical. Sinewave, square
and triangle waveforms are inverted but are less noticeable because they are symmetrical.
Normally the ramp waveform starts at the negative peak and ramps up to the positive
peak. Suspect units will exhibit an inverted ramp waveform, i.e., the ramp starts at posi-
tive peak and ramps down to the negative peak.

NOTE: Option 006/H16 is not indicated on the front panel LCD display when ann instru-
ment preset is performed. Verify if Option 006/H16 is installed by inspecting the option
labelon the rear panel of the unit.

1 of 2

AUTHOR: LHL

ENTITY: 1000

© 1991 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.

S

E

R

V

I

C

E

N

O

T

E

ADMINISTRATIVE INFORMATION

MODIFICATION RECOMMENDED

SERVICE NOTE CLASSIFICATION:

ACTION
CATEGORY:

LOCATION
CATEGORY:

CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER

RESPONSIBLE UNTIL: 1000 UNTIL: 01 January 1997

ADDITIONAL INFORMATION:

IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME

AVAILABILITY:

PRODUCT’S SUPPORT LIFE

RETURN
SCRAP
SEE TEXT

SERVICE
INVENTORY:

RETURN
SCRAP
SEE TEXT

USED
PARTS:

STANDARDS:

LABOR: 1.0 Hours

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.